    K Number
    K110590

    Validate with FDA (Live)

    Date Cleared
    2012-04-13

    (408 days)

    Product Code
    Regulation Number
    892.5900
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Needle Applicator set (comprising the Needle Applicator and guide shaft) is indicated for use in combination with the INTRABEAM® System to intraoperatively administer radiation to tissue including irradiation of intracranial tumors.

    Device Description

    The proposed device is a modification of the INTRABEAM® System with Balloon Applicator, K090584, incorporating a Needle Applicator and guide shafts that are used to open a tract for the positioning of the applicator for irradiation of turnors. The INTRABEAM® System that will be used with the new Needle Applicator accessory was most recently described in K090584. There are no modifications to the INTRABEAM® System hardware or software. The INTRABEAM® Core System consists of the following components: PRS500 Control Console, XRS4 X-ray source, User terminal. Components of the Needle Applicator set are as follows: 1) Inner Sterile Packaging, 2) 5 cm guide shaft, 3) 6 cm guide shaft, 4) Needle Applicator. The physical design of the Needle Applicator consists of a polycarbonate case which fits over the probe of the XRS 4 X-ray source, a stainless steel shaft, and a polyetherimide tip for the Applicator. The guide shafts are stainless steel.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Carl Zeiss INTRABEAM® System with Needle Applicator. This document focuses on the non-clinical performance testing conducted to demonstrate substantial equivalence to predicate devices, rather than a study typically associated with AI/ML device approval which would include extensive clinical performance criteria and human reader studies.

    Therefore, many of the requested categories related to clinical study design, AI performance, ground truth establishment for a training set, and expert adjudication are not applicable or not explicitly detailed in the provided content. This document primarily addresses the safety and functional performance of the hardware modification (Needle Applicator) to an existing radiation therapy system.

    Here's an attempt to populate the requested table and information based on the provided text, indicating "Not Applicable" or "Not Specified" where the information is not present or not relevant to this type of submission.


    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance
    Sterilization ValidationCompliant with AAMI/ANSI/ISO 11137-1:2006 (development, validation, routine control), AAMI/ANSI/ISO 11137-2:2006 (establishing sterilization dose, VDmax 25 method), and AAMI/ANSI/ISO 11137-3:2006 (dosimetric aspects). Sterilization assurance level 10⁻⁶Validation performed; Dosimetric release complies. All criteria met.
    Performance Testing1. No adverse effect of sterilization and accelerated aging (equivalent to 5 years) on Needle Applicator functionality. 2. Dosimetry equivalence to the PRS400 with the XRS probe (with sheath). 3. Isotropy of tested units meets criteria.1. Five samples inspected visually before/after sterilization and accelerated aging. All applicators passed tests, were fully functional, and met engineering specifications. 2. Dose depth curves of Needle Applicators matched those of the bare probe with sheath. 3. Isotropy of tested units met the passing criteria. All criteria met.
    Biocompatibility Testing1. Cytotoxicity (L 929 cell cultures, quantitative proliferation). 2. Chemical Analysis (GC-FID, quantification of organic/inorganic leachables).1. Needle Applicator and Guide shaft: "No relevant effects observed in comparison to controls." 2. Needle Applicator: "No relevant effects observed in comparison to controls." All criteria met.
    Overall ConformityFulfills prospectively defined design and performance specifications as basis for substantial equivalence to predicate devices.Based on the non-clinical testing, the Carl Zeiss INTRABEAM® Needle Applicator is substantially equivalent to the cited predicate devices and fulfills prospectively defined design and performance specifications. This led to a substantial equivalence determination by the FDA in 2012. All key criteria met for regulatory approval based on non-clinical data.

    Study Details

    1. Sample size used for the Test Set and Data Provenance:

      • Sterilization Validation: Not explicitly stated for specific test runs, but the methodology (VDmax 25) implies a statistically relevant sample size for microbial testing.
      • Performance Testing (Functionality/Aging): 5 samples (Needle Applicators) were subjected to visual inspection, sterilization, and accelerated aging.
      • Performance Testing (Dosimetry): Not explicitly stated, but "tested units" implies multiple devices.
      • Biocompatibility Testing: Not explicitly stated, likely follows standardized protocols requiring specific sample sizes for each test type.
      • Data Provenance: The testing was conducted internally by Carl Zeiss Meditec AG, likely in Germany (country of manufacturer) or its affiliates. This is retrospective data for the submission, based on laboratory testing.
    2. Number of Experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable in the context of this submission. The "ground truth" here refers to objective measurements (e.g., sterilization levels, dose depth curves, cytotoxicity assays) against engineering specifications and industry standards, not expert clinical interpretation of medical images or patient data.
    3. Adjudication Method for the test set:

      • Not applicable. Performance was assessed against objective, predefined physical and chemical standards, not through adjudication of subjective interpretations.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This is not an AI/ML device, and no clinical studies involving human readers or comparative effectiveness were conducted for this submission ("No clinical testing was conducted to support this submission").
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is a hardware modification for a radiation therapy system, not an algorithm or AI.
    6. The type of ground truth used:

      • Objective physical and chemical measurements/standards:
        • Sterilization Assurance Level (SAL) of 10⁻⁶
        • Engineering specifications for functionality
        • Dosimetry measurements (dose depth curves, isotropy data) compared to bare probe with sheath
        • Standardized biocompatibility test results (cytotoxicity, chemical analysis) compared to controls.
    7. The sample size for the training set:

      • Not applicable. This is not an AI/ML device requiring a training set.
    8. How the ground truth for the training set was established:

      • Not applicable.

    Summary of Device and Study Context:

    This 510(k) submission (K110590) is for a hardware accessory (Needle Applicator) to an existing X-ray radiation therapy system (INTRABEAM® System). The purpose of the submission is to demonstrate that the new accessory is substantially equivalent to previously cleared predicate devices. The "study" described is a series of non-clinical, laboratory-based tests (sterilization, performance/aging, biocompatibility) designed to ensure the safety, functionality, and compatibility of the Needle Applicator with the existing system and regulatory standards. It does not involve patient data, clinical trials, or AI/ML components, hence many of the specific questions are not relevant in this context.