Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K113105

    Validate with FDA (Live)

    Date Cleared
    2011-12-15

    (56 days)

    Product Code
    Regulation Number
    870.2300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Tempus IC is intended to aid with the diagnosis of a person presenting as unwell or sick when they are in a location remote from immediate medical assistance. The device allows the User to take vital signs data from a patient and to transmit that data to medical professionals located at the response centre elsewhere. Typical examples are remote land, sea or air locations.

    Tempus IC is intended primarily to be used by medically unqualified people who have received basic training in the use of the device. Medical expertise is provided through communication with the Response Centre which would be staffed by physicians who would advise the operator on the nature of the medical incident.

    Tempus IC is intended to be used where a physician or other medically trained staff may or may not be present but where remote physician support is required.

    Tempus IC measures non-invasive blood pressure, SpO2, pulse rate, respiration rate and ETCO2, 12 Lead ECG, tympanic temperature (via a wireless external module) and blood glucose (via a wireless external module).

    Tempus IC is suitable for use on adults or children (over 10 years old and over 20kg in weight).

    Device Description

    The TempusIC™ is an advanced multi-parameter vital signs monitor designed for use in remote locations by trained non-expert users. It provides a wide range of features in a highly robust package, including integrated voice link and video camera.

    The device remains unchanged from its previous/existing clearance with the exception that the device is now capable of receiving (wirelessly over a Bluetooth™ link) both tympanic temperature readings from a tympanic thermometer (cleared under K101264) and blood glucose data from the Entra Health Systems MyGlucoHealth glucometer (cleared under K081703),

    Voice and data connections are made automatically via existing satellite or terrestrial communications systems.

    The TempusIC™ is used in conjunction with TempusNET™ software, which provides a system for receiving real-time voice and vital signs data. The system enables users to receive voice, vital signs data, and still video pictures from TempusICTM devices located anywhere in the world.

    The TempusNET™ system can be used by commercial response centre service providers or by individuals or organisations wishing to provide their own internal service.

    It also provides a synchronised user interface, and remote control of the TempusIC™.

    TempusNET™ also supports a full patient records database.

    AI/ML Overview

    The provided 510(k) summary for the RDT Tempus IC Patient Monitor indicates that the device's clearance is based on its substantial equivalence to a legally marketed predicate device (itself, prior to the addition of wireless peripherals). The study details focus on the newly added functionalities (wireless temperature and blood glucose measurement) and their reliable and accurate transmission.

    Here's an analysis based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly present a table of "acceptance criteria" with specific numerical targets for performance metrics (e.g., accuracy, precision) for each vital sign parameter. Instead, the "acceptance criteria" are implied by the demonstration of equivalence to predicate devices and the reliability and accuracy of data transmission for the new wireless peripherals.

    Parameter/Test AreaAcceptance Criteria (Implied)Reported Device Performance
    Overall Device EquivalencePerformance equivalent to legally marketed predicate device(s) (Tempus IC cleared under K082718)"the device is safe, effective and performs as well as, or better than, the legally marketed predicate device(s)."
    MyGlucoHealth Data ReceptionMyGlucoHealth data received reliably and accurately."Comparative testing... comprised benchmarking to confirm the MyGlucoHealth data was received reliably and accurately."
    Glucometer Data Transmission (Bench Testing)Transmitted data is transmitted reliably and accurately."The glucometer has been bench tested to confirm that transmitted data is transmitted reliably and accurately."
    Glucometer Wireless RangeOperates reliably and accurately at its maximum stated range."The glucometer has been tested to confirm it operates reliably and accurately at its maximum stated range."
    Glucometer Wireless Co-existenceOperates reliably in the presence of other wireless fields."The glucometer has been tested to confirm it operates reliably in the presence of other wireless fields as per the FDA Guidance for Radio-Frequency Wireless Technology in Medical Devices."
    Tympanic Thermometer FunctionalityEquivalent to the predicate (same thermometer cleared with Tempus IC Professional under K101264)"The use of the tympanic thermometer is predicated on the clearance of the same thermometer with the Tempus IC Professional (granted under K101264)." (Implies performance is already established)
    Software ConformanceConforms to FDA guidance for software in premarket submissions, IEC62366, and IEC60601-1-4."The requirements of the FDA document Guidance for the Content of Premarket Submissions for Software in Pre-Market Submissions has been applied. In addition, the requirements of IEC62366 and IEC60601-1-4 have been addressed."
    UsabilityAddresses usability as per IEC62366."Usability has been address by the application of IEC62366."

    2. Sample Size Used for the Test Set and Data Provenance:

    The document primarily refers to "testing" without providing specific sample sizes for individuals or data points for the comparative or bench tests.

    • Sample Size: Not explicitly stated. The description mentions "comparative testing," "bench testing," and "wireless range/co-existence testing," but details on the number of measurements, subjects, or data points are absent.
    • Data Provenance: Not explicitly stated. The tests are described as being performed to "benchmark" the device and confirm reliable/accurate data transmission. It's likely these were internal laboratory or engineering tests, but no geographic or retrospective/prospective details are given.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not provided in the document. The testing described primarily focuses on the technical performance of the wireless data transmission features of the new peripherals, rather than diagnostic accuracy established by medical experts on a "test set" of patients.

    4. Adjudication Method for the Test Set:

    This information is not provided. Given the nature of the testing described (benchmarking data transmission, wireless performance), an adjudication method as typically applied to clinical diagnostic studies is not relevant or described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, an MRMC comparative effectiveness study is not described. The document focuses on the device's performance and its equivalence to predicates, not on how human readers' effectiveness might improve with or without AI assistance. This device is a vital signs monitor, not an AI-powered diagnostic system.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    The document describes testing of the device's components, particularly the wireless glucometer and its data transmission. This would be analogous to a standalone performance evaluation of those specific functionalities (e.g., how accurately and reliably the glucometer transmits data). However, it's not an "algorithm-only" study in the context of an AI-powered diagnostic algorithm. It's a hardware/software system performance evaluation.

    The "bench testing" and "wireless range/co-existence testing" for the glucometer can be considered standalone performance evaluations of those specific functionalities.

    7. Type of Ground Truth Used:

    For the glucometer functionality, the ground truth would likely be established by a reference method for blood glucose measurement to determine the "accuracy" of the data received. For wireless reliability, the "ground truth" is simply whether the data was successfully transmitted and received without error, and if it matched the source data.

    • For Glucometer Accuracy (implied): A reference blood glucose measurement. The document states "benchmarking to confirm the MyGlucoHealth data was received reliably and accurately," implying a comparison to a known standard.
    • For Wireless Reliability: Successful transmission and reception of data without corruption, verified against the original transmitted data.

    8. Sample Size for the Training Set:

    This information is not applicable and not provided. The Tempus IC Patient Monitor is a vital signs monitoring device, not a machine learning or AI system that requires a "training set." Its functionality relies on established physiological measurement technologies and data transmission protocols.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable. As stated above, there is no "training set" for this device in the context of machine learning. The device's operation is based on physical sensors and software logic, not learned patterns from a training dataset.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1