Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K120625

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-07-25

    (146 days)

    Product Code
    Regulation Number
    866.3305
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Roche Elecsys HSV-1 IgG immunoassay is a test for the in vitro qualitative determination of IgG class antibodies to HSV-1 in human serum and lithium-heparin plasma, K2—EDTA plasma, and K3—EDTA plasma. The test is intended for sexually active individuals and expectant mothers as an aid in the presumptive diagnosis of HSV-1 infection. The predictive value of positive and negative results depends on the population's prevalence and the pretest likelihood of HSV-1. The electrochemiluminescence immunoassay "ECLIA" is intended for use on Elecsys and cobas e immunoassay analyzers. The test is not FDA cleared for screening blood or plasma donors. The performance of this assay has not been established for use in a pediatric population, neonates, immunocompromised patients, or for use at point of care facilities.

    PreciControl HSV is used for quality control of the Elecsys HSV-1 IgG immunoassay on the Elecsys and cobas e immunoassay analyzers.

    Device Description

    (1) The Elecsys HSV-1 IgG immunoassay is a two-step sandwich immunoassay with streptavidin microparticles, biotinylated recombinant HSV-1-specific antigen labeled with a ruthenium complex and electrochemiluminescence detection. This assay is a qualitative test based on a cut-off formula dependent on the negative and positive calibrators. Cut-off index (COI) is based on the ratio of assay signal to cut-off signal (also abbreviated s/co). COI values greater than or equal to 1.0 are considered positive for the presence of anti-HSV-1 IgG antibody. Results are determined using a two-point calibration. The test system contains the human serum-based calibrators intended for use with the system. (2) PreciControl HSV contains lyophilized control serum based on human serum. The controls are used for monitoring the accuracy of the Elecsys HSV-1 IgG immunoassay.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study details for the Elecsys HSV-1 IgG Immunoassay, extracted from the provided 510(k) summary:


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the Elecsys HSV-1 IgG Immunoassay are demonstrated through its agreement with a Western Blot (WB) assay within different cohorts. The performance is reported as Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA).

    Performance MetricCohortAcceptance Criteria (Implicit, based on reported performance)Reported Device Performance (95% CI)
    Positive Percent Agreement (PPA)Expectant Mother CohortNot explicitly stated as a numerical criterion, but demonstrated by achieving the reported value.91.0% (82.4-96.3%)
    Sexually Active CohortNot explicitly stated as a numerical criterion, but demonstrated by achieving the reported value.94.2% (91.3-96.4%)
    Low Prevalence CohortNot explicitly stated as a numerical criterion, but demonstrated by achieving the reported value.94.9% (87.4-98.6%)
    Negative Percent Agreement (NPA)Expectant Mother CohortNot explicitly stated as a numerical criterion, but demonstrated by achieving the reported value.95.7% (85.5-99.5%)
    Sexually Active CohortNot explicitly stated as a numerical criterion, but demonstrated by achieving the reported value.90.3% (85.9-93.8%)
    Low Prevalence CohortNot explicitly stated as a numerical criterion, but demonstrated by achieving the reported value.96.7% (91.8-99.1%)
    Agreement with CDC Panel(General)100% agreement.100%

    Note: The document implicitly establishes the reported performance as meeting acceptance criteria by presenting it in the context of substantial equivalence.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Expectant Mother Cohort: n=125
    • Sexually Active Cohort: n=600
    • Low Prevalence Cohort: n=200
    • CDC Panel: Not explicitly stated, but 100% agreement indicates a comparison against a reference standard panel.

    Data Provenance: The document does not explicitly state the country of origin for the data or whether it was retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The ground truth for the device's performance comparison was established using a Western Blot (WB) assay as the comparator (predicate device: Focus HerpeSelect 1 and 2 Immunoblot IgG (K000238)). This is a laboratory test, not an interpretation by human experts.

    For the "Agreement with CDC Panel," the CDC panel itself represents a reference standard, implying established truth without direct mention of human experts in this section.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method involving multiple human readers for the test set. The direct comparison is between the candidate device (Elecsys HSV-1 IgG Immunoassay) and a predicate Western Blot assay.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The study compares the performance of the device against a predicate laboratory assay (Western Blot), not against human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the performance study performed appears to be a standalone assessment of the Elecsys HSV-1 IgG Immunoassay. The assay produces a quantitative result (Cut-off Index, COI) that is interpreted as reactive (≥ 1.0 COI) or non-reactive (< 1.0 COI). This is an automated algorithmic output, and its performance is compared directly to the Western Blot without human interpretation loops being described as part of the core performance evaluation study.

    7. The Type of Ground Truth Used

    The primary ground truth used for establishing the performance metrics (Positive Percent Agreement and Negative Percent Agreement) was the Focus HerpeSelect 1 and 2 Immunoblot IgG (K000238) Western Blot assay. This is a laboratory-based serological test, which serves as a reference standard for HSV-1 IgG antibody detection. Additionally, agreement with a CDC Panel was used as a ground truth comparison.

    8. The Sample Size for the Training Set

    The document does not provide information regarding the sample size of a training set for the Elecsys HSV-1 IgG Immunoassay. As this is a laboratory immunoassay and not an AI/machine learning algorithm in the typical sense of needing a distinct training set for model development, such information is generally not applicable or explicitly provided in this type of submission. The device's cut-off is based on a formula dependent on calibrators.

    9. How the Ground Truth for the Training Set Was Established

    Since no training set for an AI/machine learning model is described, the method for establishing its ground truth is not applicable or provided in the document. The assay's operational parameters, including cut-offs, are established through internal assay development and calibration procedures.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1