510(k) Data Aggregation

    K Number
    K102555

    Validate with FDA (Live)

    Date Cleared
    2010-11-17

    (71 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CD HORIZON® Spinal System with or without SEXTANT® instrumentation is intended for posterior, non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudarthrosis; and/or failed previous fusion.

    Except for hooks, when used as an anterolateral thoracic/lumbar system, the CD HORIZON® Spinal System may also be used for the same indications as an adjunct to fusion.

    With the exception of degenerative disc disease, the CD HORIZON® LEGACY™ 3.5mm rods and the CD HORIZON® Spinal System PEEK rods and associated components may be used for the aforementioned indications in skeletally mature patients as an adjunct to fusion. The 3.5mm rods may be used for the specific pediatric indications noted below.

    When used for posterior non-cervical pedicle screw fixation in pediatric patients, the CD HORIZON® Spinal System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The CD HORIZON® Pediatric Spinal System is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

    The CD HORIZON SPIRE™ Plate is a posterior, single level, non-pedicle supplemental fixation device intended for use in the non-cervical spine (TI-S1) as an adjunct to fusion in skeletally mature patients. It is intended for plate fixation/attachment to spinous processes for the purpose of achieving supplemental fixation in the following conditions: degenerative disc disease (as previously defined); spondylolisthesis; trauma; and/or tumor.

    In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be connected to the VERTEX® Reconstruction System with the VERTEX® rod connector. Refer to the VERTEX® Reconstruction System Package Insert for a list of the VERTEX® indications of use.

    Device Description

    The CD HORIZON® Spinal System consists of a variety of shapes and sizes of rods, hooks, screws, CROSSLINK® Plates, staples and connecting components, as well as implant components from other Medtronic spinal systems, which can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case.

    A subset of CD HORIZON® Spinal System components may be used for posterior pedicle screw fixation in pediatric cases. These constructs may be comprised of a variety of shapes and sizes of rods (ranging in diameter from 3.5mm to 6.35mm), hooks, screws, CROSSLINK® Plates, and connecting components. Similarly to the CD HORIZON® implants used in adult cases, these components can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case.

    Certain components within the CD HORIZON® Spinal System are specifically excluded for use in pediatric patients. These include PEEK rods, CD HORIZON® SPIRE spinous process plate device, Shape Memory Alloy Staples, DYNALOK® bolts, and TSRH® screws and washers. All screws used in pediatric cases are only cleared for use via a posterior approach. All of the components used in pediatric cases are fabricated from medical grade stainless steel, medical grade titanium, titanium alloy, and medical grade cobalt-chromium-molybdenum alloy.

    Certain implant components from other Medtronic spinal systems can be used with the CD HORIZON® Spinal System in non-pediatric cases. These components include TSRH® rods, hooks, screws, plates, CROSSLINK® plates, connectors, staples, washers, GDLH® rods, hooks, connectors and CROSSLINK® bar and connectors; LIBERTY® rods and screws; DYNALOK® PLUS and DYNALOK CLASSIC® bolts along with rod/bolt connectors; and Medtronic Multi-Axial rods and screws. Please note that certain components are specifically designed to connect to ~ 3.5mm, ~ 4.5mm, ~ 5.5mm rods or ~ 6.35mm rods, while other components can connect to both ~ 5.5mm rods and ~ 6.35mm rods. Care should be taken so that the correct components are used in the spinal construct.

    CD HORIZON® hooks are intended for posterior use only. CD HORIZON® staples and CD HORIZON® ECLIPSE® rods and associated screws are intended for anterior use only. However, for patients of smaller stature and pediatric patients. CD HORIZON® 4.5mm rods and associated components may be used posteriorly.

    The CD HORIZON® Spinal System implant components are fabricated from medical grade stainless steel, medical grade titanium, titanium alloy, medical grade cobaltchromium-molybdenum alloy, or medical grade PEEK OPTIMA-LT1. Certain CD HORIZON® Spinal System components may be coated with hydroxyapatite.

    Never use stainless steel and titanium implant components in the same construct.

    Medical grade titanium, titanium alloy, and/or medical grade cobalt-chromiummolybdenum alloy may be used together. Never use titanium, titanium alloy, and/or medical grade cobalt-chromium-molybdenum alloy with stainless steel in the same construct.

    The CD HORIZON® Spinal System also includes anterior staples made of Shape Memory Alloy (Nitinol - NiTì). Shape Memory Alloy is compatible with titanium, titanium alloy, and cobalt-chromium-molybdenum alloy. Do not use with stainless steel. These staples are not to be used in pediatric patients.

    PEEK OPTIMA-LT1 implants may be used with stainless steel, titanium, or cobaltchromium-molybdenum alloy implants. CD HORIZON® PEEK Rods are not to be used with CROSSLINK® Plates or in pediatric patients.

    The purpose of this 510(k) was to add modified setscrews manufactured out of medical grade titanium alloy to the CD HORIZON® Spinal System for use with 4.75mm diameter rods.

    AI/ML Overview

    The provided 510(k) summary for the CD HORIZON® Spinal System (K102555) describes the acceptance criteria and the study that proves the device meets those criteria. The study focuses on the mechanical performance of modified setscrews added to the system, demonstrating their substantial equivalence to predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    Test MethodAcceptance CriteriaReported Device Performance
    Axial GripNot explicitly stated, but implied to be comparable to predicate devices and acceptable for intended use.Subject devices met the pre-determined acceptance criteria.
    Static Flexion/ExtensionNot explicitly stated, but implied to be comparable to predicate devices and acceptable for intended use.Subject devices met the pre-determined acceptance criteria.
    Flexion/Extension FatigueNot explicitly stated, but implied to be comparable to predicate devices and acceptable for intended use.Subject devices met the pre-determined acceptance criteria.

    Note on Acceptance Criteria: The document states, "The subject devices met the pre-determined acceptance criteria for all test methods." However, the specific quantitative values for these criteria are not provided in this summary. It is implied that these criteria were established based on the performance of the legally marketed predicate devices, which the new setscrews needed to meet or exceed for substantial equivalence. The relevant standard for testing was ASTM F1798.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated. The document mentions "the subject setscrews were subjected to interconnection testing" and "test methods included the following," suggesting that samples of the new setscrews were tested.
    • Data Provenance: The testing was non-clinical mechanical testing, performed in accordance with ASTM F1798. It originates from the manufacturer's (Medtronic Sofamor Danek USA, Inc.) internal testing or contracted lab testing. The country of origin for the data is not specified, but the company is based in Memphis, Tennessee, USA. The data is prospective in the sense that new tests were conducted on the modified setscrews.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    This is not applicable to this submission. The "ground truth" here is mechanical performance, established through standardized engineering tests (ASTM F1798), not through human expert assessment of clinical data or images.

    4. Adjudication Method for the Test Set

    This is not applicable. Mechanical test results are objective measurements and do not require expert adjudication in the same way clinical interpretations might.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This submission is for mechanical components (setscrews) of a spinal system and focuses on their physical performance rather than human interpretive performance with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This is not applicable. This is not an AI/algorithm-based device.

    7. The Type of Ground Truth Used

    The "ground truth" in this context is the objective mechanical performance of the setscrews as measured by standardized tests (Axial Grip, Static Flexion/Extension, Flexion/Extension Fatigue) against pre-determined acceptance criteria, which are implicitly derived from the performance of legally marketed predicate devices. This is a type of objective physical performance data or engineering test data.

    8. The Sample Size for the Training Set

    This is not applicable. There is no concept of a "training set" for this type of mechanical device submission.

    9. How the Ground Truth for the Training Set was Established

    This is not applicable as there is no training set mentioned in the context of this device submission.