Found 1 results

510(k) Data Aggregation

    K Number
    K120674

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-04-04

    (30 days)

    Product Code
    Regulation Number
    876.5540
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Mahurkar™ Elite Acute Dual Lumen Catheter is indicated for short-term central venous access for hemodialysis, apheresis, and infusion.

    The Mahurkar™ Elite Acute Triple Lumen Catheter is indicated for short-term central venous access for hemodialysis, apheresis, infusion, central venous pressure monitoring, and pressure injection of contrast media. The maximum recommended infusion rate is 5 ml/sec for power injection of contrast media.

    Device Description

    The Mahurkar™ Elite Acute Dual Lumen Catheter features a two lumen design. The proximal end has color-coded adapters to indicate arterial and venous flow. The adapters are connected to extension tubes which are available in curved or straight configurations. The extension tubes are connected to the hub which is joined to a dual lumen shaft available in pre-curved and straight configurations. The shaft extends to side slots near the distal tip. The dual lumen catheter is available in 12.0 Fr or 13.5 Fr outer diameters and a variety of implant lengths ranging from 13 cm to 30 cm. It is offered as a single device or as convenience kits.

    The Mahurkar™ Elite Acute Triple Lumen Catheter features a three lumen design. The proximal end has color-coded adapters to indicate arterial flow, venous flow, and medial infusion. The adapters are connected to extension tubes which are available in curved or straight configurations. The extension tubes are connected to the hub which is joined to a triple lumen shaft that extends to side slots near the distal tip. The triple lumen catheter is available in 12.5 Fr outer diameter and a variety of implant lengths ranging from 13 cm to 30 cm. It is offered as a single device or as convenience kits.

    AI/ML Overview

    This document describes a 510(k) submission for new catheter devices and not a study for AI/ML device. Therefore, I cannot extract the requested an AI/ML device study information from this document.

    However, I can provide available information about the device's performance testing.

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestAcceptance Criteria (Implied)Reported Device Performance
    Cleaning agent compatibilityMeet relevant product specificationsMeets relevant product specifications
    Lock solution compatibilityMeet relevant product specificationsMeets relevant product specifications
    Dynamic flowMeet relevant product specificationsMeets relevant product specifications
    Column strengthMeet relevant product specificationsMeets relevant product specifications
    Clamp functionalityMeet relevant product specificationsMeets relevant product specifications
    Shaft stiffnessMeet relevant product specificationsMeets relevant product specifications
    Tensile strength (various locations)Meet relevant product specificationsMeets relevant product specifications
    Resistance to kinkMeet relevant product specificationsMeets relevant product specifications
    Resistance to leakMeet relevant product specificationsMeets relevant product specifications
    Resistance to burstMeet relevant product specificationsMeets relevant product specifications
    Resistance to catheter collapseMeet relevant product specificationsMeets relevant product specifications
    Resistance to fatigueMeet relevant product specificationsMeets relevant product specifications
    Central venous pressure monitoringMeet relevant product specifications (for triple lumen catheter)Meets relevant product specifications (for triple lumen catheter)
    Simulated injection of contrast mediaMeet relevant product specifications (for triple lumen catheter)Meets relevant product specifications (for triple lumen catheter)
    Biocompatibility (ISO 10993)Biocompatible for intended useBiocompatible for its intended use

    2. Sample size used for the test set and the data provenance:

    • The document does not specify the sample sizes used for each of the bench-top functional and performance tests.
    • The data provenance is from bench-top functional and performance testing and biocompatibility testing, which are laboratory-based tests. There is no information about country of origin in the context of data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not applicable as the document describes physical device performance tests, not clinical studies involving expert interpretation.

    4. Adjudication method for the test set:

    • This information is not applicable for the type of testing described (bench-top and biocompatibility).

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This information is not applicable as this document describes a medical device (catheter) and not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • This information is not applicable as this document describes a medical device (catheter) and not an AI/ML device.

    7. The type of ground truth used:

    • The ground truth for the performance testing was based on "relevant product specifications". This implies pre-defined engineering and safety standards for catheter performance.
    • For biocompatibility, the ground truth was adherence to ISO 10993: Biological Evaluation of Medical Devices.

    8. The sample size for the training set:

    • This information is not applicable as this document describes a medical device (catheter) and not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established:

    • This information is not applicable as this document describes a medical device (catheter) and not an AI/ML device that requires a training set.