510(k) Data Aggregation

    K Number
    K113523

    Validate with FDA (Live)

    Date Cleared
    2011-12-20

    (21 days)

    Product Code
    Regulation Number
    888.3660
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SMR Reverse Shoulder System is indicated for primary, fracture or revision total shoulder replacement in a grossly rotator cuff deficient joint with severe arthropathy.

    The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device.

    The SMR Reverse Shoulder System humeral bodies, cemented non-finned humeral stems and cemented revision stems are intended for cemented use. The SMR Reverse Shoulder System humeral bodies, finned humeral stems and uncemented revision stems are intended for cementless use. The SMR Reverse Shoulder System metal-backed glenoids and glenospheres are intended for uncemented press-fit use only with the addition of screws for fixation.

    Device Description

    The SMR Reverse Shoulder System (cleared via 510(k): K110598) consists of a humeral stem, a reverse humeral body, a reverse liner, a metal-back glenosphere and a connector with screw. Bone screws are used for the fixation of the metal-back glenoid to the scapula. Humeral stems and reverse humeral bodies are provided for both cementless fixation. The SMR Reverse Shoulder System metal back glenoids and glenospheres are intended for uncemented press-fit use only with the addition of screws for fixation.

    This submission is to add the SMR Revision Stems, cleared in K111212, as compatible humeral stem components of the SMR Reverse Shoulder System.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device called the SMR Reverse Shoulder System. It is concerned with demonstrating substantial equivalence to a predicate device, rather than presenting a study of device performance against specific acceptance criteria. Therefore, most of the information requested in your prompt is not available in this document.

    Here's what can be extracted and what cannot:

    1. Table of acceptance criteria and reported device performance:
    Not applicable. This document is not a clinical study or a performance validation report against specific acceptance criteria. Its purpose is to demonstrate "substantial equivalence" of a modified device to a previously cleared predicate device.

    2. Sample size used for the test set and the data provenance:
    Not applicable. No "test set" in the context of device performance testing is mentioned. The submission relies on a "print review" and comparison of taper dimensions.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    Not applicable. No ground truth establishment by experts is described for performance testing.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
    Not applicable. No adjudication method is described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    Not applicable. This device is a shoulder implant, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
    Not applicable. For the purpose of this 510(k) submission, the "ground truth" equivalent would be the existing characteristics and performance of the predicate device, against which the modified device is compared to establish equivalence.

    8. The sample size for the training set:
    Not applicable. No training set is mentioned as this device is not an AI/ML algorithm.

    9. How the ground truth for the training set was established:
    Not applicable. No training set is mentioned.


    Summary of available information related to "acceptance criteria" and "study":

    The "acceptance criteria" for this 510(k) submission are based on demonstrating substantial equivalence to a predicate device.

    • Predicate Device: Limacorporate SMR Reverse Shoulder System (K110598)
    • "Study" conducted: A non-clinical "print review" was performed.
      • This review compared the taper dimensions and tolerances of the SMR Revision Stems (the modification) to the taper dimensions and tolerances of the SMR humeral stems cleared for use with the predicate device.
      • Conclusion: The comparison indicated that the SMR Revision Stem tapers were substantially equivalent to the SMR Humeral Stem tapers.
    • Clinical Testing: The document explicitly states: "Clinical testing was not necessary to demonstrate substantial equivalence of the SMR Reverse Shoulder System with modified labeling to the predicate device."

    In essence, for this specific 510(k) submission, the "acceptance criteria" were met by a successful non-clinical engineering comparison demonstrating that the modification (addition of SMR Revision Stems) maintained the equivalence in design and dimensions to previously cleared components within the existing system.