Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K112523
    Date Cleared
    2012-01-06

    (128 days)

    Product Code
    Regulation Number
    866.5750
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For in vitro diagnostic use with the IMMULITE® 2000 Analyzer -- for the quantitative measurement of allergen-specific IgE in human serum, as an aid in the clinical diagnosis of IgE-mediated allergic disorders. The test results are to be used in conjunction with clinical findings and other laboratory tests.

    Device Description

    IMMULITE® 2000 3gAllergy™ Specific IgE is a solid-phase, two-step, chemiluminescent immunoassay that exploits liquid phase kinetics in a bead format. The allergens are covalently bound to a soluble polymer matrix, which in turn is labeled with a ligand. The use of an amino acid co-polymer amplifies the amount of allergen that the matrix can support.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the IMMULITE® 2000 3gAllergy™ Specific IgE Assay, based on the provided text:

    Acceptance Criteria and Device Performance

    The study evaluates the analytical performance and clinical performance of the device.

    Analytical Performance Acceptance Criteria:

    The key acceptance criteria for analytical performance are related to precision, linearity, and specificity. While explicit quantitative criteria were provided for precision, linearity had implied criteria based on excellent regression statistics, and specificity was demonstrated by successful inhibition.

    Acceptance Criteria CategorySpecific CriteriaReported Device PerformanceComments
    PrecisionIntra-assay Variability: Less than 15% CV for all allergen lots.All allergens met this criterion. For example, nOle e 1 (Lot 201) had within-run CVs ranging from 3.95% to 6.23%.This indicates consistent results within a single test run.
    Inter-assay Precision: Less than 15% CV for all allergen lots.All allergens met this criterion. For example, nOle e 1 (Lot 201) had total CVs ranging from 5.85% to 7.03%.This indicates consistent results across multiple test runs over different days.
    Lot-to-Lot Precision: Less than 20% CV for all 5 allergens.All 5 allergens met this criterion. For example, combined nOle e 1 lots showed total CVs up to 6.94%.This demonstrates consistency of results when using different batches (lots) of the allergen.
    Negative Control: Average negative control remained negative (< 0.10 kU/L) during precision testing.The average negative control remained negative (< 0.10 kU/L) throughout precision testing on all allergens.This confirms proper function and absence of non-specific binding for negative samples.
    LinearityExcellent regression statistics (slope close to 1, intercept close to 0, R close to 1) comparing observed to expected data across a range of concentrations.The reported R values were all 0.995 or higher, slopes were close to 1 (e.g., 0.997 for nOle e 1, 0.975 for Cat Serum Albumin), and intercepts were close to 0.This indicates that the assay accurately measures IgE concentrations proportionally across its analytical range.
    Specificity (Inhibition)Demonstrate concentration-dependent inhibition by relevant inhibitor extract, with a target of 50% inhibition at the highest inhibitor concentration. No significant inhibition by unrelated allergen extracts.The inhibition plots demonstrated concentration-dependent inhibition and the target 50% inhibition was met for all tested allergens. Inhibition with unrelated extracts did not generate any significant inhibition.This confirms that the assay specifically detects the intended allergen-specific IgE and does not show significant cross-reactivity with unrelated allergens.

    Clinical Performance Acceptance Criteria:

    The acceptance criteria for clinical performance are based on the agreement of the device's results with clinical diagnosis, measured by sensitivity, specificity, and overall agreement.

    Acceptance Criteria CategorySpecific CriteriaReported Device PerformanceComments
    Clinical AgreementSensitivity, Specificity, and Overall Agreement with clinical diagnosis within acceptable confidence intervals (though no explicit numerical target for the confidence interval width or minimum performance metric was given, the reported values demonstrate acceptable performance given the context of the device).Sensitivity: 76.7% (Lower CI: 71%, Upper CI: 83%)This indicates the proportion of individuals with allergic disorders that the test correctly identifies as positive.
    Specificity: 95.8% (Lower CI: 94%, Upper CI: 98%)This indicates the proportion of individuals without allergic disorders that the test correctly identifies as negative.
    Agreement: 90.5% (Lower CI: 88%, Upper CI: 93%)This indicates the overall accuracy of the test compared to clinical diagnosis.

    Study Details

    2. Sample sizes used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Analytical Test Set Sample Size:
      • Precision: For each allergen lot and each positive sample level, N=80 measurements were performed (2 aliquots per run, 2 runs per day, 20 days). For combined lot precision, N=240 (80 measurements x 3 lots) where applicable. Negative controls were also tested.
      • Linearity: 11 or 12 data points (depending on the allergen) from 2-fold serial dilutions, including undiluted (neat) samples.
      • Specificity (Inhibition): Not explicitly stated how many unique serum samples were used "a single serum sample or pool of sera". For cross-reactivity, one positive sample was tested with three unrelated allergen extracts.
    • Clinical Test Set Sample Size:
      • Total Patients: 693 individuals were tested (193 clinically positive, 500 clinically normal).
    • Data Provenance: Not explicitly stated for either analytical or clinical studies. However, the manufacturer (Siemens Healthcare Diagnostics Inc. and Siemens Healthcare Diagnostics Products, LTD) has locations in Tarrytown, NY, and Llanberis, Gwynedd, UK. It is likely that the data originates from studies conducted or coordinated by these facilities, potentially involving samples from different geographic locations. The document does not specify whether the data was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    • Analytical Studies: No external experts were used; ground truth was established by the experimental design (e.g., known concentrations for linearity, known positive/negative samples for precision and specificity).
    • Clinical Studies: The ground truth for clinical performance was "clinical documentation of presence or absence of signs, symptoms and other diagnostic evidence of allergen sensitivity." The number and qualifications of experts (e.g., allergists, physicians) involved in making these clinical diagnoses are not specified in the provided text.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Clinical Studies: The document does not describe any specific adjudication method for establishing the "clinical diagnosis." It refers to "clinical findings and other laboratory tests," implying a standard clinical diagnostic process rather than a formalized adjudication among multiple experts for the purpose of the study.
    • Analytical Studies: Adjudication is not applicable as ground truth is based on known analytical properties.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is an in-vitro diagnostic assay for quantitative measurement of allergen-specific IgE in human serum. It is an automated laboratory test, not an imaging device or AI-assisted diagnostic tool that would involve human readers or AI assistance in interpretation. The study evaluates the performance of the analytical device itself.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • This is a standalone diagnostic assay. The IMMULITE® 2000 3gAllergy™ Specific IgE Assay, when run on the IMMULITE® 2000 Analyzer, operates as an automated system. The performance data presented (precision, linearity, specificity, and clinical agreement) reflects the "algorithm only" or "device only" performance, without human interpretive input for the final result beyond operating the analyzer and reviewing the quantitative output. The results are quantitative IgE levels, not qualitative interpretations requiring human judgment (e.g., image reading).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Analytical Studies: The ground truth was based on known properties of the samples and experimental design:
      • Precision: Replicates of the same sample, the true value of which is the mean of the replicates.
      • Linearity: Gravimetrically prepared dilutions with known expected concentrations.
      • Specificity: Known inhibitor extracts and known unrelated allergens.
    • Clinical Studies: The ground truth for clinical performance was "clinical documentation of presence or absence of signs, symptoms and other diagnostic evidence of allergen sensitivity." This can be classified as a form of clinical diagnosis and outcomes data, likely based on patient history, physical examination, symptom evaluation, and other relevant diagnostic tests, rather than a single definitive pathology result.

    8. The sample size for the training set

    • Not applicable / Not explicitly stated. This document describes a 510(k) submission for new allergens to an existing IMMULITE® 2000 3gAllergy™ Specific IgE assay. It details the validation and performance studies for these additional allergens. The IMMULITE® 2000 system and the 3gAllergy™ assay are commercially available, validated technologies. The concept of a "training set" as understood in machine learning (where algorithms are "trained" on data) is not directly applicable here. The assay itself uses established immunochemical principles; it's not a machine learning algorithm that undergoes a training phase with a distinct dataset.

    9. How the ground truth for the training set was established

    • Not applicable. As explained above, this device does not involve an "AI algorithm" or "machine learning" in the traditional sense that would require a separate training set and ground truth establishment for that training phase. The analytical and clinical performance studies are for validation of the assay for the specific new allergens.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1