Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K121178

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-07-05

    (443 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Transcorp ACIF System is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine at one disc level. DDD is defined as neck pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies. Transcorp ACIF implants are used to facilitate fusion in the cervical spine and are placed via an anterior approach at the C3 to C7 disc levels using autogenous bone graft. Patients should have at least six weeks of non-operative treatment prior to treatment with an intervertebral body fusion device. The device must be used with supplemental fixation.

    Device Description

    The proposed modified Transcorp Anterior Cervical Intervertebral Fusion (ACIF) System includes various size implants manufactured from implant grade Solvay Zeniva ZA-500 PEEK conforming to ASTM F2026-08. The implant is hollow to allow for autogenous bone graft material. The implant is provided non-sterile.

    AI/ML Overview

    This 510(k) summary describes a medical device, the Transcorp ACIF System, and its modifications. It does not contain information about an AI/ML device or the results of a study designed to evaluate its performance against acceptance criteria using a test set, ground truth, or expert readers.

    The document describes the performance testing for a physical implantable device, not an algorithm. The relevant section states:

    "Performance testing was performed on the proposed modified IBFD. Static and dynamic axial compression, static and dynamic compression shear, and static torsion testing per ASTM F2077-11 were performed. Expulsion testing was also performed on the proposed modified IBFD. Engineering analyses were conducted to evaluate the Subsidence and Wear of the proposed modified implant. No clinical testing was performed."

    Therefore, I cannot provide the requested information about acceptance criteria and a study proving device performance in the context of an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1