510(k) Data Aggregation

    K Number
    K102512

    Validate with FDA (Live)

    Date Cleared
    2011-05-20

    (261 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    RMS Subcutaneous Needle Sets are intended for the delivery of medication to the subcutaneous tissue.

    Device Description

    The RMS Subcutaneous Needle Set is a Class II Intravascular Administration Set per 21 CFR 880.5440 and intended for the delivery of medication to the subcutaneous tissue. Each set consists of a sterile packaged kit including the infusion set and a commercially available adhesive dressing used to hold the device in place. The infusion set has a 90-degree stainless steel needle mounted to a butterfly on one end, and a luer lock on the other end, connected by medical grade tubing. Each set is equipped with a slide clamp used to stop flow immediately, as well as a snap together capture arrangement used to close the set upon completion. RMS Subcutaneous Needle Sets are available as single sets, as well as 2-needle, 3-needle, 4-needle, 5-needle and 6-needle assemblies, with the use of a low residual volume Y or multi-connector. The device is for single use only.

    AI/ML Overview

    The provided text describes a 510(k) summary for the RMS Subcutaneous Needle Set, which aims to demonstrate substantial equivalence to predicate devices rather than proving a device meets specific acceptance criteria in the sense of a novel AI or diagnostic device. Therefore, several of the requested sections about AI-specific criteria (e.g., sample size for AI test and training sets, number of experts for ground truth, adjudication methods, MRMC studies) are not applicable to this type of submission.

    However, I can extract information related to performance testing that served to support the claim of substantial equivalence.

    1. Table of Acceptance Criteria and Reported Device Performance

    For a 510(k) submission, the "acceptance criteria" are implicitly defined by demonstrating that the device performs similarly to or better than the legally marketed predicate devices, and that any differences do not raise new questions of safety or effectiveness. The reported performance is compared directly to the predicate devices.

    Performance CharacteristicAcceptance Criteria (Implicit: Comparable or Better than Predicate)RMS Device Performance (K102512)Predicate Device 1 Performance (Evans Medical, K020530)Predicate Device 2 Performance (Marcal Medical, K082818)
    Fluid Flow RateComparable or better than predicate devices1996.35 (Total for 26G)838.98 (Total for 27G), 1988.98 (Total for 24G)4041.17 (Total for 24G)
    Priming/Residual VolumeComparable or better than predicate devicesLower values across all needle counts (e.g., 0.074ml for 1-needle)Higher values across all needle counts (e.g., 0.23ml for 1-needle)Higher values across all needle counts (e.g., 0.15ml for 1-needle)
    Needle Tip DimensionsNo substantial difference from predicate devicesTop Angle: 52.3°, Side Angle: 16.5°, Length: 0.018 inTop Angle: 52.3°, Side Angle: 16.5°, Length: 0.022 inTop Angle: 51.9°, Side Angle: 16.3°, Length: 0.016 in
    Wing FlexibilityEqual or better (less force to flex) than predicate devicesMean: 22 grams (less force)Mean: 39 gramsMean: 48 grams
    Sterility 10-6PASSPASS(Not explicitly stated for predicates, assumed to meet standard)(Not explicitly stated for predicates, assumed to meet standard)
    BiocompatibilityPASS for various tests (Cytotoxicity, Irritation, Sensitization, Hemocompatibility)PASS for all listed tests(Implicitly comparable as per textual claim for thrombogenicity and hemocompatibility)(Implicitly comparable as per textual claim for thrombogenicity and hemocompatibility)
    MaterialsSimilar enough not to raise new issues of safety/effectivenessPVC, Polypropylene, 3M DressingPVC, Polyethylene, 3M DressingPVC, Polyethylene

    2. Sample Size Used for the Test Set and Data Provenance

    • Fluid Flow Rate Testing: The table shows results for 4 individual measurements for each needle set configuration (Needle 1, 2, 3, 4). The "Total" represents the sum of these four measurements for each device type. Therefore, a sample size of 4 measurements per device type/configuration was used.
    • Priming/Residual Volume Measurements: Not explicitly stated, but measurements are provided for 1-needle to 6-needle configurations for each device.
    • Needle Tip Measurements: One measurement is provided for each characteristic (Top Angle, Side Angle, Length) for each company.
    • Wing Flexibility: For each of the three manufacturers (RMS, Marcal, Evans Medical), 4 individual "Butterfly" measurements are provided for flexibility, along with a calculated mean.
    • Biocompatibility/Sterility: These are standard compliance tests, typically performed on a statistically significant sample based on the specific ISO standards, but the exact sample sizes for each test are not detailed in this summary.
    • Data Provenance: Not explicitly stated. The tests were performed according to RMS's "Needle Set Fluid Flow Rate Test Procedure" (SOP 5071) and ISO standards. The data is likely from laboratory testing conducted as part of the submission process, not from patient studies or specific geographical regions. It is retrospective in the sense that it was generated for the submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable. The device is a medical accessory (subcutaneous needle set), and the testing performed involves objective physical and material performance measurements (e.g., flow rate, volumes, dimensions, flexibility, biocompatibility), not diagnostic image analysis or clinical assessment requiring expert consensus ground truth.

    4. Adjudication Method for the Test Set

    This is not applicable as the tests are objective laboratory measurements, not subjective evaluations requiring adjudication.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

    This is not applicable. This is a 510(k) submission for a physical medical device, not an AI/diagnostic software.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    This is not applicable. This is a 510(k) submission for a physical medical device.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    The "ground truth" for the performance comparisons in this 510(k) submission consists of objective physical measurements, material properties, and compliance with specified standards (e.g., ISO for sterility and biocompatibility). The predicate devices' performance serves as the benchmark for demonstrating substantial equivalence.

    8. The Sample Size for the Training Set

    This is not applicable as there is no "training set" for physical device performance testing in this context.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable as there is no "training set" for physical device performance testing in this context.