Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K110696

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-07-07

    (115 days)

    Product Code
    Regulation Number
    874.1820
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Baxano Neuro Check™ Device is intended for use with an iO-Flex cutting and biting device for location of motor nerves in settings where visualization is compromised.

    Device Description

    The Neuro Check Device is designed to be pulled into the neural foramen by the iO-Flex Wire to enable the surgeon to direct a stimulus signal from a commercial EMG system to electrodes on the device enabling nerve root location prior to using a Baxano cutting and biting device. The Neuro Check Device is manufactured from biocompatible Pebax® tubing and stainless steel with a handle molded from ABS. The proximal end of the device handle has a pair of wires that connect to commercially available EMG systems.

    The iO-Flex Wire is manufactured from nickel-titanium wire and connects to the distal end of the Neuro Check Device for placement in the neural foramen. No changes were made to the previously cleared iO-Flex Wire design (K100958) other than packaging modifications to include it within the Neuro Check Device packaging for surgeon convenience.

    Since the previous clearance of the Neuro Check Device (K092729; October 2, 2009) changes were made to the physician labeling (the Instructions For Use) intended to provide compatibility with enhanced output stimulus signals of commercially-available intra-operative electromyography (EMG) neuromonitoring systems for motor nerve localization. Specifically, the maximum allowable EMG stimulation current amplitude to the Neuro Check Device was increased from 30 milliamps to 50 milliamps and the maximum current pulse width from 300 microseconds to 500 microseconds to support the stimulation parameters of commercially available EMG systems. There were no design modifications of the Neuro Check Device necessary to conform to the new electrical stimulation specifications.

    In addition, several design modifications were implemented since the last Neuro Check Device clearance. These modifications include integrating the separate Switch Box into the Neuro Check Device handle, eliminating the need for this external accessory. The handle itself is now molded from ABS plastic instead of being machined from the same material. Two radiopaque platinum-iridium marker bands were added to the distal flexible section to assist with fluoroscopic positioning of the device. A slight bend is made in the distal flexible segment to assist in orienting the Neuro Check Device as it is positioned with respect to the patient. Lastly, the iO-Flex Wire was added to the Neuro Check Device package for user convenience.

    AI/ML Overview

    This 510(k) summary describes a device modification, so the study primarily focuses on demonstrating that the modified device remains as safe and effective as its predicate, rather than establishing efficacy for a new device. Therefore, acceptance criteria generally relate to ensuring the device meets relevant safety and performance standards after modification.

    Here's an analysis of the provided information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state "acceptance criteria" in a quantified table with corresponding performance. Instead, it describes general non-clinical tests conducted to confirm the modified device meets design specifications and performance characteristics based on its intended use and complies with relevant electrical and safety standards. The performance is indirectly "reported" as compliance with these standards and equivalency to the predicate device.

    Acceptance Criteria (Implied)Reported (Implied) Device Performance
    Compliance with Mechanical Performance Specifications"Mechanical performance tests were conducted to verify that the device meets design specifications and performance characteristics, based upon the intended use." (Specific quantitative results are not provided in this summary).
    Electrical Safety and Compatibility to Standards"The modified Neuro Check Device was subjected to electrical safety and compatibility testing and was certified to following standards, including all applicable normative reference standards: IEC 60601-1: 1988 +A1:1991 +A2:1995 Medical Electrical Equipment Part 1: General Requirements for Safety; IEC 60601-2-40: 1998 Medical Electrical Equipment Part 2-40: Particular requirements for the safety of electromyographs and evoked response equipment; IEC 60601-1-2: 2001 + A1:2004 Medical Electrical Equipment Part 1-2: General Requirements for Safety Collateral Standard: Electromagnetic Compatibility."
    Sterility Assurance Level (SAL) of 10^-6"The sterilization process has been validated by an approved contract facility to provide a sterility assurance level (SAL) of 10^-6 in accordance with ISO 11137-1:2006 and ISO 11137-2:2006."
    Maintenance of Intended Use and Fundamental Scientific TechnologyThe document states: "The Neuro Check Device with iO-Flex Wire has the same indications for use and fundamental scientific technology as its predicate." This is demonstrated through the comparative characteristic tables (Table 2 and 3) highlighting no significant changes to core functional attributes.
    No new questions of safety or effectiveness"Based upon the indications for use, technological characteristics and performance test results, changes to the Neuro Check Device with iO-Flex Wire do not raise new questions of safety or effectiveness." This is a general conclusion based on all non-clinical testing and comparison to predicate devices.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: Not specified. The document states "Mechanical performance tests were conducted" and "The modified Neuro Check Device was subjected to electrical safety and compatibility testing," implying that specific units of the device were tested. However, the exact number of devices or test repetitions is not provided in this summary.
    • Data Provenance: The tests are described as "non-clinical tests" conducted on the modified device itself. This indicates laboratory testing, not human or observational data. The provenance is internal to the manufacturer or an approved contract facility for sterilization validation.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts:

    • Number of Experts: Not applicable/not specified. The "ground truth" for non-clinical performance and safety standards is the standard itself (e.g., IEC 60601-1). The validation of sterilization was done by an "approved contract facility." There's no mention of human expert adjudication for the non-clinical tests described.

    4. Adjudication Method for the Test Set:

    • Adjudication Method: Not applicable. For non-clinical tests like mechanical, electrical, and sterilization, the "adjudication" is typically against pre-defined engineering specifications and regulatory standards. There is no multi-reader or human consensus method involved for these types of tests.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • Was it done?: No, an MRMC comparative effectiveness study was not done or described. This submission is for a device modification (510(k)), which typically focuses on demonstrating substantial equivalence through non-clinical performance testing and comparison to predicates, rather than extensive clinical efficacy studies comparing human reader performance with and without AI assistance. The device is a "Nerve locator" which assists a surgeon, but this filing does not involve any AI component or human interpretation of images requiring an MRMC study.

    6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance):

    • Was it done?: Not applicable/not described. This device is not an algorithm, but a physical medical device. Its "performance" is primarily assessed through its physical, electrical, and sterilization characteristics in non-clinical settings, not through an algorithm's standalone performance. The device is intended to be used with a commercial EMG system and a surgeon ("human-in-the-loop"), so "standalone algorithm performance" (without human interaction) does not apply.

    7. Type of Ground Truth Used:

    • Ground Truth: For the non-clinical tests, the "ground truth" is adherence to established industry and regulatory standards (e.g., IEC 60601-1, IEC 60601-1-2, IEC 60601-2-40, ISO 11137-1:2006, ISO 11137-2:2006) and the device's own design specifications.

    8. Sample Size for the Training Set:

    • Sample Size: Not applicable. This document does not describe the development or testing of an AI algorithm, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established:

    • How Established: Not applicable, as there is no AI algorithm or training set discussed.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1