510(k) Data Aggregation

    K Number
    K112998

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-03-15

    (160 days)

    Product Code
    Regulation Number
    870.5200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AutoPulse Resuscitation System Model 100 is intended to be used as an adjunct to manual CPR, on adult patients only, in cases of clinical death as defined by lack of spontaneous breathing and pulse.

    Device Description

    The AutoPulse Resuscitation System Model 100 is an automated, portable, batterv powered device that compresses the chest of an adult human as an adjunct to manual CPR. The AutoPulse System consists of 4 primary components, a reusable Platform, a single use chest compression assembly (LifeBand), a rechargeable battery, and a battery charger. The AutoPulse Platform contains the mechanical drive mechanism, control system, software and electronics necessary to generate and control the force required to perform mechanical chest compressions. User controls and indicators are contained in the User Control Panel. The AutoPulse Platform and LifeBand are unchanged. A new battery and battery charger are the subject devices of this 510(k).

    AI/ML Overview

    The provided documentation is a 510(k) summary for the AutoPulse® Resuscitation System Model 100, focusing on changes related to its battery and charger. It is not a study reporting on the clinical performance of the device in relation to patient outcomes or a diagnostic algorithm's accuracy.

    Therefore, many of the requested fields are not applicable to this specific document as it details a submission for a modification to an already cleared device, primarily concerning electrical components, rather than a clinical effectiveness study.

    Based on the provided text, here is the information that can be extracted:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Bench Testing)Reported Device Performance
    System level compatibility with new Li-Ion battery meets operating parameters in a safe and intended manner.Data showed that the new battery chemistry did not alter the functioning of the AutoPulse in any way and that it operates identically to the NiMH Battery powered AutoPulse.
    New multi-chemistry charger capable of safely and consistently charging NiMH battery, testing the battery, and correctly identifying end of life.Bench testing showed the new charger was capable of safely and consistently charging the NiMH battery, testing the battery, and correctly identifying when the end of life has been reached.
    New multi-chemistry charger capable of safely and consistently charging, testing, maintaining, and identifying end of life conditions per Li-Ion battery specifications.The new charger successfully demonstrated its ability to charge, test, maintain, and identify the end of life conditions per the Li-Ion battery specifications.
    Li-Ion battery meets performance safety specifications (including charging/discharging characteristics, ability to power AutoPulse for specified runtime throughout life, environmental testing, electrical safety, electromagnetic compatibility).Extensive bench testing was conducted to verify ability of the Li-Ion battery to meet the performance safety specifications. (Specific quantitative results not provided in this summary, but the general statement indicates criteria were met).

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not specified in terms of clinical patient data. The testing was "extensive bench testing" on the device components (new battery and charger).
    • Data Provenance: Not applicable as this was bench testing, not human patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of Experts: Not applicable. Ground truth for electrical and mechanical performance is established through engineering specifications and standards, not expert clinical consensus.
    • Qualifications of Experts: Not applicable.

    4. Adjudication method for the test set

    • Adjudication Method: Not applicable. This was bench testing against engineering specifications and international standards, not a study requiring adjudication of clinical findings.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No. This document is about a mechanical resuscitation device's electrical component update, not an AI-assisted diagnostic tool.
    • Effect size: Not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Standalone Study: The "standalone" performance here refers to the device's functional integrity with the new battery and charger. The summary states that "the system level compatibility was verified to ensure that the AutoPulse powered with the new Li-lon battery meets the operating parameters in a safe and intended manner" and that "the new battery chemistry did not alter functioning of the AutoPulse in any way and that it operates identically to the NiMH Battery powered AutoPulse." This effectively describes the standalone performance of the modified device meeting its operational specifications.

    7. The type of ground truth used

    • Ground Truth: Engineering specifications for power output, charging characteristics, battery life, environmental resilience, electrical safety, and electromagnetic compatibility established by relevant international standards (e.g., IEC 60601-1, IEC 60068, CISPR11, IEC 62133).

    8. The sample size for the training set

    • Sample Size: Not applicable. This is not a machine learning or AI-based device requiring a training set.

    9. How the ground truth for the training set was established

    • Ground Truth Establishment: Not applicable.