    The Verify® Chemical Monitoring Strip for Resert Solutions Version 2 is a high level disinfectant concentration monitor dedicated for use with Resert® XL HLD High-Level Disinfectant. The purpose of the Verify Chemical Monitoring Strip for Resert Solutions Version 2 is to determine whether the concentration of a Resert® XL HLD High-Level Disinfectant solution is above the minimum recommended concentration (MRC) of 1.5%.

    Device Description

    The Verify Chemical Monitoring Strip for Resert Solutions Version 2 is a chemical indicator strip consisting of an absorbent paper pad impregnated with the reactive chemicals that is adhesively bonded to one end of a polymer film. The Verify Chemical Monitoring Strip for Resert Solutions Version 2 has been developed to monitor the Resert 11 XL HLD High-Level Disinfectant that has a minimum recommended concentration (MRC) of 1.5%.

    AI/ML Overview

    The provided document describes the K102296 510(k) submission for the "Verify® Chemical Monitoring Strip for Resert Solutions Version 2". This is a chemical indicator device, not an AI/ML medical device. Therefore, many of the requested elements for AI/ML device studies (such as MRMC studies, human reader improvement, and training set details) are not applicable.

    Below is a description of the acceptance criteria and study information that can be extracted from the provided text for this specific device type.


    Device Name: Verify® Chemical Monitoring Strip for Resert Solutions Version 2

    Description: A chemical indicator strip designed to monitor the concentration of Resert XL HLD High-Level Disinfectant to determine if it is above the minimum recommended concentration (MRC) of 1.5%.


    1. Table of Acceptance Criteria and Reported Device Performance

    Test of 3 LotsAcceptance Criteria @ 20° and 24°CReported Device Performance (Result)
    Performance TestingFAIL @ 1.5%, PASS @ 1.8%Pass
    Blind Study Testing100%Pass
    Simulated Use (Contaminants) Testing≥ 50%Pass
    Test Strip Life Outside the Bottle100%Pass
    Aggressive Chemical Stability Testing≥ 50%Pass
    Specificity Testing for hydrogen peroxide100%Pass
    Stability Testing in three storage environments through labeled shelf life≥ 50%Pass
    In-Use Stability Testing in three storage environments through labeled opened bottle shelf life100%Pass
    ≥ 50%

    Note on Performance Testing Acceptance Criteria: The row for "Performance Testing" includes two criteria: "FAIL @ 1.5%" and "PASS @ 1.8%". This implies that the device should indicate a "fail" when the concentration is at the MRC (1.5%) and a "pass" when it is above the MRC (1.8%), demonstrating its ability to distinguish between concentrations at or below the MRC and those above it. The "Result" for this test is simply "Pass", indicating that these criteria were met.


    2. Sample Size Used for the Test Set and Data Provenance

    The document states that a "Test of 3 Lots" was performed for all listed verification activities. This implies that the testing was conducted on samples from three different manufacturing lots of the device.

    • Sample Size (Test Set): Samples from 3 lots. Specific number of strips per lot is not provided.
    • Data Provenance: Not explicitly stated, but based on the device type (chemical indicator) and the nature of the tests (laboratory testing of chemical reactions and stability), this would be considered prospective data generated in a controlled laboratory environment. The country of origin of the data is not specified.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Not applicable. This is a chemical indicator device, and its performance is determined by its chemical reaction to a specific concentration of a disinfectant solution. The "ground truth" is the known chemical concentration of the solution, not an expert assessment.


    4. Adjudication Method for the Test Set

    Not applicable. The performance is based on the chemical reaction of the strip, not on subjective human interpretation that would require adjudication. The "Blind Study Testing" mentioned in the table suggests that the evaluators were blinded to the expected outcome, which is a form of control, but not an adjudication method for ground truth.


    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Human Readers Improvement

    Not applicable. This is not an AI/ML device that assists human readers.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    While this is a "standalone" device in the sense that it doesn't integrate with an algorithm, the term "standalone performance" typically refers to the performance of an AI algorithm without human input. This device's performance is intrinsically linked to its intended use by a human visually interpreting the result. Therefore, the concept of standalone algorithm performance does not apply. The performance testing described is the standalone performance of the device as a chemical indicator.


    7. The Type of Ground Truth Used

    The ground truth for this device is the known chemical concentration of the Resert XL HLD High-Level Disinfectant solutions. For example, for the "Performance Testing," the ground truth was solutions at 1.5% (expected FAIL) and 1.8% (expected PASS).


    8. The Sample Size for the Training Set

    Not applicable. This is a chemical indicator; its design is based on chemical principles and formulation, not on a machine learning training set.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable. No training set as described for AI/ML devices.