Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K110531

    Validate with FDA (Live)

    Device Name
    ENDOFLIP ECD
    Manufacturer
    Date Cleared
    2011-10-06

    (224 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EndoFLIP® EF-800 is an external channel for an endoscope (9.0 to 12.2 mm in diameter) used to aid in the insertion, advancement, and removal of endoscopic devices during endoscopic procedures. The EndoFLIP® ECD is single patient use, disposable. Environments of use - Hospitals, Sub-acute care institutions, Surgery Centers, doctor's offices where endoscopic procedures may be performed.

    Device Description

    The EndoFLIP® ECD EF-800 External Channel Device (ECD) is an endoscopic accessory designed to provide an additional channel external to the endoscope for inserting, advancing, and removing endoscopic devices thereby preserving the working channel of the endoscope for other instruments.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the EndoFLIP® ECD EF-800, structured as requested.

    The provided document is a 510(k) Summary, which is typically a premarket notification to the FDA to demonstrate that a device is substantially equivalent to a legally marketed predicate device. In many cases, for devices of this type (an accessory to an endoscope), the "acceptance criteria" are demonstrating equivalence in technical characteristics and performance to predicate devices, rather than establishing new clinical thresholds. The "study" often refers to non-clinical testing.


    1. Table of Acceptance Criteria and Reported Device Performance

    Given the nature of the K110531 submission, the "acceptance criteria" are implied by the comparison to predicate devices, focusing on demonstrating similar functional, material, and safety characteristics. The performance reported is primarily through compatibility and age testing to ensure the device performs as expected for its intended use.

    Acceptance Criteria (Implied from Predicate Comparison)Reported Device Performance (EndoFLIP® ECD EF-800)
    Indications for Use EquivalenceIdentified as equivalent to predicate K100081 (US Endoscopy overtube) for aiding insertion, advancement, and removal of endoscopic devices during procedures in the upper GI tract, including small intestine.
    Technology SimilaritySimilar to predicate K060923 (Smart Medical NaviAID BGE) – a simple double lumen tube with a tapered tip. Note: without the balloon accessory present in the predicate.
    Material BiocompatibilityMaterials in contact were tested to ISO 10993-1 (Cytotoxicity, Irritation, and Sensitization).
    Environment of Use EquivalenceIdentical to predicate K060923 (Smart Medical NaviAID BGE) – Hospitals, Sub-acute care institutions, Surgery Centers, doctor's offices.
    Patient Population EquivalenceIdentical to predicate K060923 (Smart Medical NaviAID BGE) – Patients undergoing endoscopic procedures.
    Shelf LifeAge testing performed. (Details of specific duration or passing criteria are not provided in this summary.)
    Compatibility with EndoscopesCompatibility testing performed to verify its ability to slide over endoscopes. (Details of specific compatibility parameters are not provided.)
    Safety and Efficacy ConcernsThe device "raises no new safety or efficacy concerns" compared to predicate devices.

    2. Sample Size Used for the Test Set and Data Provenance

    For the "Performance Testing" mentioned (Age testing and Compatibility testing), the document does not specify:

    • The exact sample size used for these tests.
    • The origin of data (e.g., country of origin).
    • Whether the data was retrospective or prospective.

    These tests are typically bench or lab-based rather than clinical studies with human participants.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information (number and qualifications of experts for ground truth) is typically relevant for studies involving subjective assessment (e.g., image interpretation). For a device like an endoscopic accessory undergoing performance testing (age testing, compatibility testing), the "ground truth" would be established by objective measurements and standardized test protocols, not by expert consensus on, for example, image interpretation. Therefore, this information is not applicable in the context of the provided regulatory submission.


    4. Adjudication Method for the Test Set

    As the performance testing described is objective (age testing, compatibility with endoscopes), there is no mention of an adjudication method (like 2+1 or 3+1). Such methods are typically used when subjective expert assessments need to be reconciled, which is not the case here.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. MRMC studies are typically clinical trials designed to compare the diagnostic performance of different methods (e.g., AI vs. human readers, or human readers with vs. without AI assistance) across multiple readers and cases. This submission focuses on demonstrating substantial equivalence of a physical medical device accessory through comparison of technical specifications and non-clinical performance tests.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, this is not applicable. The EndoFLIP® ECD EF-800 is a physical medical device accessory (an external channel for an endoscope), not a diagnostic algorithm or an AI-powered system. Therefore, the concept of "standalone algorithm performance" (without human-in-the-loop) does not apply.


    7. The Type of Ground Truth Used

    For the performance tests mentioned:

    • Age testing: The "ground truth" would be the stability and integrity of the device materials and function over a specified period, verified by empirical observation/measurement (e.g., material degradation, functional checks after accelerated aging) against predefined engineering specifications.
    • Compatibility testing: The "ground truth" would be the successful and unimpeded sliding of the device over various endoscope diameters, verified by direct observation/measurement (e.g., successful navigation, absence of damage to either device) against specified dimensional and functional tolerances.

    This is based on engineering specifications and objective physical measurements, not expert consensus, pathology, or outcomes data in a clinical sense.


    8. The Sample Size for the Training Set

    Not applicable. This device is a physical medical accessory, not an AI or machine learning model that requires a "training set."


    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As stated above, this device does not utilize a training set in the context of machine learning.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1