510(k) Data Aggregation

    K Number
    K122306

    Validate with FDA (Live)

    Date Cleared
    2013-03-08

    (219 days)

    Product Code
    Regulation Number
    870.3470
    Age Range
    All
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Synovis Collagen Matrix is intended for the reconstruction and repair of the pericardium.

    Device Description

    Synovis Collagen Matrix is an implantable biologic patch comprised of non-crosslinked bovine pericardium.

    AI/ML Overview

    The Synovis Collagen Matrix is an implantable biologic patch made from non-crosslinked bovine pericardium, intended for the reconstruction and repair of the pericardium. This device received 510(k) clearance (K122306) based on substantial equivalence to predicate devices, primarily Veritas® Collagen Matrix.

    The device's safety and performance were evaluated through non-clinical testing, including bench testing and animal studies.

    Acceptance Criteria and Device Performance:

    Acceptance Criteria CategorySpecific TestReported Device PerformanceStudy Type
    Mechanical PropertiesSuture retentionBench testing results support performance requirementsBench Testing
    ThicknessBench testing results support performance requirementsBench Testing
    Amine indexBench testing results support performance requirementsBench Testing
    TensileBench testing results support performance requirementsBench Testing
    BurstBench testing results support performance requirementsBench Testing
    Sterility & PackagingSterilization validationBench testing results support performance requirementsBench Testing
    Packaging and shelf-lifeBench testing results support performance requirementsBench Testing
    BiocompatibilityBiocompatibilityPerformed in accordance with ISO 10993-1.Bench Testing
    Systemic ToxicitySystemic toxicologyNo detectable systemic or site-specific toxicity at all time points.Swine Animal Model
    Tissue Integration/RemodelingHistological examination (Swine)Device was readily remodeled and integrated into the adjacent host connective tissue.Swine Animal Model
    Healing & Mechanical IntegrityPneumonectomy stump repair, chest wall defect repairNo systemic or implant site specific toxicity; minimal inflammation, minimal foreign body response or fibrosis; complete healing of bronchial stump with no air leaks; repaired chest wall defects had sufficient mechanical integrity and remained hernia-free for 90 days.Canine Animal Model
    Remodeling (Canine)Histological analysis (Canine)Undergoes extensive remodeling through host tissue integration, leading to connective tissue indistinguishable from adjacent host tissue.Canine Animal Model
    Adhesion FormationAdhesion formationDemonstrates minimal tissue attachment to the device in case of direct contact with viscera.Rat Animal Model

    Details of the Studies:

    1. Sample size used for the Test Set and Data Provenance:

      • Bench Testing: The document does not specify the exact sample sizes for each bench test (suture retention, thickness, amine index, tensile, burst, sterilization validation, packaging, shelf-life, biocompatibility). These tests are typically conducted in a laboratory setting. Data provenance is implied to be from the manufacturer's internal testing.
      • Animal Studies:
        • Swine Model: Used for systemic toxicology and histological examination. The exact number of swine is not specified.
        • Canine Model: Used for pneumonectomy stump repair and chest wall defect repair. The exact number of canines is not specified.
        • Rat Model: Used for adhesion formation evaluation. The exact number of rats is not specified.
      • All animal studies were conducted prospectively as part of the device evaluation process. The country of origin for the data is not specified but is assumed to be part of the manufacturer's regulatory submission process.
    2. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

      • The studies described are non-clinical (bench and animal studies). Ground truth for these studies is established through objective measurements (e.g., mechanical properties, chemical analysis) and histopathological examination by qualified veterinarians and pathologists, typically employed by contract research organizations or the device manufacturer. The specific number and qualifications of these experts are not provided in the summary.
    3. Adjudication Method for the Test Set:

      • Not applicable as these were non-clinical bench and animal studies, not human clinical trials involving expert review and adjudication of patient data for diagnostic accuracy.
    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

      • No. This type of study is relevant for diagnostic imaging devices where human readers interpret medical images with and without AI assistance. The Synovis Collagen Matrix is an implantable surgical patch, not an imaging device.
    5. Standalone Algorithm Performance:

      • Not applicable. This device is a biocompatible surgical implant, not a standalone algorithm.
    6. Type of Ground Truth Used:

      • For bench testing: Objective measurements of physical and chemical properties.
      • For animal studies: Histopathology (histological examination), clinical observation (e.g., absence of air leaks, hernia-free), and systemic toxicology (laboratory analysis).
    7. Sample Size for the Training Set:

      • Not applicable. This device is not an AI/ML algorithm that requires a training set. The term "training set" is used for machine learning models, not for traditional medical devices like surgical implants. The animal studies served as preclinical evaluation, not "training."
    8. How the Ground Truth for the Training Set Was Established:

      • Not applicable, as there is no training set for this device.