510(k) Data Aggregation

    K Number
    K101823

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-01-07

    (191 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Arthrex SwiveLock Anchors are intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow, and hip in the following procedures:

    Shoulder: Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction.
    Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair/Tendon Repair, Bunionectomy.
    Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis.
    Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction.
    Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial Collateral Ligament Reconstruction, Lateral Epicondylitis Repair.
    Hip: Capsular Repair, acetabular labral repair

    Device Description

    The SwiveLock Anchor is a two-component, knotless suture anchor comprised of an eyelet and a hollow anchor body. The SwiveLock Anchor is pre-mounted on a driver with the anchor body and eyelet physically separated on the driver shaft. FiberWire suture may also be provided with the device.

    AI/ML Overview

    The provided 510(k) summary (K101823) describes the Arthrex SwiveLock Anchors, a medical device intended for fixing suture (soft tissue) to bone in various anatomical regions. The device is seeking substantial equivalence to previously marketed predicate devices.

    1. Acceptance Criteria and Reported Device Performance:

    The primary acceptance criteria for the Arthrex SwiveLock Anchors are based on demonstrating substantial equivalence to existing predicate devices. Since this is an equivalence submission, specific quantitative acceptance criteria for absolute performance metrics are not explicitly stated in the summary. Instead, the performance is compared against the known performance of the predicate devices.

    The study aimed to demonstrate that the SwiveLock Anchors perform comparably to the predicate devices in terms of:

    Acceptance Criteria (Implied by Substantial Equivalence)Reported Device Performance
    Degradation (for Bio/BioComposite materials)26-week degradation data demonstrated substantial equivalence.
    Mechanical Pull-Out (Tensile) StrengthMechanical pull-out (tensile) testing data demonstrated substantial equivalence to the pull-out forces of the predicate devices.
    Insertion CharacteristicsInsertion testing data demonstrated substantial equivalence.
    In-vivo Performance (Histology, Pull-Out, Imaging)Animal testing (including histology, in-vivo pull-out, and in-vivo imaging) data demonstrated no new issues of safety and effectiveness, implying comparable in-vivo performance.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the specific sample sizes for each test mentioned (degradation, mechanical pull-out, insertion, animal testing). It only refers to "data" for each category.

    The data provenance is not specified. It is an industry standard to conduct these tests in a controlled laboratory environment or animal models. Neither the country of origin nor whether the data is retrospective or prospective is mentioned.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    This information is not applicable to this type of 510(k) submission. Establishing "ground truth" by experts is typically relevant for diagnostic or AI-driven devices where human interpretation is involved. For a mechanical device like a suture anchor, the "truth" is established through physical and biological testing against established standards or predicate device performance.

    4. Adjudication Method:

    This information is not applicable. Adjudication methods (like 2+1, 3+1) are used to resolve disagreements among human readers/experts in diagnostic studies, which is not relevant for the mechanical and biological testing of a suture anchor.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for diagnostic devices where human readers interpret medical images or data, and the effect of AI assistance on their performance is evaluated.

    6. Standalone (Algorithm Only) Performance Study:

    No, a standalone (algorithm only) performance study was not done. This device is a physical medical implant, not an algorithm or AI system. Therefore, standalone algorithm performance is not relevant.

    7. Type of Ground Truth Used:

    The "ground truth" in this context is established through physical and biological testing methodologies. This includes:

    • Degradation data: Measuring the breakdown of biomaterials over time.
    • Mechanical pull-out (tensile) testing data: Quantifying the force required to pull the anchor out of bone, adhering to engineering standards.
    • Insertion testing data: Assessing the ease and characteristics of device insertion.
    • Animal testing data: Observing the biological response (histology), in-vivo pull-out strength, and imaging of the device in a living system.

    The "truth" is rooted in the physical and biological properties and performance of the device as measured by these tests, compared against the known performance of predicate devices.

    8. Sample Size for the Training Set:

    This information is not applicable. "Training set" refers to data used to train machine learning algorithms. The Arthrex SwiveLock Anchor is a mechanical device, not an AI or algorithm-driven system.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable for the reasons stated above.