    K Number
    K102024

    Validate with FDA (Live)

    Date Cleared
    2010-10-15

    (88 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RPM Respiratory Gating System is used to obtain tracking of the subject respiratory pattern for respiratory synchronized image acquisition and radiation therapy treatment. It can also be used to monitor the patient position during image acquisition, simulation and treatment.

    Device Description

    The Varian RPM Respiratory Gating is an attachment to radiation therapy treatment systems such as conventional linear accelerators, proton therapy systems, radiation therapy simulators and image acquisition devices used for diagnostics and radiation therapy. The RPM device is to be used to characterize the patient's respiratory motion information to synchronize their operation with the respiratory motion. The imaging devices either trigger the image acquisition based on the signal received from RPM, or they use respiratory motion signal to trigger beam-hold and limit the beam-on time to those points in the respiratory cycle where the target volume is within acceptable motion limits. The RPM device can also be used to monitor the patient's position during the image acquisition and treatment therapy process.

    AI/ML Overview

    The RPM Respiratory Gating System is an attachment to radiation treatment systems used to characterize a patient's respiratory motion for synchronized image acquisition and radiation therapy treatment, and to monitor patient position during these processes.

    1. Acceptance Criteria and Reported Device Performance:

    The provided document is a 510(k) summary for a substantial equivalence claim, not a detailed study report with specific acceptance criteria and performance data. The table below summarizes the features and specifications of the device, comparing the previously cleared predicate device with the current device. The implication is that the performance of these features meets acceptable standards for the intended use, as no changes were made to these core functional specifications.

    Feature and/or Specification of New/Modified DeviceCleared Device Feature/SpecificationDevice With Change
    Intended UseTracking respiratory pattern for synchronized image acquisition and radiation therapy treatment; monitoring patient position during image acquisition, simulation, and treatment.No change
    Indications for UseTracking respiratory pattern for synchronized image acquisition and radiation therapy treatment; monitoring patient position during image acquisition, simulation, and treatment.No change
    Objects physically attached to patientLight plastic block with retro-reflective markers (2 or 6 dots), typically on upper abdomen or chest.No change
    Wire connections made to patientNoneNo change
    Measured parametersVertical, lateral, and longitudinal position of the marker block.No change
    Maximum signal range32 cm vertical by 41 cm lateral at 380 cm working distance (50 mm lens); 29 cm vertical by 38 cm lateral at 180 cm working distance (25 mm lens).No change
    Measurement resolution (at 380 cm, 50 mm lens)Vertical: 0.013 cm; Lateral: 0.023 cm; Longitudinal: 0.055 cm.No change
    Measurement resolution (at 180 cm, 25 mm lens)Vertical: 0.004 cm; Lateral: 0.002 cm; Longitudinal: 0.018 cm.No change
    Measurement driftNoneNo change
    Distance to patient15 ft maximum patient-camera distance.No change
    Compatible with radiation therapy treatment devicesYes (Varian accelerators)Varian Linear Accelerators, Proton therapy Systems that are capable of being gated.
    Compatible with radiation therapy simulatorsYesNo change
    Compatible with Image acquisition systemsYes (General interface); support for triggered/gated and 4D image acquisition.No change
    Supported patient coaching methodsAudio and/or visualNo change
    Supported breathing methodsFree-breathing, voluntary breath-holdNo change
    Patient position monitoringYesNo change
    Recognition and management of periodicity changesYesNo change
    Support for phase-based and amplitude-based gatingBoth phase and amplitude based gatingNo change

    2. Sample Size Used for the Test Set and Data Provenance:

    The document states, "Results of verification and validation testing demonstrate that the RPM Respiratory Gating System satisfies the intended use as described above." However, this 510(k) summary does not provide specific details on:

    • The sample size used for the test set.
    • The country of origin of the data.
    • Whether the data was retrospective or prospective.

    Given that this is a 510(k) for a device with "No change" to most of its core features and specifications compared to a predicate device, it's highly probable that the "verification and validation testing" referred to was an internal engineering and performance evaluation, rather than a clinical study with human subjects.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not provided in the 510(k) summary. Given the absence of specific clinical study details, it's unlikely that "experts" were formally used to establish ground truth in the context of clinical accuracy for this type of device in this submission. The "ground truth" would more likely refer to established engineering benchmarks and measurements of physical parameters.

    4. Adjudication Method for the Test Set:

    This information is not provided in the 510(k) summary.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This type of study is not applicable to the RPM Respiratory Gating System as described. This device is an attachment to radiation therapy systems for tracking respiratory patterns and monitoring patient position. It is not an AI-assisted diagnostic or interpretive tool that would involve human readers or affect their diagnostic performance.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    The device itself is a standalone system in the sense that it performs its function of tracking respiratory patterns and providing gating signals without direct human interpretation of the tracking data in real-time. The "verification and validation testing" mentioned would assess the standalone performance of its measurement accuracy, resolution, drift, and compatibility. However, the details of such standalone testing are not explicitly described in terms of methodology or specific results.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):

    The type of "ground truth" for this device would relate to the physical accuracy of its measurements (e.g., how accurately it measures the position of the marker block, its resolution, and drift). This would be established through engineering standards and calibrated measurement tools, not typically through expert consensus, pathology, or outcomes data, which are more relevant for diagnostic or prognostic devices.

    8. The Sample Size for the Training Set:

    This information is not provided and is likely not relevant for this device. The RPM Respiratory Gating System, as described, is a sensing and tracking system, not a machine learning or AI algorithm that requires a "training set" in the conventional sense. Its functionality is based on established optical tracking principles.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not provided and is likely not relevant for this device, as it does not appear to use a training set in the machine learning context.