Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K122064

    Validate with FDA (Live)

    Date Cleared
    2013-05-07

    (298 days)

    Product Code
    Regulation Number
    862.3650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ForSure One Step Buprenorphine Drug Cup Test Device is an immunochromatographic assay for the qualitative determination of Buprenorphine in human urine at cutoff level of 10 ng/mL. The test is intended for prescription and overthe-counter use.

    The buprenorphine assay will yield preliminary positive results when buprenorphine is ingested at or above therapeutic doses. There are no uniformly recognized drug levels for buprenorphine in urine. The ForSure One Step Buprenorphine Drug Cup Test Device shows the drug was or was not present at the cutoff level. The assay provides only a preliminary analytical test result.

    A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Mass Spectrometry (LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.

    Device Description

    ForSure" One Step Buprenorphine Drug Cup Test Device is a convenient specimen collection cup with a built-in strip holder which is able to hold the QuikStrip of Buprenorphine within the container. The membrane of the test strip is coated with goat anti-mouse antibody and Buprenorphine-Bovine serum albumin conjugate. The sample pad contains a colloidal gold-labeled mouse monoclonal anti-Buprenorphine antibody. As the test sample flows through the absorbent device, the Colloidal Gold labeled antibody-conjugate binds to the free drug in the specimen forming an antibody-conjugate in the test reaction zone and will not produce a magenta color band when the drug is above the detection level of 10 ng/ml of Buprenorphine. Unbound colloidal gold labeled antibody conjugate binds to the reagent in the negative control zone, producing a magenta color band, demonstrating that the reagents and device are functioning correctly. A Negative specimen produces two distinct color bands in both the test line and control area. A positive specimen produces only one color band in the control area. There is no meaning attributed to color or its intensity for either line. The test is easy and fast allowing the user to visually read the screen for abuse of drugs without the need for any other instrumentation to determine results. The tester will obtain a result in five minutes. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. This test is the first step in determining whether there are drugs in the urine. If the Buprenorphine result shows 1 line (preliminary positive) you should send the sample for laboratory testing.

    AI/ML Overview

    The provided document describes the ForSure™ One Step Buprenorphine Drug Cup Test Device (K122064). This summary focuses on the acceptance criteria and study information as requested.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    SensitivityAcceptance: ≥10ng/ml Buprenorphine in urine show positive result (Implied from predicate and subject device consistency)
    Total PrecisionAcceptance: No appreciable within and inter-lot variation for positive and negative spiked samples across three (3) different lots at different days (Implied from predicate and subject device consistency)
    Cutoff DeterminationAcceptance: Cutoff at 10 ng/ml. 50% below cutoff are negative. Cutoff and 25% above cutoff are positive. (Implied from predicate and subject device consistency)
    AccuracyAcceptance: Comparison study of positive and negative specimens with GC/MS shows very good correlation. (Implied from predicate and subject device consistency)
    Assay TimeAcceptance: 5 minutes (Explicitly stated in comparison table)
    MatrixAcceptance: Urine (Explicitly stated in comparison table)
    Preliminary Positive ReconfirmationAcceptance: Yes, by GC/Mass (Explicitly stated in comparison table)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective). It simply mentions "A comparison study of positive and negative specimens with GC/MS was performed" for accuracy, without providing specific numbers of specimens.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not mention the number of experts used or their qualifications for establishing ground truth. The ground truth method used (GC/MS) is an objective analytical method, not dependent on expert interpretation in the same way as, for example, image review by radiologists.

    4. Adjudication Method for the Test Set

    No adjudication method is mentioned. The ground truth is established by Gas Chromatography/Mass Spectrometry (GC/MS), which is a definitive analytical method and does not typically involve adjudication among human experts for preliminary results.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was mentioned. The device is a "visual read single use" diagnostic, where results are read directly, not through interpretation by multiple human readers that would then be compared to AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    The device itself is a standalone diagnostic tool, as it is a "visual read single use" lateral flow immunoassay. The intended use description explicitly states, "The tester will obtain a result in five minutes. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. This test is the first step in determining whether there are drugs in the urine." This indicates its standalone performance provides a preliminary result, with confirmatory testing (GC/MS or LC/MS) recommended for positive findings. The performance metrics (Sensitivity, Precision, Cutoff, Accuracy) are related to this standalone analytical performance.

    7. Type of Ground Truth Used

    The ground truth used for the accuracy study was Gas Chromatography/Mass Spectrometry (GC/MS). The text states: "A comparison study of positive and negative specimens with GC/MS was performed." It also mentions Liquid Chromatography/Mass Spectrometry (LC/MS) as another preferred confirmatory method.

    8. Sample Size for the Training Set

    The document does not provide information on a "training set" sample size. This device is a lateral flow immunoassay, which is a chemical and biological test, not an AI/machine learning algorithm that requires a training set in the conventional sense. Its development would involve chemical and biological optimization, not data-driven model training.

    9. How the Ground Truth for the Training Set Was Established

    As there is no "training set" in the context of an AI/machine learning algorithm, this question is not applicable. The development of this immunochromatographic assay relies on established biochemical principles and would involve iterative testing against known concentrations of Buprenorphine and its metabolites, similar to how the performance metrics (sensitivity, cutoff, accuracy) are assessed.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1