Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K111177

    Validate with FDA (Live)

    Device Name
    JUELL HARD CURE
    Manufacturer
    Date Cleared
    2011-09-12

    (138 days)

    Product Code
    Regulation Number
    872.3760
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    JuellCure Hard is intended for use as a permanent hard relining for total and partial dentures. Hard permanent total or partial relining for restoring partial and complete dentures. Lengthening denture margins.

    Device Description

    JuellCure hard is a cold-curing, hard relining material for permanently relining dentures.

    AI/ML Overview

    The provided text describes the 510(k) summary for the JuellCure Hard, a hard impression material for relining dentures. The submission aims to demonstrate substantial equivalence to a predicate device, Ufi Gel Hard C (K030916).

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance Criteria (Predicate Device K030916)Reported Device Performance (JuellCure Hard)
    Setting Time135-180 seconds135-180 seconds
    AppearancePaste is a colorless, homogenous liquid and the catalyst is a white homogenous paste.Paste is a colorless, homogenous liquid and the catalyst is a white homogenous paste.
    Composition of CatalystMatrix 40-60%, Filler 30-40%, Catalyst 2-4%, Modifier 6-7%, Stabilizer <0.1%Matrix 40-60%, Filler 30-40%, Catalyst 2-4%, Modifier 6-7%, Stabilizer <0.1%
    Composition of PasteMonomer: HEDMA, UDMA, Bis-GMA 30-50%, Filler 40-60%, Colorant <0.1%, Modifier 5-6%, Stabilizer <0.1%Monomer: HEDMA, UDMA, Bis-GMA 30-50%, Filler 40-60%, Colorant <0.1%, Modifier 5-6%, Stabilizer <0.1%
    Flexural Strength78 Mpa78 Mpa
    Translucency of 2mm specimens44%44%
    Adhesion to denture materialImplied to be similar to predicateImplied to be similar to predicate
    ThermocyclingImplied to be similar to predicateImplied to be similar to predicate
    Color stabilityImplied to be similar to predicateImplied to be similar to predicate
    Heat of polymerizationImplied to be similar to predicateImplied to be similar to predicate

    Note: For "Adhesion to denture material," "Thermocycling," "Color stability," and "Heat of polymerization," the document states: "The values for these performance characteristics was found to be very similar to the predicate device." While specific numerical acceptance criteria are not explicitly given for these, the "very similar" statement implies the JuellCure Hard met an unstated acceptance threshold based on predicate performance.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample size used for performance testing (e.g., for flexural strength, translucency, etc.) nor does it explicitly mention data provenance (e.g., country of origin, retrospective or prospective). The testing appears to be laboratory-based evaluation of material properties, not clinical data from patients.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The performance testing described relates to the physical and chemical properties of the material, which would be measured in a lab setting, not typically evaluated by clinical experts in the same way, for example, diagnostic imaging studies would be.

    4. Adjudication Method (e.g. 2+1, 3+1, none) for the Test Set

    The concept of an "adjudication method" (like 2+1 or 3+1 consensus) is primarily relevant for studies involving subjective human interpretation, such as medical image reading. For laboratory-based performance testing of a material, this method is not applicable and therefore not mentioned in the submission.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC study is relevant for evaluating the impact of AI on human reader performance in tasks like diagnostic interpretation. This submission is for a dental material (hard impression material), not an AI-powered diagnostic device. Therefore, an MRMC study was not conducted and is not applicable.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    This is not an AI-powered device. Therefore, a standalone (algorithm-only) performance evaluation was not done and is not applicable.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the performance testing mentioned (e.g., flexural strength, setting time, translucency) would be established through standardized material testing methods and measurements. For instance, flexural strength would be measured using an appropriate mechanical testing machine, and setting time would be determined according to an established dental materials standard. The comparison is made against the reported performance of the predicate device.

    8. The Sample Size for the Training Set

    This question is relevant for machine learning or AI-based devices. The JuellCure Hard is a material, not an AI device, so there is no training set in this context.

    9. How the Ground Truth for the Training Set Was Established

    As there is no training set, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1