Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K101262

    Validate with FDA (Live)

    Device Name
    FLEXI-LITE
    Date Cleared
    2010-06-10

    (36 days)

    Product Code
    Regulation Number
    868.5470
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Flexi-Lite Hyperbaric chamber is a hyperbaric chamber intended to be used in treating mild symptoms consistent with Acute Mountain Sickness (AMS) as prescribed by or under the direction of a physician. Caution: Federal law restricts this device to sale by or on the order of a physician.

    Device Description

    The Flexi-Lite flexible hyperbaric chamber is a mild hyperbaric chamber for pressures less than 4 psi. This lightweight and portable chamber utilizes atmospheric Air as supplied by a GAST model 0523/1023 "oil-less, breathable air compressor" to pressurize the chamber and provide a suitable environment for the occupant. The Flexi-Lite construction utilizes a dual bag design with the inner the ooonpining the pressure and an exterior bag to provide structural support. All components are attached to the inner bag utilizing bulkhead connections. It is outfitted with two externally mounted metal relief valves (set at 4 psi), two metal Air addition valves (one external & one internal), two metal depressurization valves (one external & one internal), two pressure gauges (one external & one internal), and one metal Air sampling port. In total it weighs 39 lbs (compressor adds another 54 lbs).

    AI/ML Overview

    Here's an analysis of the provided text regarding the Flexi-Lite Hyperbaric Chamber (K101262), focusing on acceptance criteria and supporting studies.

    Based on the provided 510(k) summary, the device is a hyperbaric chamber, not an AI/ML-driven device, diagnostic algorithm, or medical imaging analysis tool. Therefore, many of the requested categories related to algorithm performance, expert review, and ground truth for AI studies are not applicable to this submission.

    The document describes a traditional medical device submission for substantial equivalence to a predicate device. The "acceptance criteria" discussed are primarily related to the device's functional and safety characteristics compared to its predicate.


    1. Table of Acceptance Criteria and Reported Device Performance

    CriterionAcceptance Standard (Predicated on K051759)Reported Device Performance (Flexi-Lite)Met?
    Intended UseTreating mild symptoms consistent with Acute Mountain Sickness (AMS) as prescribed by or under the direction of a physician.The Flexi-Lite Hyperbaric chamber is a rugged & portable hyperbaric chamber intended to be used in treating mild symptoms consistent with Acute Mountain Sickness (AMS) as prescribed by or under the direction of a physician.Yes (SE)
    Intended PopulationPersons with High altitude mountain sicknessPersons with High altitude mountain sicknessYes (SE)
    Intended Environment of UseHome, Physician office, outdoor, hospital, subacute facility, EMSHome, Physician office, outdoor, hospital, subacute facility, EMSYes (SE)
    Relief ValvesPresent (Qty and material not specified for predicate, but assumed to be functionally equivalent)Qty 2 metal (Brass and stainless for materials)Yes (SE)
    Dump ValveSeparate dump valveYes - separateYes (SE)
    Operating Pressure (psi)2-4 psi2-4 psiYes (SE)
    Method of InflationCompressor (GAST Model 0523/1023 assumed for predicate or equivalent)GAST Model 0523/1023 CompressorYes (SE)
    Chamber Material(Material not explicitly detailed for predicate, but assumed to be functionally equivalent)Inner bag - 880 denier-urethane coated nylon Outer bag 2 x 2 basket weave, 2-side - urethane coated nylonYes (SE)
    Compressor TypeOil-less compressor (GAST 0523/1023 assumed for predicate or equivalent)GAST 0523/1023 oil-lessYes (SE)
    Air Filtration on CompressorYesYesYes (SE)
    Pressure GaugeYesYesYes (SE)
    Air Filtration on ChamberYesYesYes (SE)
    Hazard Analysis & Bench TestingMust pass all identified bench tests per hazard analysis."Bench testing was performed as determined by the hazard analysis. The Flexi-Lite passed all bench testing."Yes

    Note: "SE" indicates Substantially Equivalent to the predicate device, K051759.


    2. Sample size used for the test set and the data provenance

    The document explicitly states: "Bench testing was performed as determined by the hazard analysis." This indicates physical testing on the device itself.

    • Sample Size for Test Set: Not explicitly stated for each test, but it would typically involve one or more physical units of the Flexi-Lite hyperbaric chamber. The nature of bench testing implies testing of representative product units rather than a "data set" in the context of an algorithm.
    • Data Provenance: The testing was non-clinical, meaning it focused on engineering performance and safety. It doesn't involve patient data, country of origin of data, or retrospective/prospective studies in the sense of clinical trials. The data would originate from the manufacturer's internal testing facilities or an accredited testing lab.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The "ground truth" for this type of device (a physical hyperbaric chamber) is its functional performance, safety features (e.g., relief valve settings), and material specifications, as validated by engineering tests against predetermined criteria or predicate device specifications. This doesn't involve "experts" establishing ground truth in the way a diagnostic algorithm requires clinical expert consensus.


    4. Adjudication method for the test set

    Not applicable. This is not a comparative clinical study requiring adjudication of interpretations. Bench testing results are typically objective measurements or pass/fail determinations based on engineering standards.


    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is not an AI-assisted diagnostic tool. No MRMC study was performed as it is not relevant to a hyperbaric chamber.


    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is not an algorithm.


    7. The type of ground truth used

    The "ground truth" for this device's substantial equivalence determination relies on:

    • Engineering Specifications: Adherence to defined physical dimensions, material properties, pressure ranges, and component functionality (e.g., relief valve settings).
    • Safety Standards: Compliance with recognized safety standards relevant to medical devices and pressure vessels.
    • Predicate Device Equivalence: Direct comparison of features and performance parameters to a legally marketed predicate device (Flexi-Lite Hyperbaric Chamber K051759). The predicate device itself serves as the benchmark for acceptable performance and safety.
    • Hazard Analysis: Identification and mitigation of potential hazards, with "bench testing" verifying that these mitigations are effective.

    8. The sample size for the training set

    Not applicable. This is a physical device, not a machine learning model, so there is no "training set."


    9. How the ground truth for the training set was established

    Not applicable. There is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1