510(k) Data Aggregation

    K Number
    K111437

    Validate with FDA (Live)

    Date Cleared
    2011-08-10

    (78 days)

    Product Code
    Regulation Number
    888.4540
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Synthes RIA System is intended for use in adults and adolescents (12-21 years) to clear the medullary canal of bone marrow and debris and to effectively size the medullary canal for the acceptance of an intramedullary implant or prosthesis, to harvest finely morselized autogenous bone and bone marrow for any surgical procedure requiring bone graft to facilitate fusion and/or fill bone defects, and to remove infected and necrotic bone and tissue from the medullary canal in the treatment of osteomyelitis.

    Device Description

    Synthes RIA System is a flexible intramedullary reaming and bone harvesting device with simultaneous irrigation and aspiration that consists of disposable tube assemblies, disposable reamer heads, disposable drive shaft seal, disposable locking clip, disposable graft filter and reusable drive shafts. The device is designed for expedited reaming of the medullary canal in preparation for internal fixation as well as the harvesting of bone and bone marrow, and as part of the treatment regimen for osteomyelitis.

    AI/ML Overview

    The provided text does not contain information about acceptance criteria or a study designed to prove the device meets specific acceptance criteria in the typical sense of a performance evaluation with metrics like sensitivity, specificity, etc. This document is a 510(k) summary for the Synthes RIA System, focused on establishing substantial equivalence to a predicate device and expanding its indications for use.

    Instead of a traditional performance study, the substantial equivalence assertion is supported by a literature review demonstrating existing clinical practice and outcomes related to the expanded indications.

    Therefore, many of the requested fields cannot be directly answered, as the nature of the submission is not a device performance study against predefined metrics.

    Here's a breakdown of what can be extracted based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    • Acceptance Criteria: Not explicitly defined in terms of performance metrics (e.g., sensitivity, specificity, accuracy). The acceptance criteria for this 510(k) revolve around demonstrating that the expanded indications for use (osteomyelitis treatment and adolescent use) are substantially equivalent to existing, legally marketed practices or devices.
    • Reported Device Performance: No quantitative performance metrics are reported. The "performance" is implicitly demonstrated through the cited literature, which describes successful clinical application of intramedullary reaming and bone harvesting in the contexts of osteomyelitis and adolescent patients.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not applicable in the context of a traditional performance study. The "test set" here refers to the collection of patients from various retrospective studies and case studies cited in the literature review.
      • For osteomyelitis: 18 patients (Lidgren L, Torholm C), 25 patients (Ochsner PE, Gosele A, Buess P), 32 patients (Pape H-Ch, et. al.), 2 patients (Ilyas AM, Mudgal CS), 1 patient (Bellapianta J, et. al.), 11 patients (Zalavras CG, Sing A, Patzakis MJ), 1 patient (Cobbs KF) - with some overlap as Pape and Lidgren studies also reference adolescents.
      • For adolescent use: 5 patients (Mehlman CT, Bishai SK), and "several" patients in other cited articles (Gordon JE, et. al.; Jencikova-Celerin L, et. al.; Keeler KA, et. al.), plus the adolescent patients from the Pape and Lidgren osteomyelitis studies.
    • Data Provenance: Predominantly retrospective analyses and case studies, as clearly stated in the descriptions of the cited literature. The countries of origin are not specified for most studies, but the authors' names suggest a mix of international research.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Number of Experts: Not applicable. Ground truth was not established by a panel of experts for a novel test set in this context. The "ground truth" is implied by the clinical outcomes reported in the independent literature, which would have been assessed by the clinicians (e.g., surgeons, radiologists) involved in those studies.
    • Qualifications of Experts: Not specified for the individuals who originally published the cited literature. Presumably, they are orthopedic surgeons or similar medical specialists based on the nature of the research.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. There was no formal adjudication process for a "test set" in this submission. The clinical findings and conclusions of the referenced peer-reviewed articles serve as the evidence.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    • MRMC Study: No, an MRMC comparative effectiveness study was not done. The documentation focuses on demonstrating substantial equivalence through literature review, not on comparing the device's diagnostic or treatment effectiveness against a standard or human readers with AI assistance.
    • Effect Size of Human Readers Improvement with AI vs. without AI: Not applicable, as no such study was performed or presented.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done

    • Standalone Performance Study: Not applicable. The Synthes RIA System is a manual surgical instrument, not an AI algorithm or a device requiring standalone performance metrics in this sense. Its effectiveness is intrinsically linked to its use by a human surgeon.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: Clinical outcomes and expert observations documented in published medical literature (retrospective analyses, case studies). This can be categorized as a form of "outcomes data" and "expert consensus" derived from independent clinical research. For osteomyelitis, the ground truth would be the diagnosis and successful treatment (e.g., infection eradication, bone healing) reported by the treating clinicians in the cited studies. For adolescent use, it refers to the successful application of intramedullary reaming in this age group as reported in the literature.

    8. The Sample Size for the Training Set

    • Training Set Sample Size: Not applicable. This device is not an AI/ML algorithm that requires a "training set." The submission relies on existing clinical literature to support its expanded indications, not on a data-driven model.

    9. How the Ground Truth for the Training Set was Established

    • Ground Truth for Training Set Establishment: Not applicable, as there is no training set for this type of device submission.