510(k) Data Aggregation

    K Number
    K111724

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-09-12

    (84 days)

    Product Code
    Regulation Number
    872.4840
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Straumann® TiBrush is a debridement instrument for titanium dental implants subjected to peri-implantitis. It is indicated for the open debridement of titanium implant surfaces in bone defects caused by peri-implantitis.

    Device Description

    The Straumann® TiBrush is made of Grade 5 Titanium and stainless steel for single patient use and is attached to a dental handpiece. It is used during open flap surgery to remove biofilms (plaque and calculus) from the surfaces of dental implants where the surface of the implant is exposed due to peri-implantitis. The TiBrush mechanically removes the biofilms from the surface of the implants without damaging the integrity of the roughened implant surface.

    AI/ML Overview

    The provided document describes the acceptance criteria and performance testing for the Straumann® TiBrush. Here's a breakdown of the requested information:

    1. Table of acceptance criteria and the reported device performance

    TestAcceptance CriteriaReported Device Performance
    Bristle wear during cleaningFree bristle length > 0.40 mm after implant cleaningPassed
    Damage to implant surfaceSand blasted, acid etched surface cleaned by TiBrush scratch the surface less than the stainless steel curettePassed
    Corrosion testingNo corrosionPassed
    Cleaning of implant surfaceRemoves more plaque than curettePassed
    Performance at different speeds; 600, 900, 1200, and 3000 rpmNo difference in performancePassed
    Biocompatibility testingMet the requirements of ISO 10993, various partsMet the requirements
    Sterilization validationMet the requirements of ISO 11137Met the requirements

    2. Sample size used for the test set and the data provenance

    The document does not specify the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective). The performance testing section mentions "implant cleaning" and "implant surface" but does not detail the number of implants or the source of this test data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. The testing appears to be primarily bench testing (physical and mechanical properties, cleaning efficacy) rather than a study involving human expert assessment for ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided in the document, as the testing described does not appear to involve an adjudication process as typically understood in studies involving human reads.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no indication that a multi-reader multi-case (MRMC) comparative effectiveness study was done, nor is there any mention of AI assistance. The device is a mechanical cleaning instrument, not an AI or imaging diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable as the Straumann® TiBrush is a mechanical dental instrument, not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the performance testing appears to be based on:

    • Mechanical measurements/observations: Bristle length, scratch assessment, corrosion absence, plaque removal quantification (implicitly, through comparison to curette).
    • Adherence to standards: ISO 10993 for biocompatibility and ISO 11137 for sterilization.

    8. The sample size for the training set

    This question is not applicable. The device is a mechanical instrument, not a machine learning algorithm that requires a training set.

    9. How the ground truth for the training set was established

    This question is not applicable. The device is a mechanical instrument, not a machine learning algorithm.