Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K101158

    Validate with FDA (Live)

    Date Cleared
    2011-05-06

    (375 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LIC Instrument Processor when used with the Manzi Detergent MD10, the Manzi Sterilant MS10 concentrate ( MEC 0.49% PAA, minimum contact temperature of 120° F for a contact time of 5 minutes), and filtered and ozonated final rinse water, is indicated for cleaning and high level disinfection of heat sensitive semi-critical endoscopes used in health care settings by health care workers. Manual cleaning of medical devices (endoscopes) is not required prior to placement in the LIC Instrument Processor.

    The Manzi Detergent MD10 concentrate when used with the LIC Instrument Processor is indicated for the cleaning of heat sensitive semi-critical endoscopes used in health care settings by health care workers.

    The Manzi Sterilant MS10 concentrate when used in the LIC Instrument Processor is indicated for high level disinfection of heat sensitive semi-critical endoscopes used in health care settings by health care workers.

    The filtered ozonated final rinse water generated by the LIC Instrument Processor is indicated for the final rinsing in the LIC Instrument Processor of of heat sensitive semi-critical endoscopes used in health care settings by health care workers.

    Device Description

    The LIC Instrument Processing System consists of a LIC Instrument Processor, a proprietary Manzi sterilant, MS10; a proprietary enzymatic Manzi Detergent, MD10; and internally generated filtered ozonated final rinse water. The LIC Instrument Processing System requires no connectors to be attached to the endoscope.

    The LIC Instrument Processor is a self-contained stand-alone system designed to clean and provide high level disinfection of semi-critical endoscopes using the MD10 detergent, the MS10 sterilant, and a patented push-pull agitation system. The push-pull agitation system effectively scrubs the interior and exterior surfaces of the semi-critical endoscopes without the use of special connectors. The scope is placed in a processing chamber where it is exposed to a push-pull agitation cleaning cycle that provides cleaning to a protein residual level ≤ 6.4 ug / cm², followed by hot water rinses, a push-pull agitation disinfection cycle that provides high level disinfection of the instrument being processed and final rinses with filtered ozonated water.

    The hardware for the LIC Instrument Processor consists of a stainless steel processing chamber, a push-pull agitation pump, an ozonator, a neutralization tank, and a variety of components that are mounted in a movable covered frame. The cleaner-processor system utilizes accessories such as disposable water filters, reusable semi-critical endoscopes trays, baffle inserts, and printer paper.

    The LIC Instrument Processor is designed to: (1) be used in accordance with the reprocessing instructions provided in the operator's manual of the instruments being processed, and (2) facilitate the health care facility's compliance with reprocessing guidelines published by SGNA, APIC, AORN, ASGE, CDC, and other professional organizations.

    MD10 is a low foaming enzyme chemical detergent packaged in single use containers for attachment to the LIC Instrument Processor. MD10 is intended to be used with the LIC Instrument Processor only.

    MS10 is a peracetic acid based liquid chemical sterilant packaged in single use containers for attachment to the LIC Instrument Processor. MS10 is intended to be used with the LIC Instrument Processor.

    AI/ML Overview

    1. Acceptance Criteria and Reported Device Performance:

    RequirementAcceptance CriteriaReported Device Performance
    Cleaning Efficacy (Protein Laden Soil)Remaining protein < 3.7 ug/cm² when challenged with up to 1080 µg/cm²Remaining protein levels were < 3.7 ug/cm² when challenged with endoscopes contaminated with Protein Laden Soil up to 1080 µg/cm².
    Cleaning Efficacy (TOC Laden Soil)Remaining TOC < 1.5 ug/cm² when challenged with up to 160 µg/cm²Remaining TOC levels were < 1.5 ug/cm² when challenged with endoscopes contaminated with TOC Laden Soil up to 160 µg/cm².
    Cleaning Efficacy (In-use clinical soil)Remaining protein < 1.2 ug/cm²Remaining protein levels were < 1.2 ug/cm² after processing through the system's cleaning step.
    Disinfection Efficacy (Simulated Use)≥ 6 logs spore reduction with no CFUs for Mycobacterium terrae at ≥ 10^6 microbial loading≥ 6 spore log reduction with no CFUs was achieved for Mycobacterium terrae (≥ 10^6 microbial loading) when processed through the disinfection step. Additionally, the full cycle (cleaning, disinfecting, and final rinse) also achieved ≥ 6 spore log reduction with no CFUs for Mycobacterium terrae at ≥ 10^6 microbial loading.
    Disinfection Efficacy (In-use clinical soil)No viable organisms or CFUs recoveredNo viable organisms or CFUs were recovered from clinically used endoscopes following processing through the full cycle.
    Filtered Ozonated Final Rinse Water Efficacy> 6 logs spore reduction with no CFUs when challenged with > 10^0 microbial loading> 6 spore log reduction with no CFUs was achieved when challenged with > 10^0 microbial loading and processed to operational specifications.
    Toxicological Evaluation of ResiduesDetergent and sterilant residuals reduced to non-toxic levelsDetergent and sterilant residuals were reduced to non-toxic levels.
    Process Parameter TestsDevice performs as intended, achieves/maintains specified parameters, aborts for faults, detects faults, provides alarms/instructionsDevice achieved and maintained specified physical process parameters, automatically aborted cycles if parameters not achieved, detected defined fault conditions, and provided designated alarms and instructions in the event of fault condition detection.
    Manzi MS10 Sterilant PotencyPassedPassed
    Manzi MS10 Sterilant Simulated Use TestsPassedPassed
    Manzi MS10 Sterilant In-Use TestsPassedPassed
    Manzi MS10 Sterilant BiocompatibilityPassedPassed
    Manzi MS10 Sporicidal ActivitySporicidal (for Bacillus subtilis and Clostridium sporogenes)MS10 is sporicidal.
    Manzi MS10 Fungicidal ActivityFungicidal (for Trichophyton mentagrophytes)MS10 is fungicidal.
    Manzi MS10 Bactericidal ActivityBactericidal (for Salmonella choleraesuis, Staphylococcus aureus, Pseudomonas aeruginosa)MS10 is bactericidal.
    Manzi MS10 Virucidal TestingVirucidal (for Poliovirus Type I)MS10 is virucidal.
    Manzi MS10 Tuberculocidal TestingTuberculocidal (for Mycobacterium bovis)MS10 is tuberculocidal.

    2. Sample Size for Test Set and Data Provenance:

    • Sample Size: Not explicitly stated for each test beyond "endoscopes" or "inoculated endoscope" in various simulated use and in-use tests.
    • Data Provenance: The document does not specify the country of origin. The study includes both simulated use testing (laboratory-controlled conditions with specific microbial loading) and in-use testing (using endoscopes contaminated by clinical endoscopy with naturally occurring soils). This indicates a mix of prospective (potentially for the simulated use) and retrospective/observational (for in-use samples) approaches.

    3. Number of Experts and Qualifications for Ground Truth:

    Not applicable as the ground truth for this device (an instrument processing system) is based on quantitative laboratory measurements and microbiological assays, not expert interpretation of images or clinical assessments.

    4. Adjudication Method for Test Set:

    Not applicable, as ground truth is established through quantitative measurements and laboratory tests.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    An MRMC study was not done. This device is an automated instrument processing system, not a diagnostic imaging AI tool that assists human readers. Therefore, the concept of human readers improving with AI vs. without AI assistance does not apply.

    6. Standalone Performance:

    Yes, a standalone performance evaluation was done. The entire study describes the performance of the LIC Instrument Processing System (the algorithm/device itself) in cleaning and high-level disinfection. All efficacy tests (cleaning, disinfection, rinse water) and residue evaluations directly assess the device's capabilities without human intervention during the processing cycle.

    7. Type of Ground Truth Used:

    The ground truth used is primarily quantitative laboratory measurements and microbiological assays, including:

    • Protein residual levels (µg/cm²)
    • Total Organic Carbon (TOC) levels (µg/cm²)
    • Microbial loading reduction (log reduction, absence of CFUs)
    • Detection of viable organisms
    • Chemical residue levels

    8. Sample Size for the Training Set:

    Not applicable. This document describes the validation of an automated instrument processing system, not a machine learning or AI model that requires a training set. The system operates based on established physical and chemical principles, not learned patterns from data.

    9. How Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1