510(k) Data Aggregation

    K Number
    K110453

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-03-16

    (28 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reliance Endoscope Processing System is intended for washing and high level disinfection of up to two manually pre-cleaned, immersible, reusable, heat-sensitive, semi-critical devices such as GI flexible endoscopes, bronchoscopes and their accessories. High level disinfection is achieved within the 50 - 57℃ HLD Phase of the Endoscope Processing Cycle (4 minute generation sequence followed by a 6-minute exposure sequence).

    Device Description

    The Reliance Endoscope Processing System is an economical, easy-to-use high level disinfection system intended to wash and high level disinfect up to two manually cleaned, immersible, reusable, heat-sensitive, semi-critical devices such as GI flexible endoscopes and related accessories. The system utilizes Reliance™ DG Dry Germicide, a proprietary, safe, and dry peracetic acid generating oxidative chemistry. The Reliance Endoscope Processing System was designed to be versatile in meeting the growing demands of the modern flexible endoscope processing department, while offering the highest level of patient and staff safety. The Reliance Endoscope Processing System is a combination of products that are used to wash and high level disinfect flexible endoscopes and their accessories.

    • The Reliance Endoscope Processor is an electromechanical washer/high level disinfector with a microprocessor-based controller that provides for automated endoscope processing cycles and processor self-decontamination cycles.
    • Reliance DG Dry Germicide is a proprietary, two-part, dry, single-use oxidative chemistry, designed to generate the high level disinfection solution upon automatic dilution in water within the Reliance Endoscope Processor.
    • Optional washing is provided through the automated delivery of Klenzyme Enzymatic Presoak and Cleaner during the wash phase of the cycle.
    • CIP 200 Acid-Based Process and Research Cleaner, a general cleaning agent, is used in one of the two self-decontamination cycles provided by the processor.
    • Various accessories are available to accommodate the processing needs of specific endoscopes and endoscopic accessories.
    • Verify Reliance CI Process Indicator is available to monitor for the presence of the Reliance DG active ingredient, peracetic acid.
    AI/ML Overview

    The provided text describes the STERIS® Reliance® Endoscope Processing System, an automated endoscope reprocessor. The document is a 510(k) summary submitted to the FDA, demonstrating substantial equivalence to a predicate device and providing performance data.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the success of the tests conducted according to FDA guidance documents for germicidal efficacy, simulated-use, and in-use studies. The reported device performance indicates that these criteria were met.

    Acceptance Criterion (Implicit)Reported Device Performance
    Germicide Efficacy (Reliance DG Dry Germicide):
    Sporicidal activity (defined by AOAC Sporicidal Activity Test with 6 minutes exposure).Proven sporicidal in situ. Confirmatory testing completed, supplemental confirmatory testing in the processor with aged germicide.
    Tuberculocidal activity (defined by AOAC Tuberculocidal Activity Test with 6 minutes exposure).Proven tuberculocidal. Potency confirmed with aged germicide.
    Virucidal activity (reduction of poliovirus Type 1, adenovirus Type 5, and herpes simplex virus Type 1 by > 4 log₁₀).Proven to reduce viable population by > 4 log₁₀.
    Bactericidal activity (defined by AOAC Bactericidal Activity Test with 6 minutes exposure at worst case conditions).Proven bactericidal in situ or in vitro.
    Fungicidal activity (defined by AOAC Fungicidal Activity Test with 6 minutes exposure).Proven fungicidal in situ or in vitro.
    Simulated-Use Study (High-Level Disinfection performance within the processor):
    Greater than a 6 log₁₀ reduction of Mycobacterium terrae within the Reliance Endoscope Processor.Reproducibly achieved greater than a 6 log₁₀ reduction of Mycobacterium terrae in triplicate trials using selected clinically relevant flexible endoscopes and accessories (representing most challenging devices).
    In-Use Study (Clinical effectiveness after manual cleaning and high-level disinfection):
    No organisms recovered after processing clinically used endoscopes.In triplicate evaluations of each of three flexible endoscopes (representing the product labeling range) used in clinical procedures, no organisms were recovered after processing. Bioburden levels prior to processing were as high as 10⁵ CFU/device.
    Biocompatibility (Residues on medical devices):
    Residues remaining on medical endoscopes and accessories below established residue limits and non-cytotoxic.Residues remaining on medical endoscopes and accessories are below established residue limits and do not pose a risk. Biocompatibility testing of extracts from processed devices demonstrated no toxic residuals. Processor final rinse water was found to be non-cytotoxic.
    Material Compatibility:
    No deleterious effects or functional changes on medical devices (e.g., flexible endoscopes) after repeated processing cycles (e.g., 300 cycles).After 300 processing cycles, no deleterious effects were observed other than minor cosmetic changes. No functional changes in flexible endoscopes were observed.
    Processor Performance (Critical Process Parameters):
    Water temperature and volume, fresh Reliance DG Dry Germicide container detection, boot pressure, delivery of washing solutions and high level disinfection solution, and water filter integrity testing within required specifications under worst-case conditions.Each critical parameter was evaluated in replicate under worst case conditions and found to be within required specifications.
    Washing phase effectiveness: Devices pre-soiled with combination of eggs, blood, mucin, and serum in saline are visually clean and achieve greatly reduced yield of extractable protein per cm² device area (e.g., reduced from >173 to < 5 μg/cm²).Documented that after the shortest possible washing phase, devices were visually clean and achieved greatly reduced yield of extractable protein per cm² device area (reduced from >173 to < 5 μg/cm²).
    Rinse phase effectiveness: Levels of residuals remaining on devices below allowable limits and non-cytotoxic.Evaluated extracts of devices exposed to worst case conditions in the Endoscope Processing Cycle documented that levels of residuals remaining on devices were far below allowable limits and were not cytotoxic.
    Air purge phase effectiveness: Ability to remove rinse water from processed medical devices.Validated to confirm the ability to remove rinse water from processed medical devices.
    Filter integrity test system reliability: Reliably detect filter failure.Documented to reliably detect filter failure.
    D-LONG cycle effectiveness: Disinfect the processor after a high-level challenge with Pseudomonas aeruginosa followed by 5 days inactive period.Proven effective to disinfect the Reliance Endoscope Processor after a high level challenge with Pseudomonas aeruginosa followed by a 5 day inactive period.
    D-SHORT cycle effectiveness: Kill bacteria that have the potential to form biofilm.Proven effective to kill bacteria that have potential to form biofilm.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Germicide Efficacy Tests (Sporicidal, Tuberculocidal, Virucidal, Bactericidal, Fungicidal): "Standard array of microbiological tests" and "triplicate trials" are mentioned for simulated-use. Specific sample sizes for each in vitro test are not explicitly quantified in the provided text beyond these general statements.

    • Simulated-Use Study: "Triplicate trials" for Mycobacterium terrae reduction. No specific number of endoscopes or accessories is explicitly stated, other than "selected clinically relevant flexible endoscopes and their accessories" and "represented the range of most challenging devices."

    • In-Use Study: "Three flexible endoscopes representing the range of types indicated in the product labeling" were used. "In triplicate evaluations of each endoscope." This indicates a total of 9 clinical processing events (3 endoscopes x 3 evaluations).

    • Material Compatibility: "After 300 processing cycles."

    • Processor Performance: "Each critical process parameter... evaluated in replicate." Specific number of replicates not provided beyond "replicate."

    • Data Provenance:

      • In-Use Study: Conducted in a "US hospital" (prospective, clinical).
      • All other tests (Germicide Efficacy, Simulated-Use, Biocompatibility, Material Compatibility, Processor Performance, Stability): These appear to be laboratory-based studies (likely prospective, conducted by the manufacturer or contract labs). The text does not specify external data sources or retrospective analysis for primary efficacy/safety testing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The provided text does not mention the use of human experts to establish ground truth for the test sets (i.e., for determining sterility, log reduction, etc.). The ground truth for these studies is based on quantitative microbiological assays and analytical chemistry methods, which are objective and do not require expert consensus for interpretation.

    4. Adjudication Method for the Test Set

    Not applicable. The ground truth is objective (microbiological tests, chemical analysis, visual inspection for cleanliness, etc.), rather than expert-derived or needing adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an automated endoscope reprocessor, not an AI-assisted diagnostic device that would involve human readers interpreting images.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the primary studies described (Germicide Efficacy, Simulated-Use, Processor Performance, Material Compatibility) are all "standalone" in the sense that they evaluate the device's performance directly, independent of a human operator's judgment in the final assessment of disinfection efficacy or product integrity. The "in-use" study has a manual cleaning step by a human prior to automated processing, but the outcome assessment (microbiological culture) is objective and independent of human interpretation.

    7. The Type of Ground Truth Used

    The ground truth used in these studies is primarily objective laboratory measurements and outcomes data:

    • Microbiological Culture and Assays: For sporicidal, tuberculocidal, virucidal, bactericidal, fungicidal, simulated-use (Mycobacterium terrae reduction), and in-use studies (recovery of organisms).
    • Chemical Analysis: For germicide concentration, pH, chelation capacity, and detection of residues on devices.
    • Visual Inspection: For visual cleanliness after the washing phase.
    • Protein Assays: For measuring extractable protein per cm² device area.
    • Functional Testing: For assessing functional changes in endoscopes after repeated processing.
    • Measurements against Specifications: For critical process parameters (temperature, volume, pressure, filter integrity).

    8. The Sample Size for the Training Set

    Not applicable. This device is an electromechanical system with chemical efficacy, not a machine learning or AI-based system that requires a "training set" in the context of algorithm development. Its performance is based on physical and chemical principles and verified through empirical testing.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set (in the context of AI/ML) was used.