    K Number
    K120690

    Validate with FDA (Live)

    Date Cleared
    2012-06-22

    (107 days)

    Product Code
    Regulation Number
    880.5570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Insulin pen needle is intended for use with pen injector for subcutaneous injection of insulin.

    Device Description

    Insulin pen needle is a single-use device, which consists of needle tube, needle shield, needle seat, needle seat protection cover, and seat seal paper. Needle gauge includes 29G, 30G, 31G, and 32G. Needle length can be produced to various sizes. After sterilization by EO, it's sterile, pyrogen free, and non-toxic. Its shelf life is 5-year.

    AI/ML Overview

    The provided text describes a 510(k) submission for an insulin pen needle. This type of medical device is generally evaluated through non-clinical testing to demonstrate substantial equivalence to predicate devices, rather than through complex clinical studies with human participants, AI algorithms, or extensive ground truth establishment. Therefore, many of the requested categories are not applicable in this context.

    Here's an analysis based on the provided document:

    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryAcceptance CriteriaReported Device Performance
    Mechanical/Physical Performance (ISO Compliance)Conforms to ISO 11608-2 (Needle systems for medical use - Requirements and test methods - Part 2: Pen needles)Bench tests prove compliance with ISO 11608-2
    Material Composition (Needle Tube)Made of SUS 304 stainless steelMade of SUS 304 stainless steel
    Material Composition (Needle Seat, Shield, Protection Cover)Made of PPMade of PP
    Material Composition (Needle Seat Seal)Made of Dialysis paperMade of Dialysis paper
    BiocompatibilityConforms to ISO 10993 (Biological evaluation of medical devices)Biocompatibility tests prove compliance with ISO 10993
    SterilitySterility Assurance Level (SAL): 1x10^-6SAL: 1x10^-6, EtO Sterilization
    PyrogenicityPyrogen freeReported as "pyrogen free"
    ToxicityNon-toxicReported as "non-toxic"
    Shelf Life5-year shelf lifeReported as "5-year shelf life"
    Labeling and Indications for UseMeet FDA requirements for labeling and intended useMeets FDA requirements, Intended for use with pen injector device for subcutaneous injection of insulin.

    Study Details (Based on available information)

    1. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not explicitly mentioned for specific test types, but implied to be sufficient for ISO compliance and biocompatibility testing. This typically involves a statistically relevant number of samples per test as outlined by the standards.
    • Data Provenance: The device manufacturer is Tiger Medical Products Ltd. in Shanghai, China. The testing was conducted by or on behalf of the manufacturer, likely in China or by certified labs. The tests are non-clinical (bench and biocompatibility).

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. This is a non-clinical device that relies on standardized performance and biocompatibility testing, not expert interpretation of diagnostic images or clinical outcomes. The "ground truth" is established by the specifications of the ISO standards and the verifiable results of the tests.

    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • None. Adjudication methods like 2+1 or 3+1 are used in clinical studies, particularly for diagnostic devices where human interpretation is involved. For this device, performance is measured against established engineering and biological standards.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is a single-use insulin pen needle, not a diagnostic or AI-powered device involving human readers.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This device does not incorporate any algorithms or AI component.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Standardized Test Results/Compliance with ISO Standards and Biocompatibility Protocols. The "ground truth" for this device lies in its ability to meet the defined technical specifications of ISO 11608-2, ISO 9626, ISO 7864, and the biological safety requirements of ISO 10993. This is verified through laboratory testing (e.g., force measurements, flow rates, sterility tests, cytotoxicity assays).

    7. The sample size for the training set

    • Not Applicable. There is no "training set" as this is not an AI/ML device.

    8. How the ground truth for the training set was established

    • Not Applicable. There is no "training set" or corresponding ground truth to establish.