510(k) Data Aggregation

    K Number
    K102867

    Validate with FDA (Live)

    Device Name
    IROOT FS
    Date Cleared
    2010-12-03

    (64 days)

    Product Code
    Regulation Number
    872.3820
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    • Repair of Root Perforation
    • Repair of Root Resorption
    • Root End Filling
    • Apexification
    • Pulp Capping
    Device Description

    iRoot FS Fast Set Root Repair Material (iRoot FS) is a convenient ready-to-use fast setting white hydraulic premixed bioceramic paste developed for permanent root canal repair of root perforation and root resorption, and root end filling, apexification and pulp capping applications. iRoot FS is an insoluble, radiopaque and aluminum-free material based on a calcium silicate composition, which requires the presence of water to set and harden. iRoot FS does not shrink during setting and demonstrates excellent physical properties. iRoot FS is available as a preloaded syringe with disposable tips and a preloaded container.

    AI/ML Overview

    This document is a 510(k) summary for the iRoot FS device, a root canal repair material. The submission aims to demonstrate substantial equivalence to predicate devices, namely iRoot BP, iRoot BP Plus, iRoot SP, and D-PBS.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    This submission does not provide explicit numerical "acceptance criteria" for clinical performance. Instead, it focuses on demonstrating substantial equivalence in chemical and physical properties, and biocompatibility to existing predicate devices.

    Acceptance Criteria (Implied by Substantial Equivalence Claim)Reported Device Performance (vs. Predicate iRoot BP)
    FlowabilityComparable
    Working TimesComparable
    Setting TimesComparable
    Dimensional Change Following SettingComparable
    SolubilityComparable
    Shelf LifeComparable
    Biocompatibility (General)Comparable
    Biocompatibility (Non-cytotoxicity)Non-cytotoxic
    Biocompatibility (Non-mutagenic)Biocompatibility test data for iRoot BP (which has principal chemical components in common with iRoot FS) provides evidence that iRoot FS is non-mutagenic.
    Biocompatibility (Non-allergenic potential after multiple uses)Biocompatibility test data for iRoot BP provides evidence that iRoot FS does not cause an allergenic potential after multiple uses.
    Biocompatibility (Good tolerance by subcutaneous tissue)Biocompatibility test data for iRoot BP provides evidence that iRoot FS has a good tolerance by subcutaneous tissue.
    Equivalent Indications for UseiRoot FS has the equivalent indications for use as iRoot BP: Repair of Root Perforation, Repair of Root Resorption, Root End Filling, Apexification, Pulp Capping.
    Comparable Chemical CompositioniRoot FS is a modification of iRoot BP. Additional chemical components found in iRoot FS were found to be safe and effective in iRoot SP and D-PBS. (Implies overall composition is comparable and safe).
    Comparable Physical PropertiesiRoot FS and iRoot BP have comparable physical properties.
    Comparable Performance SpecificationsiRoot FS and iRoot BP have comparable performance specifications.
    Comparable Packaging Containers and Delivery SystemiRoot FS and iRoot BP Plus have equivalent device packaging containers and delivery systems.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not describe a "test set" in the context of a clinical trial or a diagnostic device where a sample size of patient data would be relevant. The testing performed was "shelf life, bench and biocompatibility testing."

    • Sample Size for Test Set: Not applicable in the context of clinical data for a diagnostic device. For bench testing, typical laboratory sample sizes would be used, but these are not specified in the summary.
    • Data Provenance: The nature of the bench and biocompatibility testing means the data is likely generated in a laboratory setting, not from patient populations. No country of origin for such data is stated, nor is it specified if it's retrospective or prospective, as these terms usually apply to clinical studies.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided as the submission focuses on non-clinical testing and substantial equivalence, not clinical performance or diagnostic accuracy validated by experts.

    4. Adjudication Method for the Test Set

    This information is not provided as there is no mention of a diagnostic test set requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for diagnostic imaging devices to evaluate human reader performance with and without AI assistance. This submission describes a dental material, not a diagnostic device.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    No, a standalone algorithm performance study was not done. The device itself is a material for dental procedures, not an algorithm.

    7. The Type of Ground Truth Used

    For the "test set" (bench and biocompatibility testing), the "ground truth" would be the established scientific and regulatory standards/methods for evaluating chemical properties, physical properties (e.g., flow, setting time, solubility, dimensional change), and biocompatibility (e.g., cytotoxicity, mutagenicity, sensitization, tissue irritation). These are determined through standardized laboratory protocols and scientific literature, not clinical ground truth like pathology or outcomes data in the typical sense for a diagnostic device.

    8. The Sample Size for the Training Set

    Not applicable. The iRoot FS is a physical dental material, not an AI algorithm, so there is no "training set."

    9. How the Ground Truth for the Training Set was Established

    Not applicable. As there is no AI algorithm, there is no training set or associated ground truth establishment process.


    In summary, this 510(k) submission for iRoot FS focuses entirely on demonstrating substantial equivalence to predicate devices through non-clinical (shelf life, bench, and biocompatibility) testing, rather than reporting on clinical performance or diagnostic accuracy assessed via expert review or human-in-the-loop studies. Therefore, most of the questions related to clinical study design, expert involvement, and AI performance are not applicable to the information provided in this document.