Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K100967

    Validate with FDA (Live)

    Date Cleared
    2011-08-05

    (484 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PERIMETER® C Spinal System is intended to be used for anterior cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one level from the C2-C3 disc to the C7-T1 disc who have had six weeks of non-operative treatment. Cervical disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. Additionally, the PERIMETER® C Spinal System implants are to be used with autogenous bone graft and supplemental fixation and implanted via an open, anterior approach.

    Device Description

    The PERIMETER® C Spinal System consists of spacers of various widths and heights, which can be inserted between two cervical vertebral bodies to give support and correction during cervical interbody fusion procedures. Additionally, this implant has six degrees of lordosis and the superior and inferior surfaces of the implant are designed with teeth which interact with the surface of the vertebral endplates to aid in resisting expulsion. The hollow geometry of the implants allows them to be packed with autogenous bone graft and is to be used with supplemental fixation in all procedures. This device is offered in PEEK-OPTIMA® (i.e., POLYETHERETHERKETONE) with tantalum markers. Medical grade titanium implants and medical grade PEEK implants may be used together. Never use titanium or titanium alloy implants with stainless steel in the same construct.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the PERIMETER® C Spinal System, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriterionReported Device Performance
    Mechanical Test Standards:Met:
    Static Torsion (per ASTM F2077-03)Results indicate acceptance criteria were met.
    Static Compression (per ASTM F2077-03)Results indicate acceptance criteria were met.
    Dynamic Compression (per ASTM F2077-03)Results indicate acceptance criteria were met.
    Dynamic Torsion (per ASTM F2077-03)Results indicate acceptance criteria were met.
    Static Compression-Shear (per ASTM F2077-03)Results indicate acceptance criteria were met.
    Subsidence (per ASTM F2267-04)Results indicate acceptance criteria were met.
    Expulsion TestingResults indicate acceptance criteria were met.
    Overall Goal:Met:
    Demonstrate substantial equivalence to predicate devicesSubstantial equivalence demonstrated through mechanical tests and engineering rationale.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided 510(k) summary focuses on non-clinical mechanical testing for the PERIMETER® C Spinal System. Therefore, there is no "test set" in the context of patient data (e.g., medical images, clinical outcomes).

    • Sample Size for Test Set: Not applicable, as this refers to mechanical test specimens, not patient data. The document does not specify the number of individual test articles (implants) used for each mechanical test.
    • Data Provenance: Not applicable. The data comes from in-vitro mechanical testing, not a country of origin for patient data. This is a retrospective analysis of device performance against established ASTM standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. For mechanical testing, the "ground truth" is defined by the established ASTM standards and engineering principles. The results are compared against these predetermined criteria, not against expert human interpretations of patient data.

    4. Adjudication Method for the Test Set

    Not applicable. Mechanical test results are measured and directly compared to quantitative acceptance criteria defined by the ASTM standards. There is no need for adjudication as you would have for subjective interpretations of clinical findings.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

    Not applicable. This submission is for a medical implant (spinal system), not an AI-powered diagnostic or assistive tool. Therefore, an MRMC study comparing human reader performance with and without AI is not relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This is not an algorithm-based device.

    7. The Type of Ground Truth Used

    The ground truth used for demonstrating the device's performance against its acceptance criteria is:

    • Standardized Mechanical Test Criteria: Specifically, the requirements outlined in ASTM F2077-03 (Standard Test Methods for Intervertebral Body Fusion Devices) and ASTM F2267-04 (Standard Test Method for Evaluating the Subsidence of an Intervertebral Body Fusion Device Under Static Axial Compression).
    • Engineering Rationale: Complementary engineering analysis to support the mechanical test results.

    8. The Sample Size for the Training Set

    Not applicable. There is no "training set" as this is not a machine learning or AI-driven device requiring data for model training.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set, there is no ground truth to be established for it.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1