Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K113338

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-08-10

    (270 days)

    Product Code
    Regulation Number
    868.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Oricare C4500 Medical Air Compressor is indicated to supply dry filtered and compressed air to a Medical Ventilator that operates within the C4500 Compressor manufacturer specifications.

    Device Description

    The Oricare C4500 Medical Air Compressor operates from an AC Voltage source and produces air from the normal environment supply compressed air for medical

    AI/ML Overview

    The Oricare C4500 Medical Air Compressor is a medical device designed to supply dry, filtered, and compressed air to medical ventilators. The submission provides a "Substantial Equivalence Summary" rather than a detailed study report with acceptance criteria and direct performance metrics. However, it implicitly demonstrates that the device meets acceptance criteria by establishing substantial equivalence to a legally marketed predicate device, the EKOM DK50 D Medical Compressor (K060781). This is achieved by comparing various operational and physical characteristics of the Oricare C4500 against those of the predicate device.

    Here's an analysis based on the provided text, outlining the acceptance criteria (inferred from predicate device characteristics) and the reported performance of the Oricare C4500:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are inferred from the specifications of the predicate device, the EKOM DK50 D Medical Compressor (K060781). The reported device performance is that of the Oricare C4500 Air Compressor, demonstrating that it either matches or has minor, non-adverse differences compared to the predicate.

    #Item of ComparisonAcceptance Criteria (Predicate: EKOM DK50 D)Reported Device Performance (Oricare C4500)Discussion of Differences
    1Intended useSupplying compressed air for medical ventilators.Supply gas for critical care ventilator.No differences
    2Indications for useClean, oil-free pressurized air for medical ventilators.Dry filtered and compressed air to a Medical Ventilator.No differences
    3Environment of UseMedical Care FacilitiesMedical Care FacilitiesNo Differences
    4Principle of operationCompression, cooling, drying, filtering of oil-free air.Compression, cooling, drying, filtering of oil-free air.No Differences
    5Output flow/pressure40 l/min at 51 psig40 l/min at 50 psigNo Differences
    6Peak Flow200 lpm for 2 sec180 Lpm for 0.6 secNo Differences (The slight difference in peak flow duration is considered non-significant for substantial equivalence).
    7Power120V/60 Hz115V/60 HzAC Mains line voltage varies from 103VAC to 127VAC. Both systems are within the range.
    8Power consumption5.6 ANominal current (amps) <600VAOricare device consumes slightly less power, meaning it is slightly more efficient.
    10Air filtration5 micron5 micronNo Difference
    11Pressure dew point@ 40 Lpm 20°C - 5°C below ambient temperature@ 40 Lpm 3°C below ambient temperatureNo real discernible difference.
    12Outlet connectionDISS1- DISSNo Difference
    13Sound level< 51 dB(A)< 52 dB(A)No real discernible difference. An increase in the dB(A) provides an advantage to the user for alarm annunciation.
    14Mode of operationContinuous - SIContinuous - SINo Difference
    15Separation of condensed waterAutomaticAutomaticNo Difference
    16Operating pressure of safety valve116 psig65 psigDevice regulates pressure at 40 psig. The lower safety valve pressure is still well above the operating pressure.
    17Adjustment of pressurePressure regulatorPressure regulatorNo Difference
    18Alarm for cooling failure / high tempAcoustic and optical if increase in internal temp > 80°CAcoustic and optical if increase in internal temp > 80°CNo Difference
    19Automatic turn-on pressureWhen central distribution pressure < 40.6 psigWhen central distribution pressure < 40.6 psigNo Difference
    20Output pressurePressure gaugePressure gaugeNo Difference
    21Alarm - loss of powerNo, but instructions require connection to equipment with this alarmYesAdvantage for Oricare product safety.
    22Additional Visual IndicatorsNot included"AC Power" light, "Ready" light, "Air Source - Wall" lightSafety Advantage for determining AC Mains Power is connected.
    23Indication of dryingPressure gaugeNoneDesign includes chiller and external water trap to remove excess moisture, thus an explicit drying indication is not deemed necessary.
    24Alarm for low pressureNone, instructions require connection to equipment with this alarmYes, triggers when output pressure <25.4 psigOricare device does not rely on 3rd party device to be connected, providing an advantage.
    25Material in gas pathwayAluminum, brass, nickel plated brass, polyurethane and silicon tubing, copper tubing, Plastic (PA, PET, POM, PBT, PC, Acetyl, Polyester), polyurethane foam, NBR rubber, Stainless steel, die cast zincAluminum, brass, nickel plated brass, polyurethane and silicon tubing, copper tubing, Plastic (PA, PET, POM, PBT, PC, Acetyl, Polyester), polyurethane foam, NBR rubber, Stainless steel, die cast zincNo Difference
    26Air tank capacity5L2LExcess capacity not needed to supply sufficient peak flow. Predicate device is larger because extra capacity requires additional weight and physical size, making the Oricare's smaller capacity an advantage.
    27Compressor TypeOil free piston and ring- positiveOil free wobble type rocker piston- positiveNo Difference
    28Type of lubricationOil-lessOil-lessNo Difference
    29Ambient Environment41 to 104 degrees F; up to 95% RH50 to 104 degrees F; 15 – 90% RHOricare device range is typical of many ventilators. The narrower humidity range is not considered an adverse difference.
    30Dimensions19(L)x20(W)x33(H) in16.22 (L)x17.4(W)x15.75(H) inOricare device is smaller. This is an advantage.
    31Weight101 lbs66 lbsOricare device is lighter. This is an advantage.
    32Electrical SafetyEN 60601-1IEC 60601-1:1988, +A1:1991, +A2:1995No difference, both comply (IEC 60601-1 is the international standard from which EN 60601-1 is derived).
    33Mechanical SafetyEN 60601-1IEC 60601-1:1988, +A1:1991, +A2:1995No difference, both comply.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document describes a "Substantial Equivalence Summary" based on a comparison to a predicate device and compliance with recognized standards. This is not a study in the typical sense of testing a specific sample size of the device to collect performance data from the field. Instead, it's a design and engineering comparison.

    • Test Set Sample Size: The document does not specify a "test set" in the context of a clinical trial or a large-scale performance study with a numerical sample size. The testing appears to be functional and safety testing as part of design verification, likely performed on a limited number of engineering or production units.
    • Data Provenance: The data primarily comes from device testing described as having been done "in accordance with various recognized standards, including ISO 10993-1:1999, AAMI BI79:1995 and IEC 60601-2-2:2007." Additionally, Air Purity Testing was conducted. This indicates the testing was likely done in a prospective manner during the device development and verification phase by Oricare, Inc. or its designated testing facilities, presumably in the USA (where Oricare, Inc. is based and the 510(k) was submitted). The "comparison to predicate device" serves as the primary "data" for substantial equivalence.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This type of submission (510(k) for a medical air compressor) typically does not involve a "ground truth" derived from expert consensus in the same way a diagnostic AI algorithm would. The "ground truth" for a device like this would be established through engineering specifications, regulatory standards, and objective physical measurements.

    • Number of Experts: Not applicable in the context of expert human review for ground truth
    • Qualifications of Experts: N/A for this type of device submission

    4. Adjudication Method for the Test Set

    Again, this is not applicable for a device of this nature where objective engineering and performance specifications are the "truth." There is no mention of "adjudication" in the context of resolving differing expert opinions.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices (especially those using AI) where human readers (e.g., radiologists) interpret cases with and without algorithmic assistance. The Oricare C4500 is an air compressor, not a diagnostic device, and does not involve human interpretation or AI assistance in a diagnostic capacity.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Not applicable. The device is an air compressor, not an algorithm, and therefore doesn't have "standalone" algorithm performance to be assessed. Its performance is measured directly by its physical output and adherence to specifications.

    7. The Type of Ground Truth Used

    The "ground truth" for this device is based on:

    • Objective Engineering Specifications: The technical requirements and design parameters of the device (e.g., flow rate, pressure, power consumption, filtration).
    • Compliance with Recognized Standards: Verification that the device meets relevant national and international medical device performance and safety standards (e.g., ISO 10993-1, AAMI BI79, IEC 60601-1).
    • Bench Testing and Performance Data: Direct measurements of the device's output and characteristics (e.g., air purity through air testing, sound level, temperature performance).
    • Predicate Device Specifications: The established performance and safety characteristics of the legally marketed EKOM DK50 D Medical Compressor, which serves as the benchmark for substantial equivalence.

    8. The Sample Size for the Training Set

    • Not applicable. This device is an air compressor and does not involve AI or machine learning that requires a "training set" of data.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. Since there is no training set, there is no ground truth established for one.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1