    K101815

    Validate with FDA (Live)

    Date Cleared
    2010-12-20

    (173 days)

    Product Code
    Regulation Number
    888.3510
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EnduRo Knee System is indicated for use in reconstruction of the diseased knee joint caused by osteoarthritis, rheumatoid arthritis, post-traumatic arthritis, the need to revise failed arthroplasties or osteotomies where pain, deformity or dysfunction persist, and for patients suffering from correctable valgus or varus deformity and moderate flexion contracture.

    Hinge knee systems are designed for use in patients in primary or revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments are absent or insufficient. The femoral and tibial augments are to be attached to their respective components with a fixation screw or screws.

    The EnduRo Knee System is intended for cemented use only.

    Device Description

    The EnduRo Knee System is a cemented prosthesis with a rotating hinge design. The femoral component, tibial plateau and extension stems are manufactured from CoCrMo. The tibial "gliding surfaces" and patella are manufactured from UHMWPE. The tibial mask is made from PEEK Optima® (LT1). The axial sleeve and femoral bushing components are produced from PEEK Optima® (LT1CA30). The system is made up of numerous components available in various sizes. The femoral, tibial plateau, and extension stems are also available with a ZrN (zirconium nitrate) coating. All components are sterile and for single use only.

    AI/ML Overview

    The provided text describes the EnduRo Knee System, a total knee replacement system, and its substantial equivalence to predicate devices, but does not contain information about acceptance criteria or a study that proves the device meets those criteria, as typically understood in the context of diagnostic AI/ML devices.

    The document is a 510(k) summary for a medical device (a knee prosthesis), not an AI/ML diagnostic. Therefore, the requested information categories (acceptance criteria, sample sizes, expert ground truth, adjudication methods, MRMC studies, standalone performance, ground truth types, training set details) are not relevant to this type of device submission and are not present in the provided text.

    Instead, the document focuses on:

    • Substantial Equivalence: Comparing the EnduRo Knee System to existing legally marketed predicate devices (NexGen Complete Knee Solution Rotating Hinge Knee and Aesculap Columbus REVISION Knee System).
    • Device Description: Detailing the components and materials of the EnduRo Knee System.
    • Indications for Use: Specifying the medical conditions for which the device is intended.
    • Technological Characteristics: Stating that the device has similar shapes, sizes, and materials as the predicate devices.
    • Performance Data: Listing a series of engineering and mechanical tests performed to support substantial equivalence. These tests include:
      • Range of Motion
      • Wear testing
      • Femoral Endurance Properties
      • Endurance Properties of Modular Femoral Stem Assembly
      • Tibial Tray Endurance Properties
      • Endurance Properties of the rotating axis component in A/P direction
      • Endurance Properties of the Modular Tibial Stem Assembly
      • Varus-Valgus Endurance Properties
      • TibioFemoral Contact Area/Stress at Different Angles of Flexion
      • PatelloFemoral Lateral Subluxation Resistance
      • PatelloFemoral Contact Area/Stress at Different Angles of Flexion

    Therefore, I cannot provide the requested table or answer the specific questions related to acceptance criteria and studies in the context of AI/ML diagnostic devices based on the provided text. The document pertains to a mechanical orthopedic implant, for which "acceptance criteria" and "studies" are typically defined by biomechanical standards and in-vitro testing, rather than clinical performance metrics with ground truth established by experts.