Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K102764

    Validate with FDA (Live)

    Date Cleared
    2010-12-22

    (89 days)

    Product Code
    Regulation Number
    876.4400
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Resolution II Clip is indicated for clip placement within the Gastrointestinal Tract for the purpose of:

    • Endoscopic Marking
    • Hemostasis for:
    • Mucosal/sub-mucosal defects < 3 cm
    • Bleeding ulcers
    • Arteries < 2 mm
    • Polyps < 1.5 cm in diameter
    • Diverticula in the colon
    • Anchoring to affix jejunal feeding tubes to the wall of the small bowel
    • As a supplementary method for closure of GI tract luminal perforations <20mm that can be treated conservatively
    Device Description

    The Resolution II Clip consists of a pre-loaded, radiopaque, single-use clip on a flexible delivery system. The Resolution II Clip is designed to be compatible with Gastroscopes with working channels ≥2.8, and/or Duodenoscopes and Colonoscopes with working channels ≥3.2mm. The Resolution II Clip jaws are engineered to open and close no more than five times, aiding in repositioning of the clip at the lesion site. Opening and closing capability may be limited by clinical circumstances and patient anatomy, among other factors. The Resolution II Clip will be offered in 155cm and 235cm lengths. The deployed clip is constructed of stainless steel while the delivery system is constructed of stainless steel, polyethylene, and polyester materials.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the Resolution II Clip, based on the provided documents:

    Acceptance Criteria and Device Performance for Resolution II Clip

    1. Table of Acceptance Criteria and Reported Device Performance

    Note: The provided documents do not explicitly list quantitative acceptance criteria in a pass/fail format with specific thresholds. Instead, they state that "All device bench test results were acceptable" and "All biocompatibility tests... passed." The table below reflects the types of tests performed and the general findings.

    Test CategorySpecific Tests PerformedAcceptance Criteria (Implicit)Reported Device Performance
    BiocompatibilityCytotoxicity, Sensitization, Intracutaneous Reactivity, Subacute (Subchronic) Toxicity (Intraperitoneal and Intravenous), Genotoxicity (Ames Assay and Mouse Lymphoma), Implantation, USP PhysicochemicalDevice materials must not cause adverse biological reactions or toxicity as per EN ISO 10993-1:2009.All biocompatibility tests passed; device considered biocompatible for intended use.
    Mechanical/PhysicalForces, Tensile Strengths, Corrosion Resistance, Dimensional, Endoscope CompatibilityDevice must meet design specifications for mechanical integrity, material properties, and compatibility with specified endoscopes as per ISO 14630:2008.All device bench test results were acceptable; sufficiently meets design specifications.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document describes bench testing and biocompatibility testing rather than a clinical study with a "test set" in the context of human data. Therefore, information on sample size for a test set and data provenance (country of origin, retrospective/prospective) is not applicable in this context. The tests were performed on the device itself and its components.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable as the evaluation was based on bench testing and biocompatibility testing of the device, not on diagnostic accuracy requiring human expert review of clinical cases.

    4. Adjudication Method for the Test Set

    This information is not applicable as the evaluation was based on bench testing and biocompatibility testing of the device, not on diagnostic accuracy requiring human expert review of clinical cases.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. The submission primarily focuses on demonstrating substantial equivalence to a predicate device through:

    • Same intended use.
    • Similar technological characteristics (though with fewer deployment steps).
    • Bench testing and biocompatibility testing.

    The document does not describe any human reader studies or evaluate the improvement of human readers with or without AI assistance, as AI is not a component of this device.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    No, a standalone performance study was not done. The Resolution II Clip is a medical device (endoscopic clipping device), not an algorithm or AI system. Therefore, the concept of "standalone performance" for an algorithm does not apply. The device's performance was evaluated through bench testing and biocompatibility assessments.

    7. The Type of Ground Truth Used

    The "ground truth" for the device's performance was established by engineering design specifications, industry standards (EN ISO 10993-1:2009, ISO 14630:2008), and material properties. The tests were designed to confirm that the device met these predefined physical, mechanical, and biocompatibility requirements.

    8. The Sample Size for the Training Set

    This information is not applicable. The Resolution II Clip is a physical medical device, not an AI or machine learning model that requires a "training set" of data.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1