Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K122338

    Validate with FDA (Live)

    Date Cleared
    2013-11-05

    (460 days)

    Product Code
    Regulation Number
    862.1345
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Prodigy Preferred Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, palm, calf or thigh. The Prodigy Preferred Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

    The Prodigy Preferred Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Prodigy Preferred Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).

    The Prodigy No Coding Test Strips are for use with the Prodigy Preferred Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, palm, calf or thigh.

    Device Description

    The Prodigy Preferred Blood Glucose Monitoring System consists of a meter and Prodigy No Coding Test Strips. The system utilizes an electrochemical method-based meter and dry reagent biosensor (test strips) for blood glucose testing. The size of the current is proportional to the amount of glucose present in the sample, providing a quantitative measurement of glucose in fresh whole blood and control solutions.

    The Prodigy Preferred Blood Glucose Monitoring System is marketed as a meter only with a carrying case, battery, Owner's Manual, Quick Reference Guide, Logbook, and Warranty Card. The Prodigy Preferred Blood Glucose Monitoring System is also marketed as a meter kit with a carrying case, battery, Owner's Manual, Quick Reference Guide, Logbook, and Warranty Card, Prodigy Lancing Device, Prodigy Lancets, Prodigy No Coding Test Strips, and Control Solution. The Prodigy No Coding Test Strips utilizes the active enzyme is Glucose Oxidase, derived from Aspergillus niger.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study information for the Prodigy® Preferred Blood Glucose Monitoring System, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    For many of the tests, the acceptance criterion specifically states "met the acceptance criteria" or "less than X" while also reporting the actual performance as "less than X" or "within acceptance criteria," implying that the stated performance is the acceptance criterion itself, or that the performance achieved the desired threshold.

    Test CategoryAcceptance CriteriaReported Device Performance
    Precision Test< 5.0 mg/dL for glucose < 75 mg/dL (pooled and maximum SD); < 5.0% for glucose ≥ 75 mg/dL (pooled and maximum CV); Maximum individual bias < 10% compared with YSI2300.Pooled and maximum SD were < 5.0 mg/dL at glucose < 75 mg/dL; Pooled and maximum CV were < 5.0% at glucose ≥ 75 mg/dL; Maximum individual bias was < 10% compared with YSI2300.
    LinearityCorrelation coefficient > 0.95; 100% of individual bias within ±10% between 20-600 mg/dL.Correlation coefficient > 0.95; 100% of the bias of individual glucose results fell within ±10%. Linearity acceptable between 20-600 mg/dL.
    System AccuracyFor blood glucose level < 75 mg/dL, 95% of results within ±15 mg/dL or 95% of results within ±20% for blood glucose level ≥ 75 mg/dL (ISO 15197 requirement).>95% of all alternative site tested results fell within the acceptance criteria (15mg/dL for blood glucose <75 mg/dL and 20% for blood glucose ≥75mg/dL). Met ISO 15197 requirements.
    InterferenceNo obvious interference (implied: bias compared to control pool within ~10%); No significant effect from specified normal/therapeutic concentrations of listed substances.All bias of test results were ~ 10% compared with the measurements of the controlled pool. No obvious interference observed at therapeutic or physiological levels at three blood glucose levels. Specified substances (uric acid, ascorbic acid, acetaminophen, etc.) at normal/therapeutic concentrations did not significantly affect results.
    HematocritAll individual difference < ±10% compared with individual YSI mean, for HCT 20-60%; All SD < 5mg/dL and all CV < 5%.When HCT between 20% to 60%, all individual difference compared with individual YSI mean were < ±10%. All individual bias compared with YSI mean in HCT 40% was < ±10%. All SD were < 5mg/dL and all CV were < 5%. Accepted HCT range 20% to 60%.
    Volume VerificationSample volumes between 0.7 to 1.5uL fall within acceptable criteria. (Implied: accurate results)Sample volumes between 0.7 to 1.5uL fell within acceptable criteria. (Required at least 0.7uL for accurate results).
    Measurement Internal Determination Test (High Temp Exposure)Glucose measurements with check strip before and after high temperature exposure < 2.0 mg/dL difference; Mean difference of before and after control solution measurements (Level N and H) < 5.0 mg/dL.Glucose measurements with check strip before and after high temperature exposure < 0.2 mg/dL difference. Mean difference of before and after control solution measurements (Level N and H) < 3.6 mg/dL. (Performance exceeded acceptance criteria). The meters passed.
    Altitude StudyIndividual results fall within ± 10% from 91 to 3,402 meters altitude.Individual results fell within ± 10% at altitude from 91 to 3,402 meters. No significant effects from 298 feet to 11,161 feet (91 to 3,402 meters).
    Storage Condition (Test Strips)All individual bias < ± 10%. (Implied: stability for 24 months unused, 90 days opened).All individual bias < ± 10%. Unused test strips stable for 24 months and 90 days in used vials even after transportation.
    Temperature for EquipmentMaximum individual bias < 10% compared with YSI 2300.Maximum individual bias of meters treated to extreme temperature measurements < 10% by using venous whole blood.
    Humidity Test for EquipmentMaximum individual bias < 10% compared with YSI 2300.Maximum individual bias of meters treated to extreme humidity measurements < 10% by using venous whole blood.
    Operating Condition (Test Strips)Performance of individual bias < ± 10%.Performance of individual bias of test strips evaluated in extreme environments < ± 10%.
    Low Temperature TestGlucose measurements with check strip before and after low temperature exposure < 2.0 mg/dL difference; Mean difference of before and after control solution measurements (Level N and H) < 5.0 mg/dL.Glucose measurements with check strip before and after low temperature exposure < 0.2 mg/dL difference. Mean difference of before and after control solution measurements (Level N and H) < 3.9 mg/dL. (Performance exceeded acceptance criteria). The meters passed.
    Humidity Detection LimitAll CV of test results < 4% for different humidity levels; Total CV of average values < 4% for "Normal" and "High" control solution (RH 10 ~ 85%).All CV of test results < 4%. Total CV of average values < 4% for "Normal" and "High" control solution for RH 10 ~ 85%.
    High Temperature TestGlucose measurements with check strip before and after high temperature exposure < 2.0 mg/dL difference; Mean difference of before and after control solution measurements (Level N and H) < 5.0 mg/dL.Glucose measurements with check strip before and after high temperature exposure < 0.2 mg/dL difference. Mean difference of before and after control solution measurements (Level N and H) < 3.6 mg/dL. (Performance exceeded acceptance criteria). The meters passed.
    Temperature TestAll precisions of test results < 4%; Total CV of average values < 4% for "Normal" and "High" control solution (10 to 40°C).All precisions of test results < 4%. Total CV of average values < 4% for "Normal" and "High" Prodigy control solution for 10 to 40°C. Passed operating (1040℃) and storage (440℃) temperatures.
    Robustness TestAppearance, structure, and function of meters regular after cleaning/disinfection (5,000 times for single-patient use, 156 times for lancing devices); All individual bias < 10 mg/dL (< 75 mg/dL glucose) and < ± 10% (≥ 75 mg/dL glucose) compared with YSI mean.Appearance, structure, and function of meters regular after cleaning/disinfection (5,000 times for single-patient use, lancing devices 156 times). All individual bias compared with YSI mean < 10 mg/dL at glucose < 75 mg/dL and < ± 10% at glucose ≥ 75 mg/dL.
    Disinfection Efficacy ValidationHBV of clinical sera could be efficaciously removed from the Blood Glucose Meter after completing the designed disinfection procedure. (Implied: robust disinfection method to protect users).HBV of clinical sera could be efficaciously removed from the Blood Glucose Meter provided by OKBiotech CO., Ltd after completing the designed disinfection procedure. Demonstrates robust method to protect users from HBV infection.
    Software ValidationMeets the requirements of FDA's guidance document "Guidance for the Content of Pre-market Submission for Software Contained in Medical Devices".Met the requirements of FDA's guidance document.
    Human Factor StudyEase of Use = high percentage, Label Comprehension = high percentage. (Specific threshold not provided, but implicitly met by reported percentages)Ease of Use = 98.5%; Label Comprehension = 99.5%.
    Electrical TestingPass EN 61236-1 EMC Test; Pass EN 60601-1-2 EMC Test; Pass FCC CFR 47 18 Subpart C.Passed all listed electrical tests (EN 61236-1 EMC, EN 60601-1-2 EMC, FCC CFR 47 18 Subpart C).

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes for the test set for each specific study. It frequently refers to "test results," "data," or "blood samples" without providing the precise number of participants or measurements.

    • Provenance: Not explicitly stated, but the studies involve "fresh capillary whole blood samples" and "venous whole blood" in a medical device context, suggesting a clinical or laboratory setting. The approval is from the US FDA, implying testing relevant to US regulatory standards. It does not state whether the studies were retrospective or prospective.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • The document mentions "glucose analyzer YSI2300" as the reference method for ground truth in several performance tests (Precision, Linearity, System Accuracy, Hematocrit, Temperature for Equipment, Humidity for Equipment, Robustness).
    • The YSI 2300 STAT Plus Glucose & Lactate Analyzer is a laboratory-grade reference instrument. Thus, the ground truth is established by a highly accurate and precise analytical instrument, not by human experts. The qualifications of the operators of this instrument are not detailed, but it's implied that they are qualified laboratory personnel.

    4. Adjudication Method for the Test Set

    • No adjudication method by human experts is mentioned for the test set. The direct comparison is against a reference instrument (YSI2300).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No multi-reader multi-case (MRMC) comparative effectiveness study is mentioned. The device is a blood glucose meter, which provides a direct numerical reading, not an interpretation like an imaging device that would typically involve human readers. Therefore, the concept of "human readers improve with AI vs without AI assistance" does not apply in this context.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • Yes, the performance tests described (Precision, Linearity, System Accuracy, Interference, etc.) represent standalone performance of the device (meter + test strip system). The device provides a direct quantitative measurement, and these studies evaluate its accuracy and performance independently of human interpretation, although human users operate the device and apply the blood sample. The "Human Factor Study" addresses ease of use and label comprehension for the human operator, but the core performance metrics are device-centric.

    7. Type of Ground Truth Used

    • The primary ground truth used for analytical performance studies is comparison to a laboratory reference instrument, specifically the YSI 2300 STAT Plus Glucose & Lactate Analyzer. This is a highly accurate and precise method for determining blood glucose levels. For the Disinfection Efficacy, the ground truth was the efficacious removal of HBV as per FDA guidance.

    8. Sample Size for the Training Set

    • The document does not provide information on a "training set" sample size. This type of medical device (blood glucose meter) is typically developed through analytical validation of its electrochemical principles and biochemical components, rather than through machine learning model training on large datasets in the way an AI-driven diagnostic would be. The "design verification and validation testing" refers to broader engineering and performance validation rather than machine learning training.

    9. How Ground Truth for the Training Set Was Established

    • As no "training set" for a machine learning model is mentioned or implied, the method for establishing its ground truth is not applicable. The device's "training" in a broad engineering sense would involve iterative design, prototyping, and testing against known glucose concentrations and reference methods to optimize its performance characteristics.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1