Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K121724

    Validate with FDA (Live)

    Date Cleared
    2012-09-20

    (100 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ENDOLIGHT LED Light Source Projector is intended to illuminate the surgical site during minimally invasive surgical procedures by producing light that is transmitted from the light source through fiber optic cable and a scope.

    Device Description

    The ENDOLIGHT LED Light Source is a fiber optic light source utilizing a single, solid state, high power, white light emitting diode (LED) to produce visible light that used to illuminate surgical sites during minimally invasive surgical procedures. The light from the ENDOLIGHT LED Light Source is transmitted through an optical cable and a scope. ENDOLIGHT LED 1.1 and 1.2 are comprised of light source and power supply cord. The units consist of aforementioned LED, cooling fan, light port, and protective housing. ENDOLIGHT LED 1.1 and 1.2 are portable and are very similar in design, format, and functionality to other LED light sources including the predicate devices. This product is exclusively intended for use by specialized medical personnel and must only be used by medically qualified and adequately trained persons.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the ENDOLIGHT LED Light Source. It focuses on demonstrating substantial equivalence to predicate devices rather than providing detailed acceptance criteria and a study that proves the device meets specific performance thresholds in a clinical or AI-assisted context.

    Therefore, many of the requested categories cannot be answered from the provided document. This document describes a medical device rather than an AI/ML powered device.

    Here's an analysis based on the information available:

    1. Table of Acceptance Criteria and the Reported Device Performance:

    The document does not present a table of specific quantitative acceptance criteria alongside reported performance data. Instead, it states that "Design verification testing demonstrates that the devices function as intended" and lists adherence to voluntary safety and performance standards.

    Acceptance Criteria CategorySpecific Criteria (from document)Reported Device Performance (from document)
    FunctionalityDevice functions as intended."Design verification testing demonstrates that the devices function as intended..."
    Safety - Electromagnetic CompatibilityConforms to IEC 60601-1-2 Edition 3:2007-03."The ENDOLIGHT LED 1.1 and 1.2 conform to the following Safety Standards: IEC 60601-1-2..."
    Safety - General RequirementsConforms to IEC 60601-1, Medical Electrical Equipment Part 1."The ENDOLIGHT LED 1.1 and 1.2 conform to the following Safety Standards: IEC 60601-1..."
    Safety - Endoscopic EquipmentConforms to IEC 60601-2-18:1996, Amendment 1 2000."The ENDOLIGHT LED 1.1 and 1.2 conform to the following Safety Standards: IEC 60601-2-18..."
    Safety & Effectiveness (General)No new issues of safety and effectiveness raised."...performance did not raise any new issues of safety and effectiveness..."
    User TrainingFormal user training not required."...formal user training is not required."

    2. Sample Size Used for the Test Set and the Data Provenance:

    Not applicable. The document describes a traditional medical device (LED light source), not an AI/ML-powered device that would typically involve test sets of data for performance evaluation. The "testing" mentioned refers to design verification and adherence to electrical safety standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

    Not applicable. Ground truth and expert consensus are relevant for AI/ML-powered diagnostic or interpretive devices, which this is not. The "ground truth" for this device would be its physical properties and adherence to engineering and safety standards.

    4. Adjudication Method for the Test Set:

    Not applicable. There is no mention of an adjudication method as it relates to evaluating data for an AI/ML-powered device.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance:

    Not applicable. This is not an AI-assisted device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    Not applicable. This is not an AI/ML algorithm. The device stands alone as a light source.

    7. The Type of Ground Truth Used:

    For this device, the "ground truth" for its performance is established by engineering specifications, international safety standards (e.g., IEC standards for medical electrical equipment), and functional design verification tests. The document explicitly states "Design verification testing demonstrates that the devices function as intended" and lists conformity to specific IEC safety standards.

    8. The Sample Size for the Training Set:

    Not applicable. This is not an AI/ML-powered device, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable.

    In summary: The provided document is for a traditional medical device (an LED light source) and outlines its substantial equivalence to predicate devices based on technological characteristics and adherence to established safety standards. It does not involve AI/ML technology, and therefore, many of the requested categories related to AI/ML performance evaluation (such as test sets, ground truth establishment by experts, adjudication, and clinical effectiveness studies with AI) are not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1