Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K120441

    Validate with FDA (Live)

    Date Cleared
    2012-05-31

    (108 days)

    Product Code
    Regulation Number
    880.6880
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Getinge 700HC-E Series Steam Sterilizer is intended for use by health care facilities to sterilize wrapped and unwrapped, porous and nonporous heat and moisture stable items such as surgical instruments and linens by means of pressurized steam. The 700HC-E Series Steam Sterilizer is available in 3 models differentiated by chamber length; 733HC-E 39 inch chamber, 733HC-E 53 inch chamber and 733HC-E 61 inch chamber.

    Device Description

    The Getinge 700HC-E Series Steam Sterilizer is designed for sterilization of heat and moisture stable materials used in healthcare facilities. The only model designation provided in the 700HC-E Series Steam Sterilizer is 733HC-E. The model 733HC-E is available in 3 chamber sizes; 39 inches long (21.5 cu ft), 53 inches long (29.3 cu ft) and 61 inches long 33.7 cu ft).

    The 733HC-E Steam Sterilizer employs both gravity/downward displacement with positive pulse conditioning and pressure/vacuum pulsing for dynamic air removal. Up to 22 cycles can be easily accessed and custom cycle names can be designated by the user (duplicate cycles are provided to allow for user designated names). All cycle phases are sequenced and monitored by the control system, providing both audible and visual notification of deviation from certain operating parameters.

    AI/ML Overview

    The acceptance criteria for the Getinge 700HC-E Series Steam Sterilizer are based on the ability to achieve sterilization of various loads under specified cycle parameters, and dryness of the load where applicable. The study proving the device meets these criteria involved comprehensive performance testing conforming to industry standards.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the Getinge 700HC-E Series Steam Sterilizer are implicitly defined by the successful demonstration of:

    • Achieving a sterility assurance level (SAL) of at least 10^-6.
    • Complete kill of biological indicators (BIs).
    • Moisture retention within specified limits for fabric and wrapped instrument packs.
    • Successful Bowie-Dick test for air removal.
    • Maintaining steady-state thermal conditions within the chamber.
    • Water loss not exceeding 50ml for liquid loads.

    The reported device performance, demonstrated through validation testing against these criteria, confirmed that the sterilizer performs as intended.

    Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
    Sterility AssuranceSterility Assurance Level (SAL) of at least 10^-6Achieved SAL of at least 10^-6 for all validated cycles.
    Microbiological EfficacyComplete kill of biological indicators (BIs)Complete BI kill demonstrated across all validated cycles.
    Temperature ProfileSteady-state thermal conditions consistent with predicated SAL in the loadDemonstrated by empty chamber testing.
    Fabric PacksMoisture retention < 3% increase in pre-sterilization test pack weight, no visible wet spotsAchieved for PREVAC and GRAVITY cycles using fabric packs.
    Wrapped Instrument PacksMoisture retention < 20% increase in pre-sterilization weight of towel, no visible wet spots on outer wrapperAchieved for PREVAC (excluding PREVAC 2 & 5 with shortened drying times) and GRAVITY cycles using wrapped instrument packs.
    Immediate Use (Flash) CyComplete BI kill. (Note: Drying not a primary criterion for these short cycles)Complete BI kill demonstrated.
    Liquid LoadsWater loss not exceeding 50mlAchieved for liquid cycles.
    Air Removal (Bowie-Dick)Successful Bowie-Dick testPerformed successfully, demonstrating proper air removal.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state a numerical "sample size" in terms of number of sterilizer units tested or specific quantities of test cycles beyond stating "all cycles were validated." However, the testing was performed for each cycle type available on the Getinge 700HC-E Series Steam Sterilizer, for all three chamber sizes (39 inch, 53 inch, 61 inch), and for various load configurations as detailed in the "Cycles and Load Chart."

    • Test Sets: The "test set" consisted of various "process challenge devices" and load configurations specified by ANSI/AAMI ST8:2008, including:
      • Fabric process challenge packs
      • Wrapped instrument process challenge devices
      • Unwrapped non-porous process challenge devices
      • 3 one-liter flasks for liquid load cycles
      • Bowie-Dick test packs
      • Empty chambers for thermal mapping.
    • Data Provenance: The testing was performed according to "industry standards and guidelines," specifically referencing ANSI/AAMI ST8:2008, which is a standard applicable within the United States. The testing appears to be prospective as it validates the performance of the new device model.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not specify the number or qualifications of experts used to establish the "ground truth" for the test set. Instead, the ground truth is established by objective, standardized methods and criteria defined by the ANSI/AAMI ST8:2008 standard and the expected biological and physical outcomes of sterilization. For example, "complete kill of biological indicators" is an objective, measurable outcome.

    4. Adjudication Method for the Test Set

    Not applicable. The "ground truth" for the performance testing is determined by objective measurements (e.g., direct observation of BI growth, temperature recordings, weight measurements) against pre-defined thresholds from the ANSI/AAMI ST8:2008 standard, rather than subjective expert adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device (steam sterilizer), not an AI/imaging diagnostic device. Therefore, MRMC studies and concepts of human reader improvement with AI assistance are irrelevant to this submission.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is a steam sterilizer, not an algorithm. However, the performance testing described is analogous to a "standalone" performance evaluation of the device itself, as it assesses the sterilizer's function without human intervention beyond operation and monitoring.

    7. The Type of Ground Truth Used

    The ground truth used for performance validation is based on:

    • Biological Indicators (BIs): Used to confirm the destruction of resistant microorganisms, providing a direct measure of microbial inactivation (sterility assurance).
    • Physical Measurements:
      • Temperature and Time: Achievement of specific temperature and exposure time combinations (expressed as F0 value) within the chamber and load.
      • Weight Measurement: To assess moisture retention in fabric and wrapped instrument packs.
      • Volume Measurement: To assess water loss in liquid loads.
    • Chemical Indicators (Bowie-Dick Test): Used to detect air removal efficacy in dynamic-air-removal steam sterilizers.
    • Industry Standards: Adherence to the criteria and methodologies outlined in ANSI/AAMI ST8:2008 for "Hospital Steam Sterilizers."

    8. The Sample Size for the Training Set

    Not applicable. This device is a classic medical device (steam sterilizer) and does not involve machine learning or AI, and therefore does not have a "training set." The listed cycles and load configurations were extensively tested as part of the validation, representing the complete range of intended uses.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as no training set was used.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1