Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K113568

    Validate with FDA (Live)

    Device Name
    MTA-FILLAPEX
    Date Cleared
    2012-08-21

    (263 days)

    Product Code
    Regulation Number
    872.3820
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MTA Fillapex is a root canal sealer intended for the permanent sealing of root canals and may be used in combination with root canal obturation materials.

    Device Description

    MTA Fillapex is a mineral trioxide aggregate (MTA) and resin root canal sealer used during endodontic treatment to permanently fill the canal system following debridement and disinfection. It consists of two component pastes that are combined in a dual barrel syringe for ease of dispensing and consistent dosage. Being hydrophilic in nature, MTA Fillapex is desirable as a root filling material because an isolated dry field is not necessary for use. Moisture does not negatively affect the sealing ability and is required for proper setting. It is used in combination with gutta-percha or silver points during root canal obturation. The device is provided non-sterile.

    AI/ML Overview

    This document describes the Angelus Indústria de Produtos Odontológicos S/A MTA Fillapex, a root canal sealer. The submission is a 510(k) premarket notification (K113568) and focuses on demonstrating substantial equivalence to previously marketed devices.

    Here's an analysis of the provided information regarding acceptance criteria and the supporting study:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Test StandardAcceptance Criteria (from ISO 6876)Reported Device Performance (MTA Fillapex)Source in Document
    BiocompatibilityISO 10993-5 (Cytotoxicity)Not explicitly stated in approval. Implied: demonstrate non-cytotoxicity or equivalence to predicates.Testing performed demonstrated substantial equivalence to predicate devices (side-by-side cytotoxicity).{1} Non-Clinical Testing
    Physical/Chemical PropertiesISO 6876 (Dental root canal sealing materials){1} Non-Clinical Testing
    FlowISO 6876(Specific value not provided in document)Testing performed; results were reported.{1} Non-Clinical Testing
    Working TimeISO 6876(Specific value not provided in document)Testing performed; results were reported.{1} Non-Clinical Testing
    Setting TimeISO 6876(Specific value not provided in document)Testing performed; results were reported.{1} Non-Clinical Testing
    Film ThicknessISO 6876(Specific value not provided in document)Testing performed; results were reported.{1} Non-Clinical Testing
    Dimensional ChangeISO 6876(Specific value not provided in document)Testing performed; results were reported.{1} Non-Clinical Testing
    SolubilityISO 6876(Specific value not provided in document)Testing performed; results were reported.{1} Non-Clinical Testing
    RadiopacityISO 6876(Specific value not provided in document)Testing performed; results were reported.{1} Non-Clinical Testing

    Important Note: The provided document states that "Performance testing that conformed to the protocols and recommended values described in ISO 6876 Dental root canal sealing materials was performed." However, it does not explicitly list the specific acceptance values (e.g., "flow > X mm") for each parameter from ISO 6876, nor does it provide the exact numerical results for MTA Fillapex for each test. It simply states that "Test reports included [list of tests]."

    2. Sample Size Used for the Test Set and the Data Provenance

    This submission is a 510(k) for a medical device and relies on non-clinical performance testing and biocompatibility data, not a clinical study with a "test set" of patient data.

    • Sample Size for Non-Clinical Testing: The document does not specify the exact number of samples (e.g., number of specimens for flow testing) used for each non-clinical test. It broadly states that "side-by-side cytotoxicity testing" and "performance testing" were performed.
    • Data Provenance: The testing was performed by or for the manufacturer, Angelus Indústria de Produtos Odontológicos S/A, based in Brazil. The data is thus considered internal manufacturer data from controlled laboratory settings. It is not retrospective or prospective patient data from a clinical setting.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Not applicable. This submission does not involve human expert interpretation of clinical data or images to establish a "ground truth." The "ground truth" for non-clinical performance tests is determined by the standardized methods and measurements defined in ISO 6876.

    4. Adjudication Method for the Test Set

    Not applicable. There is no "test set" in the context of clinical data requiring expert adjudication. The non-clinical tests follow direct measurement and evaluation against pre-defined ISO standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This is a traditional medical device (root canal sealer), not an AI/Software as a Medical Device (SaMD). Therefore, no MRMC study, AI assistance, or effect size related to human reader improvement is relevant or discussed.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This is a physical medical device, not an algorithm or AI.

    7. The Type of Ground Truth Used

    For the non-clinical performance tests (flow, working time, setting time, etc.), the "ground truth" is defined by the protocols and recommended values described in the international standard ISO 6876 Dental root canal sealing materials. For biocompatibility, the ground truth is established by the methods outlined in ISO 10993-5 Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity.

    8. The Sample Size for the Training Set

    Not applicable. There is no concept of a "training set" for physical medical devices of this type. Performance is demonstrated through direct testing against established material science standards.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set."

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1