Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K111747

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-07-12

    (20 days)

    Product Code
    Regulation Number
    870.1340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Endologix AFX Introducer Systems are intended for use to facilitate the introduction of guidewires, catheters, and other medical devices into the vasculature and minimize blood loss associated with such introduction.

    Device Description

    The Endologix AFX Introducer Systems consist of an Introducer Sheath (45 cm working length) and either one or two Dilators. Model S17-45DD contains dual dilators: a dual-lumen dilator for introduction and management of two guidewires; and, a single-lumen dilator. Model S17-45 contains only the single-lumen dilator.

    Each AFX Introducer System is enclosed in a sterile package. Each system is compatible with a standard 0.035" guidewire.

    • Introducer Sheath: The Introducer Sheath provides a conduit for introducing other interventional devices, including guidewires and interventional catheters, into the vasculature. The main components of this assembly are a hydrophilically-coated, PTFE lined, 55D Pebax introducer sheath, acrylobutylstyrene (ABS) hemostasis valve housing with butyl rubber valve, and silicone gasket assembly, 3 way stopcock/valve, and sideport diethylhexyl phthalate (DEHP)-free polyvinylchloride (PVC) tubing. The Pebax introducer sheath also includes an embedded stainless steel braid for increased strength and kink resistance, and embedded Vectrands for increased longitudinal strength. The distal tip of the introducer sheath includes an embedded radiopaque marker composed of 90% Platinum/10% Iridium. There are no changes to the Introducer Sheath from that cleared under K110090.
    • Dilators: The dilators for these systems are used to provide support and stability to the Introducer Sheath during deployment into the vasculature. The dilators are designed specifically for the Endologix AFX Introducer Systems and can accommodate a 0.035" guidewire through the central lumen. One dilator is a single lumen device and the other dilator is a dual lumen device. Both dilators are composed of 63D Pebax with 30% Barium Sulfate (BaSO4) for radiopacity and include a hydrophilic coating on the outer surface. These dilators are designed in diameter and length to be placed within the AFX Introducer Sheath component. The dilators have a tapered, atraumatic distal tip. The proximal end of the dilators includes a luer port. The dilators provide stiffness to the assembly so that the introducer sheath can be placed within the vasculature during standard vascular access. There are no changes to the Dilators from that cleared under K110090. Model S17-45 contains only the single lumen dilator; Model S17-45DD contains both dilators with the AFX Introducer System within the package.
    AI/ML Overview

    The provided text describes a 510(k) submission for the Endologix AFX Introducer System, which is a medical device. This type of submission focuses on demonstrating substantial equivalence to a predicate device rather than presenting a clinical study proving device performance against specific acceptance criteria.

    Therefore, many of the requested categories for a study report on acceptance criteria cannot be directly addressed from the provided document. The document lists performance tests conducted to establish equivalence, not to meet specific clinical performance acceptance criteria in the manner of a standalone study for a new device's efficacy or safety.

    Here's an attempt to extract relevant information and explain where the document does not provide the requested details:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in the traditional sense for clinical performance. Instead, it lists tests performed to establish equivalence to a predicate device. The "reported device performance" in this context refers to the successful completion of these tests in accordance with the listed standards, thus demonstrating equivalence and safety/functionality.

    Test DescriptionApplicable Test Standard(s)Reported Device Performance (Implied Acceptance)
    Functional/Mechanical QualificationISO 10555-1:1995The device successfully met the functional and mechanical requirements outlined in ISO 10555-1:1995, supporting its intended use for introducing guidewires, catheters, and other medical devices into the vasculature.
    EtO Gas Sterilization ValidationANSI/AAMI/ISO 11135-1:2007, AAMI TIR 28: 2009The device successfully underwent Ethylene Oxide (EtO) gas sterilization validation according to the specified standards, ensuring its sterility for clinical use.
    3-year Shelf-Life Packaging QualificationISO 11607-1:2006The device's packaging successfully met the qualification requirements for a 3-year shelf-life as per ISO 11607-1:2006, ensuring product integrity and sterility over time.
    Biocompatibility TestingISO 10993-1:2009The device demonstrated biocompatibility in accordance with ISO 10993-1:2009, indicating it is safe for contact with human tissue.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not specified. The document refers to "functional qualification" and other testing but does not provide details on the number of devices or components sampled for these tests.
    • Data Provenance: The tests were conducted to support a 510(k) submission to the FDA for a device manufactured by Endologix, Inc. in Irvine, CA, USA. The data would be considered internal company testing. The document does not specify if the tests were prospective or retrospective in the clinical study sense; rather, they are part of the device development and validation process.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    • Not applicable. This information is typically relevant for clinical studies involving expert interpretation (e.g., image analysis, diagnoses). The tests described (functional, sterilization, packaging, biocompatibility) are laboratory-based and do not involve human expert "ground truth" establishment in this manner. Evaluation is against defined technical standards and specifications.

    4. Adjudication Method for the Test Set

    • Not applicable. As above, adjudication methods are typically used in clinical studies for resolving discrepancies in expert interpretations. The device tests described are objective, pass/fail evaluations against technical standards.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, an MRMC comparative effectiveness study was not done. The document describes a 510(k) submission focused on demonstrating substantial equivalence through technical and functional testing, not a clinical study comparing human reader performance with and without AI.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • Not applicable. The Endologix AFX Introducer System is a physical medical device (introducer sheath and dilators) and does not involve an algorithm or AI. Therefore, a standalone algorithm performance study is not relevant.

    7. Type of Ground Truth Used

    • Technical standards and specifications. For the tests listed, the "ground truth" is defined by the requirements outlined in the applicable ISO and AAMI standards (e.g., ISO 10555-1 for functional aspects, ISO 10993-1 for biocompatibility). Successful completion of these tests means the device met the objective criteria set by these standards.

    8. Sample Size for the Training Set

    • Not applicable. As this is a physical medical device and not an AI/algorithm, there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. (See point 8).

    In summary, the provided text is a 510(k) summary for a medical device that demonstrates substantial equivalence to a predicate device through engineering and laboratory testing. It is not a report on a clinical trial with acceptance criteria for clinical efficacy or diagnostic performance, which would typically involve details about human experts, ground truth, and study designs like MRMC or standalone performance.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1