    Manufacturer
    Date Cleared
    2011-07-19

    (29 days)

    Product Code
    Regulation Number
    892.1720
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended for use by a qualified/trained doctor or technologist on both adult and pediatric patients for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with patient sitting, standing or lying in the prone or supine positions. Not intended for mammography.

    Device Description

    This device is simply the combination of two cleared devices, the Wireless Portable Detector FD-W17 (K090625) marketed by Philips Medical Systems and the Easy Moving Plus, Mobile Diagnostic X-Ray (K090322) made by Sedecal. The x-ray source is a motor driven mobile x-ray and the x-ray receptor panel is a digital wireless unit. The Wireless Portable Detector FD-W17 consists of three main parts: Portable radiography detector (x-ray sensitive part), docking station which is directly connected to the radiographic workstation, backup cable which can connect the detector to the docking station if the wireless connection cannot be used. Detector size: 35 x 43 cm (14 x 17") Image matrix size: 3000 pixels x 2400 pixels. Pixel size 144 um、Image resolution up to 3.5 LP/mm

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the MobileDiagnost wDR device:

    This 510(k) summary focuses on demonstrating substantial equivalence for the MobileDiagnost wDR as a combination of two already cleared predicate devices. It does not contain a traditional clinical study with detailed performance metrics against specific acceptance criteria for a novel algorithm. Instead, the "study" is a comparison to predicate devices, focusing on the absence of significant differences.


    1. A table of acceptance criteria and the reported device performance

    The submission does not explicitly define acceptance criteria as quantitative performance metrics for the MobileDiagnost wDR itself (e.g., sensitivity, specificity, accuracy). Instead, the acceptance is based on demonstrating substantial equivalence to its predicate devices, implying that if it performs comparably to already cleared devices, it meets the required safety and effectiveness.

    CharacteristicAcceptance Criteria (implied by predicate)Reported Device Performance (MobileDiagnost wDR)
    Intended UseSame as Sedecal Easy Moving Digital (K090322)SAME
    ConfigurationBattery or line operated mobileSAME
    Performance Standard21 CFR 1020.30SAME
    GeneratorHigh frequency made by SedecalSAME
    Generator power levels20 to 50 kw (4 models)20 to 50 kw (4 models)
    CollimatorRalco R221 DHHSRalco 108F DHHS (equivalent)
    Image acquisitionDigital CANON CDXI-50G (K031447)Philips Wireless Portable Detector FD-W17 (K090625)
    Electrical safetyElectrical Safety per IEC-60601, UL listedSAME
    Clinical ImagesNo significant differences compared to predicateNo significant differences
    Software ValidationAll tests passedTests performed, results indicate safety and effectiveness
    Risk AnalysisAcceptable risk profileAnalysis performed
    Hardware ComplianceUS Performance Standards, CSA CertifiedConforms to US Performance Standards, CSA Certified

    2. Sample size used for the test set and the data provenance

    • Sample Size: The document states "Clinical images were acquired and compared to our predicate images." It does not specify the number of clinical images acquired.
    • Data Provenance: Not explicitly stated (e.g., retrospective/prospective, country of origin). It's implied that the images were taken for the purpose of this comparison.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This was a direct comparison of images from the new device to those from predicate devices, likely for visual assessment of image quality, not for diagnostic accuracy against a ground truth diagnosis established by experts. The comparison involved assessing clinical images for "significant differences," which would typically be done by qualified personnel familiar with diagnostic image quality, but no specific number or qualification of experts is mentioned for this task.

    4. Adjudication method for the test set

    Not applicable. There's no mention of an adjudication process, as the goal was a direct comparison of image quality between two systems, not a diagnostic accuracy assessment requiring consensus on diagnoses.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    No. A MRMC study was not performed. The study described is a comparison of "Clinical images... acquired and compared to our predicate images." This indicates a technical comparison rather than a human reader performance study.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, in a way. The "algorithm" here refers to the overall device's image acquisition and processing. The comparison of clinical images and software validation are assessments of the device's standalone performance in producing images and functioning correctly, without explicit human assistance in the "performance" measurement.

    7. The type of ground truth used

    Not explicitly stated. For the "clinical images were acquired and compared to our predicate images" statement, the "ground truth" would be the expected image quality and diagnostic information provided by the predicate device. The comparison aimed to show that the new device produced images of essentially the same quality and diagnostic utility as the predicate. It did not involve comparing readings against a histopathological diagnosis or clinical outcomes.

    8. The sample size for the training set

    Not applicable. The device is a combination of existing cleared devices (a mobile X-ray unit and a digital detector). This is not an AI/ML algorithm that requires a separate "training set." The software validation refers to functional testing, not model training.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for an AI/ML algorithm mentioned in this submission.