    Intended Use

    The TotalCare® Bed System is intended to be used to treat or prevent pulmonary or other complications associated with immobility; to treat or prevent pressure ulcers; or for any other use where medical benefits may be derived from either Continuous Lateral Rotation Therapy or Percussion/Vibration Therapy. The TotalCare® Bed System is intended to provide a patient support to be used in health care environments. The TotalCare® Bed System may be used in a variety of settings including, but not limited to, acute care, including critical care, step down/progressive care, medical/surgical, high acuity sub-acute care, post anesthesia care unit (PACU), and sections of the emergency department (ED). The TotalCare® Bed System is capable of being used with a broad patient population as determined appropriate by the caregiver or institution.

    The TotalCare® Bed System, dependent upon model, is capable of supporting patient populations up to 500 lbs.

    Device Description

    The TotalCare® Modular Therapy (Bed) System is a control unit (GUI) and frame which can house several therapeutic surfaces. The system can articulate to provide different positions for treatment, sleeping, up-in-bed, and other physician recommended positions. The Graphical Caregiver Interface/ Graphical User Interface can be set for alarms such as patient egress and head-of-bed angle.

    AI/ML Overview

    The Hill-Rom TotalCare® Modular Therapy (Bed) System is a medical bed designed to provide patient support and therapy.

    Here's an analysis of the provided text regarding acceptance criteria and supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySourceReported Device Performance
    Functional, Performance, Safety, and EfficacyHill-Rom's verification and validation based on international standardsMeets functional, performance, safety, and efficacy specifications and requirements in compliance with listed international standards.
    Compliance with International StandardsIEC 60601-1 (General Safety)Compliant
    IEC 60601-1-2 (EMC Requirements)Compliant
    IEC 60601-1-4 (Programmable Electrical Medical Systems)Compliant
    IEC 60601-2-38 (Safety of Electrically Operated Hospital Beds)Compliant
    Substantial Equivalence to Predicate DevicesFDA's 510(k) review processDetermined to be substantially equivalent to legally marketed predicate devices.
    Intended UseDevice Description and Indications for UseFunctions as a patient support system in healthcare environments; treats/prevents pulmonary/immobility complications, pressure ulcers; provides Continuous Lateral Rotation Therapy or Percussion/Vibration Therapy; supports patients up to 500 lbs.

    2. Sample Size for Test Set and Data Provenance

    The provided summary does not mention any specific sample sizes for a test set related to clinical performance or algorithm evaluation. Instead, it explicitly states:

    • "Clinical testing was not required for determination of substantial equivalence."
    • "A literature review was compiled of articles related to safe patient handling and benefits of therapy surfaces and beds of this type for patients experiencing prolonged immobility."

    This indicates that the device's acceptance was primarily based on non-clinical performance and substantial equivalence to predicate devices, without a dedicated clinical test set for this specific submission.

    3. Number of Experts and Qualifications for Ground Truth

    Given that "Clinical testing was not required," there is no mention of experts used to establish ground truth for a test set in the context of this 510(k) summary. The evaluation relied on compliance with engineering standards and comparison to existing products.

    4. Adjudication Method for the Test Set

    As there was no clinical test set requiring human assessment, no adjudication method is described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done as clinical testing was not required. The device is a medical bed, not an AI-assisted diagnostic tool for which such studies are typically performed.

    6. Standalone (Algorithm Only) Performance Study

    No standalone performance study was done as the device is a medical bed and not an algorithm-based system requiring such evaluation. The evaluation focused on the physical device's compliance with safety and performance standards.

    7. Type of Ground Truth Used

    The "ground truth" for the device's acceptance was:

    • Compliance with international engineering and safety standards: This forms the basis for the non-clinical performance summary (e.g., IEC 60601 series).
    • Substantial equivalence to predicate devices: This involves comparing the new device's technological characteristics and intended use to existing, legally marketed similar devices, rather than a clinical ground truth like pathology or outcome data.

    8. Sample Size for the Training Set

    Since there are no indications of a machine learning or AI algorithm being developed or deployed within this device, there is no mention of a training set sample size.

    9. How Ground Truth for the Training Set Was Established

    As there is no mention of a training set, the question of how its ground truth was established is not applicable.

    In Summary:

    The Hill-Rom TotalCare® Modular Therapy (Bed) System's acceptance was based on non-clinical performance validation against established international safety and performance standards, and demonstrating substantial equivalence to predicate medical beds. Clinical testing was not deemed necessary for this 510(k) submission, and therefore, typical metrics associated with AI device evaluation (like test set size, expert ground truth, MRMC studies, or training sets) are not present in this documentation.