    K Number
    K111529

    Validate with FDA (Live)

    Date Cleared
    2011-07-29

    (57 days)

    Product Code
    Regulation Number
    876.1075
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The M-CORE disposable biopsy needle is intended for use with the Bard® Magnum® biopsy instrument in obtaining biopsies from soft tissues such as liver, kidney, prostate, breast, spleen, lymph nodes and various soft tissue tumors. The M-CORE needle is not intended for use in bone.

    Device Description

    The M-CORE disposable core biopsy needle is a sterile, disposable biopsy needle featuring stainless steel cutting cannula and stylet, with sample is a lotenio, disposable biopsy needle leading stames steaturing statiness steat themoplastic removable spacer is provided to allow easy insection of both hubs into a Bard@ Magnum® reusable biopsy instrument. An LDP of the county the neder in both huss in a Barde Magnum® reuseble available in gauges from 14 - 18 and lengths from 10 - 25 centimeters.

    AI/ML Overview

    The provided 510(k) summary for the M-CORE disposable core biopsy needle does not contain the detailed information necessary to fully answer all aspects of your request regarding acceptance criteria and a study proving those criteria are met. The document focuses on demonstrating substantial equivalence to predicate devices based on design and materials, and mentions "Performance testing summary" but does not elaborate on the specific tests or their results within this summary.

    Here's a breakdown of what can be extracted and what is missing, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The 510(k) summary does not explicitly state acceptance criteria in terms of specific performance metrics (e.g., biopsy sample quality scores, tensile strength, or cutting efficiency thresholds). Instead, it relies on demonstrating that the device is "substantially equivalent" to predicate devices, implying similar performance.

    The summary states: "Performance testing confirms that the quality of the samples obtained with the M-CORE disposable core biopsy needle is substantially equivalent to the predicate device(s)." This is a general statement of performance, not a quantified measurement against an acceptance criterion.

    The table provided in the document compares device characteristics, not performance metrics:

    CharacteristicReported Device (M-CORE Needle) Performance (Implicitly equivalent to predicate)Acceptance Criteria (Not explicitly stated in this document)
    Cannula and Stylet Material304SS(Implicitly: Same as predicate)
    Cannula and Stylet Hub MaterialPolycarbonate(Implicitly: Same as predicate)
    SheathLDPE(Implicitly: Same as predicate)
    Stylet Sample Notch19mm(Implicitly: Same as predicate)
    SpacerThermoplastic(Implicitly: Same as predicate)
    Non-PyrogenicYes, LAL TestYes (Implicitly: Demonstrated by LAL Test)
    Sterile PackagingTyvek pouch(Implicitly: Same as predicate)
    Sterilization MethodGamma, 11137, VDMax25, SAL 10-6 Dose: 25 - 40 kGy(Implicitly: Equivalent or superior to predicate's EO SAL 10-6 or Gamma)

    2. Sample Size Used for the Test Set and Data Provenance

    The 510(k) summary does not provide any information regarding the sample size used for performance testing (e.g., number of biopsies performed, number of needles tested). It also does not specify the data provenance (e.g., country of origin, retrospective or prospective).

    3. Number of Experts and Qualifications for Ground Truth

    No information is provided about experts used to establish ground truth. The summary mentions "the quality of the samples obtained," suggesting an evaluation of biopsy specimens, but does not detail how this quality was assessed or by whom.

    4. Adjudication Method for the Test Set

    No information is provided regarding any adjudication method for performance testing.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC study is mentioned or implied. This device is a biopsy needle, not an imaging or diagnostic algorithm that would typically involve human readers.

    6. Standalone (Algorithm Only) Performance Study

    Not applicable. This device is a physical medical instrument, not an algorithm.

    7. Type of Ground Truth Used

    While not explicitly called "ground truth," the performance testing likely involved assessing the "quality of samples obtained." This would typically be evaluated through:
    * Histopathology/Pathology: Examination of the tissue samples by a pathologist to determine adequacy for diagnosis.
    * Visual Inspection/Gross Examination: Assessment of the sample size, integrity, and appearance.
    The document does not specify which of these, or any other method, was used.

    8. Sample Size for the Training Set

    Not applicable. This device is a physical medical instrument, not a machine learning algorithm that requires a training set.

    9. How Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for a physical medical device.

    In summary, the provided 510(k) notification primarily focuses on demonstrating substantial equivalence through a comparison of design and materials with predicate devices, and a general statement about performance testing confirming sample quality. It lacks the specific details about acceptance criteria, study sizes, methodologies, and expert involvement that your detailed questions request. This level of detail is often found in the full test reports referenced (e.g., "See Section 18" for Performance Testing Summary), but not typically within the concise 510(k) summary itself.