510(k) Data Aggregation

    K Number
    K122702

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-03-18

    (195 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BioFriend™ BioMask™ N95 surgical respirator is a single use NIOSH-approved, disposable N95 surgical respirator with a hydrophilic plastic coating on the outer layer (active ingredient: citric acid 2% w/w, a pH-lowering agent), and a second inner layer treated with metal ions (active ingredients: copper 1.6% w/w and zinc 1.6% w/w which form ionic bonds with negatively-charged side-groups on influenza viruses).

    The BioFriend™ BioMask™ N95 surgical respirator inactivates 99.99% of tested influenza viruses on five minutes contact with the surface of the respirator in laboratory (in vitro) tests against the following seasonal, pandemic, avian, swine and equine influenza viruses: Influenza A subtypes and strains H1N1 (the 2009 pandemic flu subtype A/Californial04/2009, A/Brisbane/59/2007, A/Wisconsin/10/1998, A/New Jersey/8/1976, A/PuertoRico/8/1934), H3N2 (A/Brisbane/10/2007, A/Wisconsin/67/2005), H2N2 (A/2/Japan/305/1957); the bird flu subtypes: H5N1 (NIBRG-14), H9N2 (A/TurkeyWisconsin/1966), H5N2 (A/Duck/PA/10218/84); the swine fly sublype: H1N1 (A/Swine/1976/1931); the equine flu subtype: H3N8 (A/Equine/2/Miami/1963); and Influenza B strains: (B/Florida/4/2006, B/Lee/1940), under tested contact conditions. Correlation between in vitro testing results and any clinical event has not been tested.

    The BioFriend™ BioMask™ N95 surgical respirator, Model: Professional BF-200-3013AN is flatfolded and expands into a convex-shaped mask with polyamide/spandex elastic head-loops to secure the mask to the user's face, and a malleable aluminum strip positioned above the nose for a tighter seal around the nose and face. The device is intended to be worn during seasonal Influenza A or Influenza B, and an Influenza A or Influenza B pandemic. It is intended for occupational use, to help reduce wearer exposure to pathogenic biological airborne particulates, and to protect against the transfer of micro-organisms, body fluids, and particulate material.

    Device Description

    The BioFriend™ BioMask™ N95 surgical respirator, Model: Professional (BF-200-3013AN) is flatfolded and expands into a convex-shaped mask with polyamide/spandex elastic head-loops to secure the mask to the user's face, and a malleable aluminum strip positioned above the nose for a tighter seal around the nose and face. The respirator is comprised of four layers of material: an outer layer of spun-bond polypropylene, a second layer of cellulose/polyester, a third layer of melt-blown polypropylene filter material and an inner (fourth) layer of spun-bound polypropylene. All of the construction materials used in this device are typical construction materials commonly used in surgical facemasks and N95 surgical respirators and being used in current legally marketed devices. The outer active layer of the respirator is coated with a hydrophilic plastic. The second inner layer is treated with copper and zinc ions. Both layers inactivate influenza viruses using different mechanisms of action.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the BioFriend™ BioMask™ N95 Surgical Respirator, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The provided document details performance against recognized standards, specifically ASTM F2100-11 for Medical Face Masks and NIOSH 42 CFR 84 for Respirator Certification. It also includes virucidal efficacy claims.

    DescriptionTest MethodAcceptance Criteria (Predicate/Standard)Reported Device Performance (N95 Surgical Respirator Professional BF-200-3013AN)
    ASTM F2100-11 Medical Face Mask Standards (Level 3)
    Bacterial Filtration Efficiency (BFE)ASTM F2101> 99% (3M N95); 99.9% (BioMask Predicates)> 99.9 %
    Sub-micron (0.1 µm) Particulate Filtration EfficiencyASTM F2299N/A (3M N95); 99.7% (BioMask Predicates)> 99.9 %
    Differential Pressure (Delta-P)MIL-M-36954C4.9 mm H2O/cm² (3M N95); 2.6 mm H2O/cm² (BioMask Predicates)4.9 mm H2O/cm²
    Resistance to Penetration by Synthetic BloodASTM F1862160 mm Hg160 mm Hg
    Flame Spread16 CFR Part 1610Class 1Class 1
    NIOSH 42 CFR Part 84 - N95 Respirator Standards
    Particulate Filtration Efficiency (NaCl)42 CFR 84.181Pass (3M N95); N/A (BioMask Predicates)98 %
    Breathing Resistance (Inhalation)42 CFR 84.180Pass (3M N95); N/A (BioMask Predicates)8.0 mm H2O
    Breathing Resistance (Exhalation)42 CFR 84.180Pass (3M N95); N/A (BioMask Predicates)8.9 mm H2O
    Virucidal EfficacyIn vitro laboratory tests99.99% viral inactivation at 5 minutes99.99% (≥4-logs) inactivation of 15 influenza strains

    2. Sample Sizes Used for the Test Set and Data Provenance:

    The document does not explicitly state the sample sizes for each specific test (e.g., number of masks tested for BFE, PFE, etc.). However, it indicates that "All testing was carried out in compliance with the Good Laboratory Practice (GLP) regulations," implying standardized and sufficient sample sizes as per GLP.

    • Data Provenance: The tests were conducted by independent laboratories:
      • Nelson Laboratories Inc., Salt Lake City, UT, USA: For bacterial filtration efficiency, differential pressure, sub-micron particulate filtration efficiency, resistance to penetration by synthetic blood, and flammability.
      • Microbiotest Inc., Sterling, VA, USA: For Virucidal Efficacy testing.
    • Retrospective or Prospective: These appear to be prospective nonclinical laboratory tests specifically conducted for the device's clearance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    This section is Not Applicable as the studies described are non-clinical, laboratory-based performance tests of the device's physical and antimicrobial properties, not assessments requiring expert interpretation of medical images or patient outcomes. The "ground truth" is established by the standardized test methods themselves.

    4. Adjudication Method for the Test Set:

    This section is Not Applicable for the same reasons as point 3. There is no human interpretation or adjudication required for these objective performance tests.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The studies presented are laboratory performance tests of the device's material properties and virucidal efficacy, not clinical studies involving human readers or patient cases.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

    This concept typically applies to AI/software as a medical device. Since the device is an N95 surgical respirator, and not an AI algorithm, this section is Not Applicable. The performance described is the standalone performance of the physical device.

    7. The Type of Ground Truth Used:

    The ground truth used for these studies is based on:

    • Standardized Test Methods and Definitions: The "ground truth" for parameters like BFE, PFE, differential pressure, fluid penetration, and flame spread is defined by the specific ASTM, MIL, 16 CFR, and NIOSH standards and their associated methodologies.
    • Laboratory-Confirmed Viral Inactivation: For virucidal efficacy, the "ground truth" is the empirically measured reduction in viable virus particles (e.g., cell culture assays) after contact with the mask material under controlled laboratory conditions, as observed and quantified by Microbiotest Inc.

    8. The Sample Size for the Training Set:

    This section is Not Applicable. The BioFriend™ BioMask™ N95 Surgical Respirator is a physical medical device, not an AI/machine learning model that requires a 'training set.' Its performance characteristics are determined through manufacturing specifications and subsequent testing against established standards.

    9. How the Ground Truth for the Training Set Was Established:

    This section is Not Applicable as there is no training set for this type of device.