510(k) Data Aggregation

    K Number
    K112660

    Validate with FDA (Live)

    Device Name
    POCKETCPR
    Manufacturer
    Date Cleared
    2012-05-10

    (240 days)

    Product Code
    Regulation Number
    870.5210
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PocketCPR Intended Use: To assist users in the performance of effective CPR on a patient 8 years or older.

    Device Description

    Effective cardiopulmonary resuscitation (CPR), when administered quickly after the on-set of cardiac arrest, is effective at saving lives and enabling complete neurological recovery, especially when followed by early defibrillation. The PocketCPR® was developed to provide users with real-time feedback during the administration of CPR, to assist CPR instructors during class exercises, and to provide CPR students with verbal instructions and feedback for at-home practice. When the PocketCPR is turned on, the default protocol being driven by the 2010 American Heart Association's (AHA) Guidelines, reminds the user to call for help. Once placed on the patient's chest and chest compressions have begun, PocketCPR generates a metronome at the recommended rate of chest compression, gives verbal and visual feedback on the quality of CPR being delivered and reminds the user to give two breaths after every 30 compressions. Measurement of actual chest wall movement is made by a state-of-the-art accelerometer micro-chip and microprocessor performing continuous analysis.

    AI/ML Overview

    The provided text describes the PocketCPR® device, an aid for Cardiopulmonary Resuscitation (CPR), and its substantial equivalence to a predecessor device. However, it does not include specific acceptance criteria (e.g., target performance metrics like sensitivity, specificity, accuracy for CPR feedback) or a detailed study proving the device meets such criteria.

    The bulk of the "Types of Testing" section lists various bench tests, simulated use on a manikin, usability studies, and biocompatibility, but it doesn't provide performance results against specific, quantifiable acceptance criteria.

    Therefore, I cannot fully answer your request based on the provided text for certain points as the information is not present. I will indicate where information is missing.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria for CPR Feedback: Specific numerical acceptance criteria (e.g., what percentage accuracy is required for depth, rate, or release feedback) are not explicitly stated in the provided document. The document states the device "generates a metronome at the recommended rate of chest compression, gives verbal and visual feedback on the quality of CPR being delivered and reminds the user to give two breaths after every 30 compressions." This describes the functionality but not the performance targets or the measured performance against those targets.

    Reported Device Performance: Similarly, specific quantitative performance data (e.g., measured accuracy of depth detection, rate measurements, or percentage of correct feedback instances) is not reported. The document notes "Measurement of actual chest wall movement is made by a state-of-the-art accelerometer micro-chip and microprocessor performing continuous analysis," but actual performance results are absent.

    Feature/MetricAcceptance Criteria (As stated in document)Reported Device Performance (As stated in document)
    CPR Depth Feedback AccuracyNot explicitly stated (Implied: "gives verbal and visual feedback on the quality of CPR being delivered" meeting AHA 2010 Guidelines)Not reported quantitatively. States "Measurement of actual chest wall movement is made by a state-of-the-art accelerometer micro-chip and microprocessor performing continuous analysis."
    CPR Rate Feedback AccuracyNot explicitly stated (Implied: "generates a metronome at the recommended rate of chest compression" meeting AHA 2010 Guidelines)Not reported quantitatively.
    Full Release Detection AccuracyNot explicitly stated (Implied: "allow for 'Full Release' on 'UP' stroke" feedback)Not reported quantitatively.
    Breaths Prompt AccuracyNot explicitly stated (Implied: "reminds the user to give two breaths after every 30 compressions" meeting AHA 2010 Guidelines)Device "maintains communicating 'Open Airway - Give Two Breaths' after each sequence of 30 compressions."
    Orientation Guidance AccuracyNot explicitly stated (Implied: "assist the caregiver in selecting proper orientation")Device "will also prompt 'Analysis Halted' if improper orientation has occurred and will stay silent awaiting proper orientation."

    2. Sample size used for the test set and the data provenance

    • Test Set Description: The document mentions "Simulated Use on Computerized Manakin for Data Collection" and "Usability Studies performing CPR using PocketCPR."
    • Sample Size for Test Set: The sample size for these tests is not specified.
    • Data Provenance: The document does not specify the country of origin for the data or whether the studies were retrospective or prospective. Given the mention of "Computerized Manakin" and "Usability Studies," it's highly likely the data was collected prospectively in a controlled environment.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. The type of ground truth used (manikin performance, expert observation) is not detailed, nor is the number or qualifications of any experts involved in assessing performance.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This information is not provided in the document.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: An MRMC comparative effectiveness study, which typically compares human performance with and without AI, is not described in the document. The device is a direct feedback tool, not an "AI" in the typical sense of interpreting complex diagnostic images or data for a human reader. Its function is to directly guide the user.
    • Effect Size: Therefore, effect size related to human reader improvement with AI assistance is not applicable or reported here.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: The document mentions "Measurement of actual chest wall movement is made by a state-of-the-art accelerometer micro-chip and microprocessor performing continuous analysis." While this implies an internal algorithm for measurement, no standalone performance evaluation of this algorithm's accuracy (e.g., against a gold standard for force or depth measurement) is explicitly detailed with specific metrics. The testing rather focuses on the device's output (feedback) in simulated use.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The document implies that the ground truth for the "Simulated Use on Computerized Manakin for Data Collection" and "Usability Studies" likely involves:
      • Manikin-based measurements: Computerized manikins are designed to accurately measure CPR parameters (depth, rate, recoil). This would serve as a form of objective, engineered ground truth.
      • Compliance with AHA Guidelines: The ultimate "ground truth" for the device's feedback is its adherence to the 2010 American Heart Association (AHA) Guidelines for CPR.

    8. The sample size for the training set

    • The document describes the device as being "programmed to meet the 2010 AHA Guidelines." It does not mention a "training set" in the context of machine learning. The device's "programming" is based on the AHA guidelines rather than learned from a dataset.
    • Therefore, a "training set sample size" in the conventional machine learning sense is not applicable or provided.

    9. How the ground truth for the training set was established

    • As above, there is no mention of a "training set" for a machine learning model. The device's operational logic and feedback parameters are derived directly from the 2010 American Heart Association (AHA) Guidelines for CPR. These guidelines represent the accepted "ground truth" for effective CPR practices.