510(k) Data Aggregation
(106 days)
The CD HORIZON® Spinal System with or without SEXTANT® instrumentation is intended for posterior, non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radio- graphic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); turnor; pseudarthrosis; and/or failed previous fusion. Except for hooks, when used as an anterolateral thoracic/lumbar system, the CD HORIZON® Spinal System may also be used for the same indications as an adjunct to fusion. With the exception of degenerative disc disease, the CD HORIZON® LEGACY™ 3.5mm rods and the CD HORIZON® Spinal System PEEK rods and associated components may be used for the aforementioned indications in skeletally mature patients as an adjunct to fusion. The 3.5mm rods may be used for the specific pediatric indications noted below. When used for posterior non-cervical pedicle screw fixation in pediatric patients, the CD HORIZON® Spinal System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The CD HORIZON® Pediatric Spinal System is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
The CDHORIZON SPIRE™ Plate is a posterior, single level, non-pedicle supplemental fixation device intended for use in the non-cervical spine (T1-S1) as an adjunct to fusion in skeletally mature patients. It is intended for plate fixation/attachment to spinous processes for the purpose of achieving supplemental fixation in the following conditions: degenerative disc disease (as previously defined); spondylolisthesis; trauma; and/or tumor. In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be connected to the VERTEX® Reconstruction System with the VERTEX® rod connector. Refer to the VERTEX® Reconstruction System Package Insert for a list of the VERTEX® indications of use.
The purpose of this Traditional 510(k) is to add Threaded Rods to the CD HORIZON® Spinal System.
The CD HORIZON® Spinal System is intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the thoracic, lumbar, and/or sacral spine.
The CD HORIZON® Spinal System consists of a variety of shapes and sizes of rods, hooks, screws, CROSSLINK® Plates, staples and connecting comportents, as well as implant components from other Meduonic spinal systems, which can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case.
A subset of CD HORIZON® Spinal System components may be used for posterior pedicle screw fixation in pediatric cases. These constructs may be comprised of a variety of shapes and sizes of rods (ranging in diameters from 3.5mm to 6.35mm), hooks, screws. CROSSLINK® Plates, and connecting components. Similarly to the CD HORIZON® implants used in adult cases, these components can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case. Certain components within the CD HORIZON® Spinal System are specifically excluded for use in pediatric patients. These include PEEK rods, CD HORIZON® SPIRE spinous process plate devices, Shape Memory Alloy Staples, DYNALOK® bolts, and TSRH® screws and washers. All screws used in pediatic cases are only cleared for use via a posterior approach. All of the components used in pediatric cases are fabricated from medical grade stainless steel, medical grade titanium alloy, and medical grade cobalt-chromiummolybdenum alloy.
Certain implant components from other Medtronic spinal systems can be used with the CD HORIZON® Spinal System in non-pediatric cases. These components include TSRH® rods, hooks, screws, plates, CROSSLINK® plates, connectors, staples, washers, GDLH® rods, hooks, connectors and CROSSLINK® bar and connectors; LIBERTY® rods and screws; DYNALOK® PLUS and DYNALOK CLASSIC® bolts along with rod/bolt connectors; and Medironic Multi-Axial rods and screws. Please note that certain components are specifically designed to connect to 3.5mm, 4.5mm, 5.5mm diameter rods or 6.35mm diameter rods, while other components can connect to both 5.5mm rods and 6.35mm diameter rods. Care should be taken so that the correct components are used in the spinal construct.
CD HORIZON® hooks are intended for posterior use only. CD HORIZON® staples and CD HORIZON® ECLIPSE® rods and associated screws are intended for anterior use only. However, for patients of smaller stature and pediatric patients, CD HORIZON®4.5mm rods and associated components may be used posteriorly. The CD HORIZON® Spinal System implant components are fabricated from medical grade stainless steel, medical grade titanium, titanium alloy, medical grade cobalt-chromium-molybdenum alloy, or medical grade PEEK OPTIMA-LT1. Certain CD HORIZON® Spinal System components may be coated with hy-droxyapatite. No warranties expressed or implied are made. Implied warranties of merchantability and fitness for a particular purpose or use are specifically excluded.
Never use stainless steel and titanium implant components in the same construct. Medical grade titanium, titanium alloy, and/or medical grade cobalt-chromium-molybdenum alloy may be used together. Never use titanium, titanium alloy, and/or medical grade cobaltchromium-molybdenum alloy with stainless steel in the same construct. The CD HORIZON® Spinal System also includes anterior staples made of Shape Memory Alloy (Nitinol - NiTi). Shape Memory Alloy is compatible with titanium, titanium alloy, and cobalt-chromium-molybdenum alloy. Do not use with stainless steel. These staples are not to be used in pediatric patients. PEEK OPTIMA-LT1 implants may be used with stainless steel, titanium, or cobalt-chromium-molybdenum alloy implants. CD HORIZON@PEEK Rods are not to be used with CROSSLINK@Plates or in pediatric patients. To achieve best results, do not use any of the CD HORIZON® Spinal System implant components with components from any other system or manufacturer unless specifically allowed to do so in this or another Medtronic document. As with all orthopaedic and neurosurgical implants, none of the CD HORIZON® Spinal System components should ever be reused under any circumstances.
The provided document is a 510(k) summary for the CD HORIZON® Spinal System, specifically for the addition of Threaded Rods. It details the device, its intended use, and the testing conducted to demonstrate substantial equivalence to predicate devices.
1. Table of Acceptance Criteria and Reported Device Performance:
| Test | Acceptance Criteria (from ASTM F1717) | Reported Device Performance |
|---|---|---|
| Construct Compression Static Bending | Per ASTM F1717 (2010), "Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model." This typically involves measuring the stiffness and strength of the construct under compressive bending loads. The acceptance criteria would be defined by comparison to the predicate device's performance, ensuring the new device is mechanically equivalent or superior. The specific numerical criteria are not provided in this summary but are implicit in meeting the ASTM standard. | Successfully met all acceptance criteria |
| Construct Static Torsion | Per ASTM F1717 (2010). This involves measuring the stiffness and strength of the construct under torsional loads. The acceptance criteria would be defined by comparison to the predicate device's performance, ensuring the new device is mechanically equivalent or superior. The specific numerical criteria are not provided in this summary but are implicit in meeting the ASTM standard. | Successfully met all acceptance criteria |
| Construct Compression Fatigue Bending | Per ASTM F1717 (2010). This involves subjecting the construct to cyclic compressive bending loads to assess its fatigue life and resistance to failure over time. The acceptance criteria would be defined by comparison to the predicate device's performance, ensuring the new device is mechanically equivalent or superior. The specific numerical criteria are not provided in this summary but are implicit in meeting the ASTM standard. | Successfully met all acceptance criteria |
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size: The document does not specify the exact sample size (number of constructs tested) for each non-clinical test. However, tests were performed "per ASTM F1717 (2010)," which outlines methodologies for such testing. Adherence to a recognized standard suggests appropriate sample sizes were used for mechanical testing to ensure statistical significance, though the exact numbers are not explicitly stated in this summary.
- Data Provenance: The data is from non-clinical (benchtop) testing of the device components. Therefore, there is no country of origin for patient data, nor is it retrospective or prospective in a clinical sense.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:
This section is not applicable as the study was non-clinical (benchtop mechanical testing) and did not involve human subjects or expert assessment of clinical data. The "ground truth" in this context is the physical performance of the device against predefined mechanical standards and comparison to predicate devices.
4. Adjudication Method for the Test Set:
This section is not applicable as the study was non-clinical (benchtop mechanical testing) and did not involve human experts making judgments on clinical cases. The "adjudication" was based on objective mechanical measurements and comparison against established criteria within the ASTM standard.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size:
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This study solely involved non-clinical benchtop testing of the device's mechanical properties.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:
This section is not applicable as the device is a medical implant (spinal system components), not an algorithm or software. The "standalone performance" refers to the mechanical performance of the physical device components.
7. The Type of Ground Truth Used:
The "ground truth" used was objective mechanical performance data derived from standardized laboratory tests (ASTM F1717) measuring properties such as static bending, static torsion, and fatigue bending. This data was compared against acceptance criteria presumably established from the performance of legally marketed predicate devices and the requirements of the ASTM standard.
8. The Sample Size for the Training Set:
This section is not applicable. There was no "training set" in the context of machine learning. The device underwent non-clinical mechanical testing.
9. How the Ground Truth for the Training Set Was Established:
This section is not applicable. There was no "training set" or "ground truth" in the context of an algorithm. The device's performance was evaluated against a recognized industry standard (ASTM F1717) and by comparison to predicate devices; the "ground truth" for the predicate device's performance would have been established similarly through non-clinical mechanical testing at the time of their 510(k) clearances.