Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K103540

    Validate with FDA (Live)

    Device Name
    PSORIA-LIGHT
    Manufacturer
    Date Cleared
    2011-02-10

    (70 days)

    Product Code
    Regulation Number
    878.4630
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Psoria-Light Model PS1000 is indicated for use in targeted PUVA photochemistry and UVB phototherapy for the treatment of skin conditions including psoriasis, vitiligo, atopic dermatitis (eczema), and seborrheic dermatitis. In addition, the system UVB channel is indicated for the treatment of leukoderma.

    Device Description

    Psoria-Light™ Model PS1000 delivers targeted ultraviolet (UV) light to small regions of affected skin (2.88 cm), sparring healthy skin. The system is capable of delivering either UVB light (300 to 320nm) or UVA light (350 to 395nm). Psoria-Light™ Model PS1000 generates both UVA and UVB light via LED technology, avoiding environmental concerns associated with disposal and exchange of mercury gas lamps, Xenon, and Chlorine gases commonly associated with targeted UV phototherapy devices. Psoria-Light's™ UVB LEDs emit light between 300 and 320 nm, avoiding light emissions below 300nm. Psoria-Light™ Model PS1000 incorporates a unique safety feature, a non-contact sensor located about the aperture of the handpiece that detects human skin. The trigger mechanism for UV dosage administration (located on the handpiece) is disabled until a suitable surface is detected by this sensor.

    AI/ML Overview

    The Psoria-Light Model PS1000 is a targeted UVA/UVB phototherapy system for dermatologic disorders. The provided document is a 510(k) summary for the device, focusing on its substantial equivalence to previously cleared predicate devices rather than new clinical effectiveness studies.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Clinical Performance:
    New Indications for UseNo new indications for use introduced by Psoria-Light Model PS1000. All indications (psoriasis, vitiligo, atopic dermatitis (eczema), seborrheic dermatitis, and leukoderma for UVB) have been previously cleared for predicate devices.
    Product Performance (Non-Clinical):
    Conformance to IEC 60601 series of standardsThe Psoria-Light Model PS1000 shall have completed all testing to these standards prior to sale.
    Conformance to 21 CFR Part 1040.10 (Performance Standard for Light-Emitting Products)The Psoria-Light Model PS1000 shall have completed all testing to these standards prior to sale.
    Certification to UL 60601-1 Medical Electrical Equipment classification standardThe Psoria-Light Model PS1000 shall have completed all testing to these standards prior to sale.
    Compliance with EMC requirements (FCC rules part 15, IEC 60601-1-2, EN55011)The Psoria-Light Model PS1000 shall have completed all testing to these standards prior to sale.
    Application of risk management per ISO 14971The Psoria-Light Model PS1000 has been reviewed for risk management utilizing ISO 14971, ensuring all aspects are reviewed for potential hazards.
    Manufacturing in an FDA-registered facilityProducts shall be produced and distributed by a facility that has been registered with the FDA to manufacture medical devices.
    Substantial Equivalence:
    Equivalent design, composition, and function to predicate devicesThe characteristics of the Psoria-Shield Psoria-Light Model PS1000 are substantially equivalent to those of the predicate devices (TheraLight Inc. DuaLight UV120-2, PhotoMedex, Inc. XTRAC Ultra2 Excimer Laser System).
    Equivalent output wavelengths, power, and light sources to predicate devicesOperation of the predicate devices, including output wavelengths, power, and light sources, received 510(k) clearance previously. The Psoria-Light Model PS1000 performs in a substantially equivalent manner.

    2. Sample Size Used for the Test Set and Data Provenance

    No clinical performance test set (i.e., patient data) was used for this 510(k) submission. The submission explicitly states that "duplicative clinical data is not required as a condition of granting market clearance" because the device does not introduce new indications for use and performs in a substantially equivalent manner to predicate devices. The "study" for acceptance criteria was based on demonstrating substantial equivalence to these previously cleared devices through non-clinical performance testing and comparison of device specifications.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. No clinical test set requiring expert ground truth was employed for this 510(k) submission. The device's safety and effectiveness were argued based on substantial equivalence to predicate devices, which had presumably undergone such evaluations during their own clearance processes.

    4. Adjudication Method for the Test Set

    Not applicable. No clinical test set requiring adjudication was employed.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance

    Not applicable. This device is a phototherapy system, not an AI-assisted diagnostic or interpretive tool. Therefore, an MRMC study related to human reader performance with or without AI assistance is not relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This device is a hardware phototherapy system, not an algorithm. Its performance is evaluated based on its physical characteristics, light output, and compliance with safety and electrical standards.

    7. The Type of Ground Truth Used

    For the non-clinical product performance testing, the "ground truth" would be established by the specifications and requirements of the referenced international and national standards (e.g., IEC 60601, 21 CFR Part 1040.10, UL 60601-1, ISO 14971, FCC rules). The device's performance is measured against these established engineering and safety benchmarks. For the clinical indications, the "ground truth" was established by the previous 510(k) clearances of the predicate devices, which had demonstrated clinical effectiveness for the stated indications.

    8. The Sample Size for the Training Set

    Not applicable. This device is a hardware phototherapy system and does not involve AI or machine learning that would require a 'training set.'

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. No training set was used.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1