Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K103775

    Validate with FDA (Live)

    Date Cleared
    2011-05-27

    (151 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Suction Coagulator is a single use electrode attachment, designed to be attached to the PEAK PlasmaBlade TnA handpiece and is intended for use in surgical procedures, such as general and otolaryngology (ENT) surgery procedures, where the coagulation of tissue and suction of fluids are desired. It is not indicated for the removal of tonsils and adenoids.

    Device Description

    The PEAK Suction Coagulator consists of of an active suction lumen, a bendable shaft, and a finger grip for attachment to the TnA handpiece and is to be used for the coagulation of tissues and aspiration of fluids during electrosurgical procedures. The device is activated by pressing the coag button (blue) on the TnA handpiece.

    AI/ML Overview

    The provided text describes the 510(k) submission for the PEAK Suction Coagulator. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study of the device's performance against specific acceptance criteria in the manner typically seen for novel diagnostic or AI-driven devices.

    Therefore, the information required to populate most of the acceptance criteria and study details (like sample size for test/training sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, etc.) is not available in the provided document.

    Instead, the submission relies on:

    • Comparison to a predicate device (Valleylab Suction Coagulator): The primary claim is that the PEAK Suction Coagulator is "substantially equivalent" due to similar output energy, delivery system, blade specifications, and the use of monopolar electrosurgery for tissue coagulation.
    • Compliance with recognized standards: The device met various IEC and ISO standards related to electrical safety, electromagnetic compatibility, surgical equipment safety, sterilization, and biocompatibility.
    • Histological studies: These were conducted to compare the thermal effects of the PEAK Suction Coagulator to the predicate device, aiming to show similar safety and effectiveness. The specific acceptance criteria for these histological studies are not detailed, nor are the number of samples, ground truth establishment, or expert involvement.

    Here's a breakdown of what can be gathered from the document, and what is missing:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (as inferred/stated)Reported Device Performance
    Substantial EquivalenceSimilar output energySimilar to predicate device
    Similar delivery systemSimilar to predicate device
    Similar blade specificationsSimilar to predicate device
    Monopolar electrosurgical deviceSimilar to predicate device
    RF powered distal endsSimilar to predicate device
    Safety (Thermal Effects)Thermal effects comparable to predicate deviceHistological studies demonstrated substantial equivalence to predicate device.
    FunctionalityFunction as intended (coagulation of tissue, aspiration of fluids)Laboratory and performance tests executed to ensure function as intended.
    Electromedical StandardsIEC 60601-1 (General Safety)Complies with standard
    IEC 60601-1-2 (EMC)Complies with standard
    IEC 60601-2-2 (HF Surgical Eq.)Complies with standard
    SterilizationISO 11135-1 (EO Sterilization)Complies with standard
    BiocompatibilityISO 10993-1 (Guidance on Tests)Complies with standard
    ISO 10993-7 (EO Residuals)Complies with standard

    2. Sample Size for the Test Set and Data Provenance

    • Sample Size: Not specified for the histological studies or laboratory/performance tests.
    • Data Provenance: Implied to be laboratory tests and histological studies, conducted for this regulatory submission. No information on country of origin or whether it's retrospective/prospective.

    3. Number of Experts Used to Establish Ground Truth & Qualifications

    • Number of Experts: Not specified.
    • Qualifications: Not specified.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not specified. Given the nature of histological comparisons, it would typically involve expert pathologist review, but the method isn't described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • MRMC Study: No, this type of study is not mentioned. The device is a surgical instrument, not a diagnostic imaging AI, so an MRMC study comparing human readers with and without AI assistance is not applicable here. The comparison is between the device and a predicate device/standards.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • Standalone Performance: Not applicable in the context of an electrosurgical device. The device itself is a tool used by a human surgeon. Its performance is evaluated through its physical and electrical characteristics, and its effects on tissue.

    7. The Type of Ground Truth Used

    • For the "Safety (Thermal Effects)" criterion, the ground truth was likely derived from histological examination by qualified personnel (e.g., pathologists) comparing tissue effects between the PEAK Suction Coagulator and the predicate device.
    • For other criteria, the "ground truth" would be established by compliance with published international standards (IEC, ISO) and internal design specifications.

    8. The Sample Size for the Training Set

    • Sample Size: Not applicable. This document describes a medical device requiring regulatory clearance based on substantial equivalence, not an AI/ML algorithm that undergoes "training."

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth Establishment: Not applicable, as there is no "training set" in this context.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1