Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K100697

    Validate with FDA (Live)

    Date Cleared
    2010-05-13

    (63 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    powder free vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-06.

    AI/ML Overview

    Here's an analysis of the provided text, focusing on the acceptance criteria and study data for the "Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored)":

    This device, a patient examination glove, is a Class I medical device, which typically requires less rigorous testing than higher-class devices. The primary focus of the testing is to demonstrate substantial equivalence to a predicate device and adherence to recognized standards.

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicsAcceptance Criteria (Standard)Reported Device Performance
    DimensionASTM standard D 5250-06Meets
    Physical PropertiesASTM standard D 5250-06Meets
    Freedom from pinholes21 CFR 800.20Meets
    Powder ResidualASTM standard D 5250-06, D6124-06<2mg/glove
    Biocompatibility (Irritation)Primary Skin Irritation in rabbitsPasses (Not a Primary Skin Irritation)
    Biocompatibility (Sensitization)Dermal sensitization in the guinea pigPasses (Not a Dermal Sensitization)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes used for each test (e.g., for dimension, physical properties, pinholes, powder residual, or biocompatibility). It references adherence to standards like ASTM D5250-06, 21 CFR 800.20, and ISO10993-10, which would typically specify minimum sample sizes for individual tests.

    The data provenance implies that the testing was conducted by the manufacturer, SHIJIAZHUANG UNIVERSAL CHANNEL PLASTIC CO., LTD, located in China. The testing appears to be prospective as it was conducted specifically for this 510(k) submission to demonstrate compliance with the standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of device (patient examination glove) does not typically involve human expert interpretation of results in the way, for example, a diagnostic imaging AI algorithm would.

    • For physical and chemical tests (Dimension, Physical Properties, Freedom from pinholes, Powder Residual), the "ground truth" is established by the specifications of the referenced ASTM standards and 21 CFR regulations. The results are quantitative measurements compared against predefined thresholds.
    • For biocompatibility tests (Primary Skin Irritation, Dermal Sensitization), the "ground truth" is established by observing animal responses (rabbits for irritation, guinea pigs for sensitization) against established protocols (ISO 10993-10) and interpretation by trained toxicologists or scientists. The number and specific qualifications of these experts are not stated in this summary, but would be implicit in the conduct of these standardized tests.

    4. Adjudication Method for the Test Set

    Not applicable in the conventional sense. The "adjudication" is essentially the direct comparison of measured device characteristics against the objective, pre-defined criteria in the referenced standards. There isn't a scenario described where multiple interpretations of data would require an adjudication process by human experts.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. This type of study is typically conducted for diagnostic devices where human readers (e.g., radiologists) interpret medical images or data, and the study compares their performance with and without AI assistance. A patient examination glove is a simple barrier device, not a diagnostic tool, so an MRMC study is not relevant.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    No. This is also not relevant. The device is a physical product (a glove), not an algorithm. The performance is assessed through laboratory testing of its physical and biological properties.

    7. The Type of Ground Truth Used

    The ground truth for this device's performance is established by:

    • Standard Specifications: For mechanical properties (dimension, physical properties, pinholes, powder residual), the ground truth is the quantitative limits defined by ASTM standards (D 5250-06, D6124-06) and 21 CFR 800.20.
    • Biological Response Protocols: For biocompatibility, the ground truth is determined by the observed biological reactions in animal models following the established protocols of ISO 10993-10 (often interpreted as "Pass" or "Fail" based on whether reactions exceed certain thresholds).

    Essentially, it's a combination of objective measurement against predefined criteria and standardized biological testing outcomes.

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI algorithm, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1