Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K122984

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-10-22

    (26 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Atlas Gold PTA Dilatation Catheter is indicated for use in Percutaneous Transluminal Angioplasty of the iliac arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for post-dilatation of stents and stent grafts in the peripheral vasculature. This catheter is not for use in coronary arteries.

    Device Description

    The Atlas Gold PTA Dilatation Catheter is balloon catheter consisting of an over the wire catheter with a balloon fixed at the distal tip. The proprietary non-compliant, low profile balloon is designed to provide consistent balloon diameters and lengths even at high pressures. Two radiopaque markers delineate the working length of the balloon and aid in balloon placement. The coaxial catheter includes an atraumatic tip to facilitate advancement of the catheter to and through the stenosis. The over the wire catheter is compatible with 0.035" guidewire and is available in 80 and 120 cm working lengths. The proximal portion of the catheter includes a female luer lock hub connected to the inflation lumen, and a female luer-lock hub connected to the guidewire lumen.

    Packaged with every product is a profile reducing sheath that is positioned over the balloon for protection before use. A re-wrapping tool is also provided on the catheter shaft. This product is not manufactured with any latex.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Atlas Gold PTA Dilatation Catheter, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document explicitly states that the device "met all the predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs." However, it does not provide a specific table detailing the acceptance criteria values for each test alongside the performance results. Instead, it lists the types of tests performed.

    Therefore, a table cannot be constructed with specific numerical acceptance criteria and performance data. The reported device performance is broadly stated as follows:

    Acceptance Criteria CategoryReported Device Performance
    In Vitro Tests"The results from these tests demonstrate that the technological characteristics and performance criteria of the Atlas Gold PTA Dilatation Catheter is comparable to the predicate device and that it can perform in a manner equivalent to devices currently on the market for the similar intended use."
    Dimensional VerificationMet criteria (implied by overall statement)
    Media InteractionMet criteria (implied by overall statement)
    Equipment InterfaceMet criteria (implied by overall statement)
    Sheath CompatibilityMet criteria (implied by overall statement)
    Rated Burst PressureMet criteria (implied by overall statement)
    Balloon Burst ModeMet criteria (implied by overall statement)
    Catheter Shaft LeaksMet criteria (implied by overall statement)
    Catheter Shaft BurstMet criteria (implied by overall statement)
    FatigueMet criteria (implied by overall statement)
    Balloon Nominal PressureMet criteria (implied by overall statement)
    Balloon DistensibilityMet criteria (implied by overall statement)
    Inflation TimeMet criteria (implied by overall statement)
    Deflation TimeMet criteria (implied by overall statement)
    Balloon to Shaft TensileMet criteria (implied by overall statement)
    Catheter Shaft ElongationMet criteria (implied by overall statement)
    Hub to Shaft TensileMet criteria (implied by overall statement)
    Tip TensileMet criteria (implied by overall statement)
    Tip MorphologyMet criteria (implied by overall statement)
    TrackabilityMet criteria (implied by overall statement)
    Tip VisibilityMet criteria (implied by overall statement)
    Catheter Shaft VisibilityMet criteria (implied by overall statement)
    Marker Band VisibilityMet criteria (implied by overall statement)
    Marker Band AlignmentMet criteria (implied by overall statement)
    Balloon Burst Strength in Stent/Stent GraftMet criteria (implied by overall statement)
    Fatigue in Stent/Stent GraftMet criteria (implied by overall statement)
    Balloon Removal from Stent/Stent GraftMet criteria (implied by overall statement)
    Visual Inspection of PackagingMet criteria (implied by overall statement)
    Dye PenetrationMet criteria (implied by overall statement)
    Pouch Tensile StrengthMet criteria (implied by overall statement)
    Biocompatibility Tests"The results from these tests demonstrate that the subject device...is comparable to the predicate device and that it is considered safe and biocompatible for its intended use."
    CytotoxicityMet criteria (implied by overall statement)
    SensitizationMet criteria (implied by overall statement)
    Intracutaneous ReactivityMet criteria (implied by overall statement)
    Acute Systemic ToxicityMet criteria (implied by overall statement)
    HemocompatibilityMet criteria (implied by overall statement)
    Material Mediated PyrogenicityMet criteria (implied by overall statement)

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not specify a distinct "test set" in the context of clinical data for algorithmic performance. The tests described are in vitro (laboratory) tests on the device itself and biocompatibility tests. The document does not provide sample sizes for these tests, nor does it mention country of origin or whether the data is retrospective or prospective, as these are typically considerations for clinical studies involving patient data, not mechanical or biocompatibility testing.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    This information is not applicable. The evaluations were based on engineering specifications and laboratory testing protocols, not expert consensus on medical images or clinical outcomes.

    4. Adjudication Method for the Test Set:

    This information is not applicable. There's no mention of an adjudication process as would be used for subjective assessments or expert reviews of clinical data.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, an MRMC comparative effectiveness study was not conducted or mentioned. This type of study is relevant for evaluating the impact of AI on human reader performance, which is not the focus of this device's submission.

    6. Standalone (Algorithm Only) Performance Study:

    No, a standalone (algorithm only) performance study was not conducted or mentioned. The device is a physical medical instrument (catheter), not a software algorithm.

    7. Type of Ground Truth Used:

    The "ground truth" for the performance evaluation relies on established engineering specifications, test protocols, and recognized standards (e.g., ISO 10993-1:2009 for biocompatibility). For mechanical and performance tests, the ground truth is often defined by the design requirements and the physical properties expected of the device.

    8. Sample Size for the Training Set:

    This information is not applicable. There is no software algorithm or AI that requires a training set for this device. The evaluation is based on direct device testing.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable, as there is no training set mentioned for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1