510(k) Data Aggregation

    K Number
    K112937

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-06-01

    (242 days)

    Product Code
    Regulation Number
    866.1640
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MRSA/SA ELITe MGB® is a qualitative in vitro diagnostic test for the direct detection of Staphylococcus aureus (SA) and methicillin-resistant Staphylococcus aureus (MRSA) using DNA purified from nasal swabs. MRSA/SA ELITe MGB® is intended to aid in the prevention and control of MRSA infections in healthcare settings. It is not intended to diagnose, guide or monitor MRSA infections, or provide results of susceptibility to oxacillin/methicillin. A negative result does not preclude MRSA/SA (Staphylococcus aureus) nasal colonization. Concomitant cultures are necessary to recover organisms for epidemiological typing or for further susceptibility testing.

    Device Description

    MRSA/SA ELITe MGB is a real-time, multiplex polymerase chain reaction (PCR) assay for the in vitro qualitative detection of MRSA and SA DNA extracted from human nasal swab samples. In this system, sample preparation and amplification/real-time detection are completed on separate instruments. Sample processing is completed on the bioMérieux NucliSENS® easyMAG® instrument with bioMérieux NucliSENS Nucleic Acid Extraction Reagents according to the manufacturer's instructions. Following processing, the extracted sample is placed in the well of a 96 well plate to which "monoreagent" is added. The monoreagent contains the primers and probes for the genes of interest and the internal control combined with master mix. The assay is performed on an Applied Biosystems 7500 FAST Dx System that consists of the 7500 FAST Dx instrument, a personal computer, 96-well plates and seals. The total system run time is 150 minutes consisting of 60 minutes for sample processing and about 90 minutes for the real time amplification and detection steps. The instrument never comes into contact with any fluids within the 96well plate. Each disposable plate is intended to test up to 96 samples, controls or any mixture thereof. The 96-well plates are not re-usable and are specific to the system. The kit contains enough reagents for 100 reactions. One positive and one negative control are required for each PCR run ; a Negative Processing Control and a Positive Processing Control are recommended to be run in each extraction run. The design of the assay includes systems to identify both the gene responsible for methicillin resistance and for a conserved portion of a gene unique to S. aureus. Thus, for a true "MRSA," both targets will be identified in roughly equal proportions. Results are determined by using an algorithm that compares output, Cq, from the cycler (called Ct in the output from the cycler.)

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the MRSA/SA ELITe MGB® device, based on the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly derived from the performance goals demonstrated in the clinical study. The summary doesn't explicitly state pre-defined acceptance criteria values but rather presents the achieved performance.

    Performance MetricAcceptance Criteria (Implicit)Reported Device Performance
    MRSA Detection
    SensitivityHigh (e.g., >90%)92.3% (88.08%-95.16%)
    SpecificityHigh (e.g., >90%)95.2% (94.32%-95.87%)
    Positive Predictive Value (PPV)Adequate (e.g., >50%)58.9% (53.67%-63.95%) (Initial) 65.2% (60.08%, 70.04%) (After Discrepant Analysis)
    Negative Predictive Value (NPV)Very High (e.g., >95%)99.4% (99.04%-99.62%) (Initial) 99.4% (99.08%, 99.65%) (After Discrepant Analysis)
    SA Detection
    SensitivityHigh (e.g., >90%)96.1% (94.48%-97.25%)
    SpecificityHigh (e.g., >90%)95.1% (94.16%-95.89%)
    PPVHigh (e.g., >80%)86.2% (83.74%-88.36%)
    NPVVery High (e.g., >95%)98.7% (98.16%-99.09%)
    Analytical Sensitivity (LoD)Consistent detection at low concentrations (e.g., 95% confidence)Average LoD: 165 CFU/mL of a swab eluate
    Analytical Reactivity (Inclusivity)Detection of diverse MRSA/MSSA strains (>95%)97.3% (All MSSA positive for SA, negative for MRSA; All MRSA positive for MRSA; 2 BORSA isolates SA positive, MRSA negative)
    Analytical Specificity (Exclusivity)No cross-reactivity with common nasal microflora/pathogens (100%)100% (with exceptions for two interfering organisms)
    ReproducibilityConsistent results across runs, operators, days, and lotsSee detailed SD and CV% values for Ct in the table (generally low variability)
    Carry-Over / Cross-ContaminationMinimal to no false positives/negatives (e.g., <5%)Zero false negatives (high MRSA samples), One false positive (negative samples) out of 55 each.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Clinical Performance Study): A total of 3271 nasal swab specimens were collected. After exclusions, 3174 eligible specimens were included in the statistical analyses.
    • Data Provenance: The data was collected from a prospective investigational study conducted at three (3) sites in unique geographies. The specific countries are not mentioned, but the submitter is based in Bothell, WA, USA, suggesting the studies were likely conducted within the USA. The patients were those eligible for MRSA screening according to the facilities' policies.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The ground truth for the clinical performance study was established using a reference culture method involving latex agglutination and cefoxitin susceptibility tests. This method is a standard laboratory procedure, and the expertise would reside with trained laboratory personnel performing these microbiological tests, rather than external clinical experts (e.g., radiologists). The document does not specify the number or qualifications of individuals performing these reference tests, as it's a standard laboratory process.

    4. Adjudication Method for the Test Set

    • Initial Ground Truth: The initial ground truth for MRSA was defined as a specimen where MRSA was identified by both latex agglutination and cefoxitin susceptibility test (after broth enrichment). For MSSA, it was negative for cefoxitin susceptibility and positive for latex agglutination.
    • Discrepant Analysis: For specimens with discordant MRSA results between the reference culture method and the MRSA/SA ELITe MGB™, further investigation was performed using sequencing of the SCCmec right extremity junction. This method served as a higher-level adjudication for refining the ground truth for discordant MRSA cases. The document does not specify if multiple sequencers were involved or if there was a consensus process for the sequencing results.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. This device is a fully automated nucleic acid amplification test (NAAT) for direct detection, not an AI-assisted diagnostic device that is interpreted by human readers. Therefore, the concept of human reader improvement with/without AI assistance is not applicable here.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, a standalone performance evaluation was performed. The MRSA/SA ELITe MGB® is an in vitro diagnostic (IVD) device that provides qualitative results (MRSA+, SA+, SA-). Its performance metrics (sensitivity, specificity, PPV, NPV) were calculated by comparing the device's output directly against the reference culture gold standard, without any human interpretation of the device's raw data that could influence the final result. The "algorithm" in this context is the automated Cq (Ct) comparison logic within the system that determines the positive/negative call for MRSA and SA targets.

    7. The Type of Ground Truth Used

    The primary ground truth used for the clinical performance study was expert consensus pathology / laboratory reference method. Specifically, it was defined by:

    • MRSA: Latex agglutination and cefoxitin susceptibility test (after broth enrichment) confirming MRSA presence.
    • MSSA: Negative cefoxitin susceptibility test and positive latex agglutination test.
    • Discrepant cases: Further adjudicated by sequencing of SCCmec right extremity junction.

    8. The Sample Size for the Training Set

    The document does not explicitly state the sample size used for a training set. This type of 510(k) submission typically focuses on validation data rather than the development and training phases of the algorithm, especially for established PCR technologies. The "Non-Clinical Performance Evaluation" references analytical studies using specific strains and dilutions, but these are for analytical validation, not for training a machine learning model.

    9. How the Ground Truth for the Training Set Was Established

    As no explicit training set is detailed for an AI/ML algorithm, the method for establishing its ground truth is not provided. However, for the analytical studies (e.g., analytical sensitivity, reactivity), the ground truth for microbial strains was established by:

    • Quantification: Cultures of strains were quantified (CFU/swab).
    • Characterization: Strains were well-characterized isolates obtained from sources like the Network on Antimicrobial Resistance in Staphylococcus aureus (NARSA) Program, American Tissue Culture Collection (ATCC), or gifts from research institutions (e.g., Medical College of Wisconsin), implying established methods for strain identification and resistance profiles.