Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K111611

    Validate with FDA (Live)

    Date Cleared
    2011-07-29

    (50 days)

    Product Code
    Regulation Number
    878.3610
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MERIT ENDOTEK™ EndoMAXX™ Fully Covered Esophageal Stent is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors and for occlusion of esophageal fistulae.

    Device Description

    The MERIT ENDOTEK™ EndoMAXX Fully Covered Esophageal Stent is comprised of two components: the radiopaque self-expanding Nitinol stent and the deployment catheter. The stent is completely covered with a biocompatible silicone membrane. The stent expansion results from the physical properties of the metal and the proprietary geometry. The stent is designed with a larger diameter at the distal and proximal ends to reduce the possibility of migration. The overall stent geometry is designed to minimize foreshortening upon expansion, thus facilitating improvement in deployment accuracy. The proximal and distal ends of the stent are threaded with a suture intended for use in proximal and distal repositioning of the stent. The stent is deployed with a dedicated deployment catheter. The deployment catheter consists of two coaxial sheaths attached to a deployment handle. The handle permits one-handed positioning and deployment via a trigger mechanism. The exterior sheath serves to constrain the stent until the sheath is retracted during deployment. A radiopaque tip and marker on the inner shaft proximal to the stent aid the operator in determining stent position to the deployment threshold. Once deployment is initiated, the stent cannot be reconstrained. The stent can be repositioned proximally until the first deployment trigger is deployed for a two deployment trigger device or until the second deployment trigger is deployed for a three deployment trigger device. The inner tube of the coaxial sheath catheter contains a central lumen that will accommodate a 0.035" (0.89mm) guide wire. This feature is designed to allow safe guidance of the deployment catheter to the intended implant site while minimizing the risk of esophageal injury from the deployment catheter tip.

    AI/ML Overview

    The Merit Medical Systems, Inc. EndoMAXX™ Fully Covered Esophageal Stent is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors and for occlusion of esophageal fistulae. The device was found substantially equivalent to the ALIMAXX-ES™ Esophageal Stent.

    1. A table of acceptance criteria and the reported device performance:

    The document lists numerous tests that were successfully conducted, indicating the device met the established acceptance criteria for each without providing specific quantitative values for acceptance criteria or specific performance beyond "successfully completed."

    Acceptance Criteria CategoryReported Device Performance
    Mechanical/Physical Testing
    DeploymentSuccessfully completed
    Expansion ForceSuccessfully completed
    Compression ForceSuccessfully completed
    DimensionalSuccessfully completed
    CorrosionSuccessfully completed
    Tensile StrengthSuccessfully completed
    Guidewire compatibilitySuccessfully completed
    Distal Tip InsertionSuccessfully completed
    Stent foreshorteningSuccessfully completed
    Deployment Catheter ProfileSuccessfully completed
    Deployment Trigger StrokeSuccessfully completed
    Trigger safetySuccessfully completed
    Stent Deployment ForceSuccessfully completed
    Stent Repositioning after partial deploymentSuccessfully completed
    Deployment AccuracySuccessfully completed
    Stent repositioning, after full deploymentSuccessfully completed
    Stent expansion and condition after deploymentSuccessfully completed
    System integritySuccessfully completed
    Radius of leading stent strutsSuccessfully completed
    Stent Springback after purse stringing sutureSuccessfully completed
    Compression Force Pre-fatigueSuccessfully completed
    Expansion Force Pre-fatigueSuccessfully completed
    Stent in-fold distanceSuccessfully completed
    Stent in-fold forceSuccessfully completed
    Stent flexibilitySuccessfully completed
    Suture cutting with forcepsSuccessfully completed
    Suture tensileSuccessfully completed
    Stent FatigueSuccessfully completed
    Cover Integrity after FatigueSuccessfully completed
    Stent Springback after Fatigue (dimensional)Successfully completed
    Compression Force after FatigueSuccessfully completed
    Expansion Force after FatigueSuccessfully completed
    Stent TensileSuccessfully completed
    First trigger to floater tensileSuccessfully completed
    Floater to internal connector tensileSuccessfully completed
    Second trigger to internal connector tensileSuccessfully completed
    For 3-trigger device, 1st trigger to second trigger tensileSuccessfully completed
    Handle to outer support tensileSuccessfully completed
    Handle to metal hypotube and inner shaft bond tensileSuccessfully completed
    Outer shaft to internal connector bond tensileSuccessfully completed
    Outer shaft tensileSuccessfully completed
    Pod to outer shaft tensileSuccessfully completed
    Tip to inner sheath bondSuccessfully completed
    Stent Anti-migration Strut HeightSuccessfully completed
    Biocompatibility Testing
    CytotoxicitySuccessfully completed
    SensitizationSuccessfully completed
    ImplantationSuccessfully completed
    Sub-Chronic ToxicitySuccessfully completed
    Visibility & Compatibility
    Visualize stent with fluoroscopySuccessfully completed
    Fluoroscopic visibility of deployment catheterSuccessfully completed
    MR CompatibilitySuccessfully completed
    Packaging & Sterilization
    Packaging ConditionSuccessfully completed
    Packaging performance (after accelerated aging and simulated shipping/handling)Successfully completed

    2. Sample size used for the test set and the data provenance:

    The document is a 510(k) summary and primarily focuses on non-clinical testing. It does not provide details on sample sizes for clinical test sets as clinical trials are typically not required for 510(k) applications demonstrating substantial equivalence. The tests performed are engineering and benchtop tests, therefore, information regarding data provenance (e.g., country of origin of the data, retrospective or prospective) is not applicable in the context of clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    This information is not applicable. The study described is not a clinical study involving diagnosis or interpretation by experts to establish ground truth. It is a series of non-clinical, benchtop and laboratory tests.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    This information is not applicable as the study involves non-clinical testing rather than clinical data requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No MRMC comparative effectiveness study was done. This 510(k) summary describes non-clinical testing for a medical device (esophageal stent) and does not involve AI assistance or human reader studies.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This information is not applicable. The device is a physical medical implant (esophageal stent), not a software algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    For the non-clinical tests conducted, the "ground truth" refers to the established engineering specifications, material properties, and biological safety standards (e.g., ASTM, ISO, USP standards) against which the device's performance was measured. For example:

    • Mechanical/Physical Tests: Ground truth is defined by the requirements of the specific ASTM standards (e.g., mechanical strength, deployment forces, corrosion resistance) and internal specifications for device dimensions and functionality.
    • Biocompatibility Tests: Ground truth is established by the requirements of ISO 10993 standards for cytotoxicity, sensitization, and implantation effects.

    8. The sample size for the training set:

    This information is not applicable. The study describes non-clinical performance and safety testing of a physical medical device, not the training of an algorithm or AI model.

    9. How the ground truth for the training set was established:
    This information is not applicable for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1