Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K120964

    Validate with FDA (Live)

    Date Cleared
    2012-05-25

    (56 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Unstream GR Catheter is intended to be used in conjunction with a steerable guidewire to access discrete regions of the peripheral vasculature and for guidewire exchange.

    Device Description

    The Upstream GR Catheter is a sterile, single-use, dual-lumen catheter, having anchoring balloon and funnel at its distal tip, and Y-connector at its proximal end. The Upstream GR Catheter is intended for use with up to 0.035" guidewires. The Upstream GR Catheter is provided in 100 cm effective length and its shaft's outer diameter is 1.7mm (5F). A hub at the Y-connector center port, at the proximal end of the Upstream GR Catheter, allows guidewire access. The Y-connector side port is used for balloon inflation and deflation. The distal balloon is used for catheter anchoring and centering. The distal funnel is used as soft tip and soft stopper when an occlusion is reached.

    AI/ML Overview

    The provided text describes the Upstream Peripheral Technologies GR Catheter and its 510(k) summary. It details the device's intended use, technological characteristics, and the safety and performance data used to demonstrate substantial equivalence to a predicate device.

    However, the information provided does not pertain to an AI/ML device or a study involving human readers or ground truth established by experts. Instead, it describes a medical device (a catheter) and the engineering bench testing and biocompatibility assessments performed to support its marketing clearance.

    Therefore, many of the requested elements for describing acceptance criteria and a study for an AI/ML device, such as sample sizes for test sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, training set details, and ground truth establishment methods relevant to AI/ML, are not applicable or present in this document.

    Here's an attempt to answer the questions based only on the provided text, while highlighting where the requested information is not available due to the nature of the device and study described:

    Acceptance Criteria and Device Performance Study (for a non-AI medical device: Upstream GR Catheter)

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Test/RequirementAcceptance Criteria (Implicit)Reported Device Performance
    BiocompatibilityCytotoxicityBiocompatibleMeets requirements
    SensitizationBiocompatibleMeets requirements
    Irritation/Intracutaneous ReactivityBiocompatibleMeets requirements
    Systemic Toxicity (Acute)BiocompatibleMeets requirements
    Complement ActivationBiocompatibleMeets requirements
    Hemocompatibility (Hemolysis and Thromboresistance)BiocompatibleMeets requirements
    PyrogenicityBiocompatibleMeets requirements
    Bench Testing (ISO-10555-1 & 4, and Internal)Tensile Force TestingMeets specificationsMeets intended performance
    Air Leakage TestingMeets specificationsMeets intended performance
    Corrosion Resistance TestingMeets specificationsMeets intended performance
    Liquid Leakage Pressure TestingMeets specificationsMeets intended performance
    Catheter Torque TestingMeets specificationsMeets intended performance
    Kink TestingMeets specificationsMeets intended performance
    Tip TestMeets specificationsMeets intended performance
    Freedom from LeakageMeets specificationsMeets intended performance
    Balloon Burst TestingMeets specificationsMeets intended performance
    Surface and Dimensional AnalysisMeets specificationsMeets intended performance
    Radiopacity TestingMeets specificationsMeets intended performance
    Package TestingMeets specificationsMeets intended performance
    Sterilization Validation TestingMeets specificationsMeets intended performance

    Note: The document states "All of these tests demonstrated that the Upstream GR Catheter meets its intended performance specifications." The specific quantitative acceptance criteria or results for each test are not detailed in this summary.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not applicable/not provided for this type of medical device clearance. The "test set" in this context refers to physical units of the device subjected to various engineering and biocompatibility tests, not a dataset for an AI/ML algorithm. The provenance of the data would be the results from these lab tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable/not provided. The "ground truth" for this device is established by engineering specifications and recognized biocompatibility standards (e.g., ISO 10993, ISO 10555-1, ISO 10555-4), not by expert medical review of patient data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable/not provided. Adjudication methods like 2+1 or 3+1 refer to agreement among multiple expert readers, which is not relevant for the type of testing described (bench testing and biocompatibility).

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. This device is a physical medical instrument (a catheter), not an AI algorithm intended to assist human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No, a standalone performance evaluation in the context of an "algorithm" was not done. This is a physical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance is based on:

    • Compliance with international standards (e.g., ISO 10993 for biocompatibility, ISO 10555-1 and ISO 10555-4 for catheters).
    • Internal engineering and design specifications for mechanical and handling characteristics.
    • The performance of the predicate device (Endocross Ltd Enabler-P Support Catheter (K082339)) as a benchmark for substantial equivalence.

    8. The sample size for the training set

    This information is not applicable/not provided. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established

    This information is not applicable/not provided. As above, this is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1