Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K120537

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-05-10

    (78 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Aquamantys3 9.5 XL Bipolar Sealer is a single use, sterile, bipolar device intended to be used in conjunction with the Aquamantys3 Pump Generator for delivery of radio-frequency (RF) energy and saline for blunt dissection and for haemostatic sealing and coagulation of soft tissue and bone at the operative site. It is intended for, but not limited to, open abdominal, orthopaedic, and thoracic surgery. The device is not intended for contraceptive tubal coagulation (permanent female sterilization).

    Device Description

    The Aquamantys3 9.5 XL Bipolar sealer represents a line extension to the predicate Aquamantys 9.5 XL Bipolar Sealer (cleared as the Double Cone Bipolar Sealer in K101057) and the Aquamantys product family. Similar to the predicate devices, it is a sterile, single-use hemostatic sealing device. The device employs bipolar radio-frequency (RF) energy and saline for blunt dissection and for hemostatic sealing and coagulation. The device is equipped with a dual electrode tip with saline apertures at its distal end. Saline and RF energy are supplied to the device from lines on the proximal end of the handpiece and connection to the Aquamantys3 Pump Generator via the distally attached cassette interface.

    AI/ML Overview

    This document describes the Aquamantys3 9.5 XL Bipolar Sealer, a medical device, and its acceptance criteria based on non-clinical testing. This 510(k) submission primarily focuses on demonstrating substantial equivalence to a predicate device, rather than a novel AI system. As such, many of the requested categories related to AI performance, ground truth, and expert evaluation are not applicable to this submission.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that "All testing met the acceptance criteria" for the listed tests. However, the specific quantitative acceptance criteria for each test are not provided in the summary. The reported performance is implicitly that the device successfully passed all tests.

    Test TypeAcceptance Criteria (Quantified)Reported Device Performance
    Static cable pullsNot specified in documentMet acceptance criteria
    Dynamic cord pullsNot specified in documentMet acceptance criteria
    Static saline tube pullsNot specified in documentMet acceptance criteria
    Air leak and flowNot specified in documentMet acceptance criteria
    Hipot testingNot specified in documentMet acceptance criteria
    Saline flow testingNot specified in documentMet acceptance criteria
    ContinuityNot specified in documentMet acceptance criteria
    Biocompatibility assessmentNot specified in documentMet acceptance criteria

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document does not specify the exact sample size for each bench performance test. It broadly states that the device "has undergone bench performance testing."
    • Data Provenance: The testing was "bench performance testing," implying it was conducted in a laboratory setting by Medtronic Advanced Energy. The country of origin of the data is not explicitly stated but would likely be the location of the submitter (Portsmouth, NH, USA). The data is prospective in the sense that these tests were performed on the device to assess its performance before market approval.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable as the evaluation was based on bench performance testing against device specifications, not expert interpretation of clinical data or imagery. Therefore, there was no "ground truth" in the context of expert consensus that needed to be established.

    4. Adjudication Method for the Test Set

    This is not applicable as the evaluation was based on bench performance testing against objective specifications, not subjective interpretation requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. This device is a surgical instrument (Electrosurgical Bipolar Sealer), not an AI-powered diagnostic or assistive tool for human readers. No MRMC study was performed.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. This device is a physical instrument, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's evaluation was based on engineering specifications and performance standards for electrosurgical devices. The bench tests measured objective physical and electrical properties of the device.

    8. The Sample Size for the Training Set

    This is not applicable. The device is not an AI model that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable. The device is not an AI model that requires a "training set."

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1