Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K122916

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-10-19

    (28 days)

    Product Code
    Regulation Number
    870.4875
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The JETSTREAM System is intended for use in atherectomy of the peripheral vasculature and to break apart and remove thrombus from upper and lower extremity peripheral arteries. It is not intended for use in coronary, carotid, iliac or renal vasculature.

    Device Description

    The Jetstream Systems are rotational atherectomy catheter systems designed with either a fixed (Jetstream G3 SF, Jetstream G3 SF 1.6) or an expandable (Jetstream Navitus L, Jetstream Navitus) cutting tip intended for use in debulking and treating vascular disease in the peripheral vasculature. Separate lumens within the Catheter allow for continuous aspiration and infusion during device use. Excised tissue, thrombus, and fluid are aspirated from the peripheral treatment site through a port in the Catheter tip to an external collection bag located on the Console. The distal portion of the Catheter also possesses infusion ports that provide continuous infusion of sterile saline during the atherectomy procedure.

    The Jetstream Systems consist of two primary components: a Catheter with Control Pod and a Console, which are packaged separately. Each of these system components is described generally as follows:

    • Jetstream Catheter with Control Pod: A sterile, single-use unit consisting of an electrically-driven Catheter with attached Control Pod. As with the predicate device, the modified Jetstream Catheter utilizes a differentially cutting tip and includes both aspiration and infusion capabilities and the Control Pod provides a user interface with keypad controls. The unit, its electrical connectors, tubing, and aspirant collection bag are packaged in a double pouched tray.
    • PV Console: A reusable compact PV Console, with two (2) peristaltic pumps for aspiration and infusion, power supply, system controller, keypad interface, and LED indicators for device operational status. The PV Console mounts on a standard I.V. stand and remains outside the sterile field during the procedure.

    The primary modification of this 510(k) removes the tachometer function (i.e., speed/RPM inout) in the Control Pod, and correspondingly removes the rotational speed displays on the PV Console. This modification applies to the entire family of Jetstream Systems.

    AI/ML Overview

    The provided 510(k) summary (K122916) describes modifications to the JETSTREAM atherectomy systems and compares them to predicate devices. This submission focuses on engineering testing to demonstrate substantial equivalence, rather than a clinical study evaluating diagnostic or treatment efficacy. Therefore, much of the requested information about clinical study design, ground truth, and expert evaluation is not applicable or not present in this document.

    Here's an analysis based on the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    System Reliability/Life TestResults demonstrate that the technological and performance characteristics of the modified Jetstream Systems are comparable to the predicate Jetstream Systems.
    Aspiration Efficiency & Crossing TimeResults demonstrate that the technological and performance characteristics of the modified Jetstream Systems are comparable to the predicate Jetstream Systems.
    Label Cleanability and DurabilityResults demonstrate that the technological and performance characteristics of the modified Jetstream Systems are comparable to the predicate Jetstream Systems.
    Auditory FeedbackResults demonstrate that the technological and performance characteristics of the modified Jetstream Systems are comparable to the predicate Jetstream Systems.

    Note: The document states that performance standards do not currently exist for these devices and none are established under Section 514. The "acceptance criteria" here are derived from the identified in vitro tests performed to support substantial equivalence.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the exact sample sizes (e.g., number of catheters tested, number of cycles) for the in vitro tests. The testing performed was in vitro (bench testing), not on human subjects. Therefore, data provenance related to country of origin or retrospective/prospective is not applicable.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This submission concerns engineering and performance testing of a physical device, not an AI or diagnostic device requiring expert interpretation for ground truth.

    4. Adjudication Method for the Test Set

    Not applicable. There was no clinical test set requiring expert adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The submission focuses on in vitro engineering validation of a device modification, not a clinical study comparing human reader performance with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This device is a mechanical atherectomy system; it does not involve algorithms or AI that would have a "standalone" performance.

    7. The Type of Ground Truth Used

    The "ground truth" for the in vitro engineering tests would be derived from quantifiable physical measurements and established engineering standards to determine if the modified device performed comparably to the predicate device. For example, in aspiration efficiency tests, the ground truth would be precise measurements of tissue removal rate or pressure differentials, compared against established performance benchmarks or the predicate device's performance.

    8. The Sample Size for the Training Set

    Not applicable. This device does not use machine learning or AI, and therefore does not have a "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set for an AI model, ground truth establishment for a training set is irrelevant.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1