510(k) Data Aggregation

    K Number
    K102021

    Validate with FDA (Live)

    Device Name
    CRYO-TOUCH II
    Manufacturer
    Date Cleared
    2010-09-17

    (60 days)

    Product Code
    Regulation Number
    882.4250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MyoScience Cryo-Touch II is used to produce lesions in peripheral nervous tissue by the application of cold to the selected site for the blocking of pain. The Cryo-Touch II is not indicated for treatment of central nervous system tissue.

    Device Description

    The MyoScience Cryo-Touch II is a hand-held, single patient-use disposable, cryogenic device used to produce lesions in peripheral nervous tissue by the application of cold to the selected site for the blocking of pain. The device is based on introduction of a cryoprobe cooled by the cryogenic fluid (liquid nitrous oxide) to the target nervous tissue and controlled by the user via a freeze control. An iceball forms around the tip of the needle assembly and the adjacent nerve. The Cryo-Touch II may be used in conjunction with a standard off-the-shelf nerve stimulator device in applications where precise nerve location is desired. The device is comprised of three main components: a control unit (controller), probe assembly, and a cryogen cylinder assembly (Nitrous Oxide). The entire device is hand-held and for single patient-use. The Cryo-Touch II controller is battery operated; the battery powers electronics within the controller and provide feedback to the user when the device is ready to use. An assortment of needle assemblies is available for the Cryo-Touch II, ranging from a single to a triple needle, from 25 gauge to 30 gauge and 6mm in length. All needle assemblies are made of stainless steel and have a closed-tip that fully contains the cryogen so that it does not enter the target tissue. The needle assembly is the sterile patient contacting component of the Cryo-Touch II. The user removes the needle assembly from the sterile packaging and attaches it to the controller by inserting the needle hub end into the distal end of the controller. The cryogen is provided as a nitrous oxide cylinder. The cartridge is filled with liquid N₂O (83%) and the cylinder treats for approximately 180 seconds. The cylinder, once empty, may be replaced with a new cylinder by the user.

    AI/ML Overview

    The provided 510(k) summary for the MyoScience Cryo-Touch II is for a cryosurgical unit and accessories, which is a Class II device. This type of device relies on established physical principles for its operation (cryotherapy) rather than complex algorithms or AI for diagnostic/interpretive functions. Therefore, many of the requested categories related to algorithm performance, ground truth, expert consensus, and comparative effectiveness studies with AI are not applicable to this type of medical device submission.

    The "acceptance criteria" for a device like the Cryo-Touch II are typically its ability to function as intended, consistently apply the specified temperature, and meet safety and performance standards. The "study" mentioned refers to verification testing to ensure these criteria are met.

    Here's the breakdown based on the provided text, with "N/A" for categories not applicable to this type of device submission:

    Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied from submission)Reported Device Performance
    Device meets design requirements for accurate temperature profilingVerification testing demonstrated the product met the design requirements for accurate temperature profiling during simulated use conditions.
    Consistent application of cold to selected site for blocking pain(Demonstrated through functional testing, consistent with predicate device)
    Operational reliability (e.g., battery operation, cryogen flow, iceball formation)(Demonstrated through functional testing, consistent with predicate device)
    Patient-contacting materials are biocompatible(Confirmed as identical to predicate, implying biocompatibility established)
    Treatment Temperature Range of -55° C to -75° C(Confirmed as identical to predicate)

    Detailed Information:

    1. Sample size used for the test set and the data provenance:

      • The document mentions "Verification testing was performed on the Cryo-Touch II device to demonstrate that the product met the design requirements for accurate temperature profiling during simulated use conditions." It does not specify a sample size for this testing nor the data provenance (country of origin, retrospective/prospective). This type of non-clinical testing typically involves a set number of devices or components tested under controlled laboratory conditions to ensure manufacturing and functional specifications are met.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • N/A. This device does not involve diagnostic interpretation or AI algorithms that require expert-established ground truth for a test set in the traditional sense. The "ground truth" for a cryosurgical device is its physical performance (e.g., temperature output, iceball size, durability), which is assessed through engineering and functional testing against design specifications, not expert interpretation of outputs.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • N/A. Adjudication methods are typically used in studies where human interpretation (e.g., reading medical images) is being evaluated and discrepancies exist. For performance testing of a physical device like this, results are objective measurements against specifications.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • N/A. This device is a surgical tool, not an AI-assisted diagnostic or interpretive system. Therefore, no MRMC study comparing human reader performance with or without AI assistance would have been conducted or is relevant for its 510(k) clearance.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • N/A. This is not an algorithm-only device. It is a handheld surgical instrument operated by a medical professional.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For this device, the "ground truth" or reference standard for its performance is its engineering specifications and functional requirements (e.g., achieving a specific temperature range, proper cryogen flow, iceball formation). These are objective, measurable parameters, not subjective expert consensus, pathology, or outcomes data.
    7. The sample size for the training set:

      • N/A. This is not an AI/machine learning device; therefore, there is no "training set."
    8. How the ground truth for the training set was established:

      • N/A. As there is no training set, this question is not applicable.